Document 8V0LyGOaxkdLnoX8waEvXY4ny
DISTRIBUTION Akron ALGC
ALTC
Brecksvilie
Calvert City Cleveland Henry Long Beach Louisville Niagara Falls Orange Pedricktown Port Neches Shawinigan
E. B. Katzenmeyer, Jr. R. W. Strassburg 0. G. Desrosiers
J. M. Whitney M. 0. Rider E. G. OeCaplta J. A. Nikora M. M. O'Mara A. L. Schultz-L. B. Crider J. B. Pausch 0. E. Ley A. R. Berens
E. E. Atkins J. W. Bloodworth J. A. Klupar C. McCrosky D. E. Giffin C. W. Ball R. R. Taylor
J. 0. Fannin C. Reid J. H. Beck D. T. Wright D. P. O'Keefe C. W. Wiedenfeld J. P. Gregoire
Corporate Environmental Service Project 850^-76 October 14, 1976
as- mo
MOV 1
RESEARCH NOTE
T o o z e te
Evaluations of Charcoal Badges as Monitors for Vinyl Chloride Monomer
by
John W. Born
w
BFG20734
B.F.GOODRICH Research end
Development Center
Corporate Environmental Service 8504-76, RN, 10-14-76
Evaluations of Charcoal Badges as Monitors for Vinyl Chloride Monomer by
John W. Born
SUMMARY Two activated charcoal badges have proved to be effective VCM monitors under laboratory conditions. The conditions included exposure to 1, 5, and 10 ppm con centrations of VCM in air for 6 hours at 50$ relative humidity. The two badges' calibration factors were 6.0 and 7.8 ppm per 0.10 mg of vinyl chloride monomer. The calibration curves were straight lines through measured weights of adsorbed VCM up to 0.16 mg. The 0.16 mg of VCM corresponded to a 360-minute exposure to a 10 ppm VCM concentration. It is reasonable to assume that exposing a badge for 40 hours to a i.6 ppm VCM concentration would give satisfactory results. Such week-long exposures to higher VCM concentrations may prove feasible. Further evaluations are needed to prove that.
t:
BFG20735
li
B.F. GOODRICH Research end
Development Center
Corporate Environmental Service -2- 8504-76, RN, 10-14-76
CONCLUSIONS
1. The Louisiana State University (LSU) charcoal badge served well as a VCM monitor for 360-minute exposures to 1, 5, and 10 ppm concentrations of VCM in air at 50$ relative humidity.
2. The calibration curves for the two LSU badges suggest that an LSU badge might * serve continuously for five days before analysis.
3. The LSU badge appears promising as an area VCM monitor both inside and outside B.F.Goodrich plants.
4. The LSU badge may prove to be an effective monitor for other volatile pollutants.
RECOMMENDATIONS
1. Evaluate the LSU badges as VCM monitors under BF6 plant operating conditions compared with present monitors.
2. Continue to evaluate the LSU badges in the laboratory as interest warrants; for example:
a. study the effect of humidity, b. study the effect of longer-term VCM exposures such as five days, c. study the concurrent effect of other vapors in the air on the
VCM monitoring, and d. study the LSU badges as monitors for other volatile pollutants.
3. Extend all such laboratory evaluations to the interested B.F.Goodrich plants if the badges prove to be effective.
4. Arrange the assembly of the LSU badge to permit filling with charcoal through the open face before placing the diffusion membrane.
eoozevzz
BFG20736
B.F. GOODRICH Reseerch and
Development Center
Corporate Envfronmental Sarvice -3- 8504-76, RN, 10-1W6
INTRODUCTION
With proper calibration and use, personal monitor tubes meet many air sampling requirements. Such tubes include the Bendlx Personnel Monitoring Collection Colum (PMCC) and the Mine Safety Appliances Company Organic Vapor Sampling Tube (MSA tube). However, these tubes require the use of a battery-operated air sampling pump. The pump battery requires recharging after 6 hours or less of use. The pumping rate usually varies during the sampling period. The inability to measure the flow rate through the tube continuously makes the sampled air volume somewhat uncertain. The concentration of the pollutant in the air is therefore also uncertain. There is thus a need for a safe solid-state monitor which does not require a vacuum pump. Paul M. Zakriski has obtained two such monitors from Louisiana State University (LSU).
Description of the LSU Monitor Badges
The Ethyl Corporation and LSU have developed a vinyl chloride monitor badge. The badge consists of a cylindrical case containing about one gram of activated charcoal granules. See Figure 1. The LSU badge has H_ essential parts. The shell which
B
contains the activated charcoal granules has eight parts. They are the base, the
diffusion membrane, the protective ring, the screws secure the latter five parts together
screen, and the cover ring. Three in that order. A Teflon plug closes
tx)
the fiHer hole in the base. A spring clamp attaches the badge to a person's clothing or to some area fixture.
