Document 8mgvxeJmzby4kXgyN0nG2nam
IDENTIFICATION OF CHLORINATED BIPHENYLS IN BIOLOGICAL RESIDUES
Object
This method describes a procedure for the identification of chlorinated biphenyls. The samples to be analyzed normally are the residues from biological extractions but the method is appropriate to other systems.
Principle
The components of the residue in an extract suspected of containing chlorinated biphenyls are first separated by gas chromatography. The separated materials are then identified by rapid scan mass spectrom etry.
Instrumental Details
The gas chromatographic separations are achieved with a F&M 5751-B chromatograph using a flame ionization detector. The 1/8 inch stain less steel columns contain SE-54 packing on 80-100 mesh Chromosorb W. The injection and detector blocks are maintained at 280C. The column temperature is programmed linearly from 170 to 280C at 4 per minute. Programming the temperature sharpens the peaks with longer retention times and facilitates their identification via mass spec trometry. The helium flow rate through the column is 30 ml per minute. The column effluent is split 1:10 between the flame detector and the mass spectrometer.
A 0.010 inch i.d. stainless steel capillary connects the chromatograph outlet to a Biemann type helium separator. This line is heated to 300C and the separator temperature is 280C. The helium separator is connected to the ion source of an Atlas CH 4B mass spectrometer. The electron impact source is operated at 70 ev and 60 amp filament emission. The source temperature is 280C. The ion source also contains a total ionization monitor whose analog output is proportional to the concentration of ionizable material in the source. This recorded signal produces a trace which resembles the original chromatogram.
Procedure
1.
Residual solvent is evaporated to concentrate the residue to about 0.01%. A 20-30 p liter aliquot of this concentrate is injected into the chro matograph and the temperature program started. The solvent quickly passes through the column and is pumped away through the glass frit in the Biemann separator. A valve between the separator and the mass spectrometer is then opened so that column effluent can pass into the ion source. A mass spectrum is scanned each time a peak is observed on the total ion monitor trace. After each component of the extract has been separated and scanned, the spectra are examined and the com ponents identified.
Identification of Chlorinated Biphenyls
The mass spectra of chlorinated biphenyls are characterized by relatively abundant molecular ion peaks. Since naturally occurring chlorine con sists of two isotopes, C1^5 and C1^7f in the ratio of 3:1, the molecular ion peak actually is a multiplet of peaks. The distribution and relative
intensities of the peaks within this multiplet allow the identification of the number of chlorine atoms per molecule. Using this information and
DSW 152710
STLCOPCB4040054
2- the measured molecular weight, the empirical formula can be determined readily. These spectra can easily be distinguished from the spectra of pesticides including chlorine containing products.
4-18-69 db
DSW 152711
STLCOPCB4040055
STLCOPCB4040056
< /&<?'? /-/?/< C ST -s-V /'scs,/?
fJU-'l 6rn$s
gc.
DSW 1 5 2 7 1 2 .0 1
STLCOPCB4040057
ri
DSW 152713
STLCOPCB4040058
Cc. frtoH tfS/6,c,
S f - 5"'/
Co^uka/' <c-
DSW 1 5 2 7 1 3 .0 1
STLCOPCB4040059
STLCOPCB4040060
IO X 10 T O T H E C E N TIM E TE R 4 6 1 5 1 2
Qh<i Q ti
Qfit
OS'S
O^'S
Ob'e
0t t
STLCOPCB4040061
/Itfoc-isO/S- /? < 0 /~oT/ik j>
DSW 152716
STLCOPCB4040062
STLCOPCB4040063
1 0 X lO T O T H C fc.N T l M t i l ICR
STLCOPCB4040064
DSW 152719
STLCOPCB4040065
QUESTIONS ON JENSEN'S MASS SPECTRAL WORK
jX. What were his GC conditions for GC/MS - supports, Temperatures, sample size, flow rate, split ratio to MS?
2. Did he program the column temperature - was column bleed a problem?
3. Did he use any unusual MS conditions?
4. Did he use stainless steel columns - if so, did he observe decomposition of some of the pesticides he studied?
' MWD, JPM
PESTICIDES ANALYSIS BY MS/GC
The determination of several pesticides, including p,p' DDT and two of its metabolic products, was attempted. Under the chromatographic conditions used to identify chlorinated biphenyls, it was found that many of these pesticides decomposed to varying degrees. The chromato grams of Dieldrin, Aldrin, Lindane, and pp' DDE contained predominantly one peak, but pp' DDT, pp' DDD, and Heptaclor gave numerous broadened peaks. The decomposition is probably due to a combination of high temperatures and the catalytic effect of the stainless steel column and transfer line. The use of a glass column and transfer line might solve this decomposition problem.
COMPARISON OF ATLAS AND LKB INSTRUMENTS
Maximum Resolution (10% valley)
'
Fastest Scan Speed (magnetic)
Mass Range (2 mu)
Sensitivity for GC Inlet (pg)
Carrier Gas Separator and Max. Temp. (C)
Mass Marker
Field Ion Source
Atlas CH 433 2800
LKB 9000 1000
14-448 in 2.2 sec. 12-400 in 1.7 sec.
2-2400 0.001
2-1400 0.01
Biemann-Watson 300
Yes
Yes
Becker-Ryhage 300
Available as Option
No
db 4-18-69
DSW 152720
STLCOPCB4040066