O
$FG20,73,7
B.F. GOODRICH Research and
Development Center
Corporate environmental Service -4- 8504-76, RN, 10-14-76
Charging the Badges with VCM
The dynamic vapor generator (see Reference 1) supplied the VCM-air stream for charging the activated charcoal (Nuchar WV-H, 8x30 mesh) in each badge. We added an in-line chamber between the dilution chamber and the supply manifold to contain the badge. The badge chamber (s Pyrex glass with spherical ground-glass joints. The connections are gas-tight. The flow rate of the VCM-air stream was 4.6 1pm. The inner diameter of the chamber in the plane of the badge is 75 mm. The flow rate there converts to 0.039 miles per hour. The VCM-air stream flows around the badge and then out through the five ports of the supply manifold. We thus charged five MSA tubes during each badge charging period.
We charged Badges No. 1 and No. 2 separately for 6 hours each at three VCM concen trations. The VCM concentrations were 1, 5, and 10 ppm. We obtained the concen trations by mixing pure 50% R.H. air with 9*0, 45, and 90 ml of 508 ppm VCM to give
a 4.6 Ipm stream. We charged the five MSA tubes at the nominal flow rate of 10 ml per minute during each badge charging. One purpose for the MSA tubes was to monitor the VCM concentration in the air stream. A second purpose was to provide data for later comparisons at BFG plants.
VCM Analyses
Ail analyses involved desorbing the VCM from the charcoal with carbon disulfide. We transferred all of the charcoal from an MSA tube into a 1 cc glass vial (Supelco,
Inc. catalog no. 3*123). We added 1.0 ml of carbon disulfide and closed the vial with a gas-tight septum seal. A 7-ml vial (Pierce catalog no. 13028) with a Mininert screw cap (Pierce catalog no. 10145) was used to contain ail of the charcoal from a badge. We put 3*0 ml of carbon disulfide into the vial through.the septum seal with a syringe. The MSA tube contains a nominal 150 mg of charcoal. We filled each badge with what we considered to be a full load of charcoal (Nuchar WV-H) each time. The weights of charcoal each time (Table I) were obtained after analysis by drying it and weighing by difference.
TABLE I
Summary of Charcoal Weights of LSU Badges
Badge No.
1 1 1 2 2 2
Exposure (ppm)
1 5 10 1 5 10
Wt. of Vial Plus Charcoal (g)
14.0268 14.0201 13.8903 14.3257 14.3703 14.1377
Wt. of Vial (g)
12.8550 12.8221 13.0980 12.9606 12.9533 12.8651
Wt. of Charcoal (g)
1.1718 1.1980 0.7923 1.3651 1.4170 1.2726
The variations of charcoal weight in Table 1 resulted from several factors the 8x30 mesh size range of charcoal granules was too wide. Not all particles would go through the filler hole of the badge. Second, one cannot see far enough into the
badge to know how nearly full it is. Also, one cannot see how well-packed the gran ules are inside the badge. Sieving and trials indicated that an 18x30 mesh range would permit easier and faster filling and better packing.
a
0) N g w Ul
BFG20738
B.F. GOODRICH Reieerch
end Development Center
Corporate Environmental Service -5- 8504-76, RN, 10-14-76
Three 5-Pl aliquots of each solution of VCM from a badge were analyzed with a gas chromatograph (Hewlett-Packard Model 5711A). We used a 1/8"x36" Porapak OS column
and a flame Ionization detector. The temperature program included 2 minutes at 80C, a rise of 32#C per minute, and then 2 minutes at 160C. The results of analysis appear in Table II and Figure 2.
TABLE II
Calibration of the VCM Monitor Badges
.SU Badge Number
VCM Weight (mg)
VCM Concentration in Air (ppm)
Calibration F (ppm/mg)
1 1 1 Average:
0.021$
0.057s 0.127
1.0 5-0 10.
(47)* 86
79 T2 "4
2 2 2 Average:
0.023? 0.082$
0.15s
1.0 5.0 10.
(42)* 61
63 62 1
* This value was omitted in calculating the average for the reasons discussed below. Note that this omission results in a conservative value for VCM personnel exposures at about 1 ppm.
The best plot through the data points for Badge No. 1 in Figure 2 gives a cali bration factor of 7.8 ppm per 0.10 mg of VCM analyzed. That value agrees well with the 8.2 ppm per 0.10 mg average value in Table II. The best plot through the data points for Badge No. 2 in Figure 2 gives a calibration factor of 6.0 ppm per 0.10 mg of VCM analyzed. Again, that value agrees well with the 6.2 ppm per 0.10 mg average value in Table It. The calibration factors for Badges I and 2 at 1 ppm are about 60% of the factors at 5 and 10 ppm. These lower calibration factors suggest that the diffusion rate is slower at 1 ppm than at 5 and 10 ppm. However, the following re sults of our standardization of the VCM-air mixtures show that all three calibration factors for each badge are the same within the limits of precision.
We standardized the three VCM-air mixtures as follows. We first charged a series
of eight standard MSA tubes directly from a 508 ppm VCM-air cylinder. Second, we
calibrated the gas chromatograph with these MSA tubes. The results appear in
Table III. The calibration factors for standard MSA tubes 1 through 4 averaged
1.51 0.18 ng of VCM per area unit (under the VCM peak on the GC recorder chart).
The corresponding calibration factors for standard MSA tubes 5 through 8 averaged
1.88 0.03 ng of VCM per area unit. Then we standardized the VCM-air mixtures by
charging MSA tubes with the same VCM-air stream used to charge the badges. Table IV
gives the results. The empirical result of 0.97 ppm (tubes 26 through 30) agrees
with the nominal VCM concentration of 1.0 ppm. However, the results of 3.4 ppm
(tubes 16 through 20) and 8.3 ppm (tubes 21 through 25) differ from the nominal
values of 5.0 and 10. ppm. If we should use the empirical ppm values, the results
in Table (I would become 46, 58, and 66 ppm per mg of VCM analyzed for Badge No. 1
and 41, 41, and 52 ppm per mg for Badge No. 2. Then the average calibration factors
would be 57 7 ppm per mg for Badge No. 1 and 45 5 ppm per mg for Badge No. 2.
See Figure 3-
b.f.goodrich
BFG20739
Research and
Development Center
9 o o ze w r? :
-6-
Corporate Environmental Service 8504-76, RN, 10-14-76
Figure 2. Calibration of L5U VCM Monitor Badges
22432007
Corporate Environmental Service -7- 8504-76, RN, 10-14-76
Figure 3. Recalibration of LSU VCM Monitor Badges
8 in )z e > z
gp(j20741
Development Center
VCM Badge
No.
1 1 Average:
1 1 Average:
1. 1 1 1 Average:
2 2 Average:
2 2 Average:
2 2 Average:
Corporate Environmental Service -8- 8504-76, RN, 10-14-76
TABLE III
Calibration of the Gas Chromatograph
VCM Concn. (ppm)
MSA Tube No.
VCM
Wt. (nV)
G.C. Peak Area
(units)
1.0
Std. 7
441
1.0
Std. 8
499
230 266
G.C. Calibration Factor (Fcal)
(nq per area unit
1.92 1.88 1,90 0.02
5.0
Std. 5
205
5.0
Std. 6
272
111 147
1.85 1.85 1.85 0.00
10.
Std. 1
316
10.
Std. 2
453
10.
Std. 3
930
10.
Std. 4
455
171 312
602 378
1.85 1.45 1.54 1.20
1.51 0.18
1.0
Std. 7
441
1.0
Std. 8
499
230 266
1.92 1.88 1.90 * 0.02
5.0
Std. 7
441
5.0
Std. 8
499
230 266
1.92 1.88 1.90 1 0.02
10.
Std. 7
441
10.
Std. 8
499
230 266
1.92 1.88 1.90 0.02
r> V
22432009
BFG20742
B.F.GOODRICH Research and
Development Center
MSA Tube Number
26a 27 28 29 30 Average:
I6b 17 18 19 20 Average:
21C 22 23 24 25 Average:
Corporate Environmental Sarvlca
-9- 8504-76, RN, 10-14-76
TABLE IV Standardization of the VCM-Air Mixture
VCM Weight (mg)
0.0050 0.013 0.0092 0.0075 0.0072
Air Volume Sampled (1)
1.90 4.57 3.47 3.46 3-19
VCM Concentration (ppm)d
1.02 1.09 1.03 0.84 0.89 0.97 0.088
0.0396
0.0239 0.0345 0.0274 0.0268
3.64 2.78
4.39 3.59 3-42
4.26 3-36 3.07 2.99 3.07 3.38 0.38
0.0836 0.0918
0.0579 0.0645 0.0549
3-38 4.32 2.92 3.14 2.74
9.68
8.31 7-76 8.04 7.84
8.33 0.54
a Tubes 26-30 were exposed to a 1.0 ppm VCM-air mixture.
** Tubes 16-20 were exposed to a 5.0 ppm VCM-air mixture.
c Tubes 21-25 were exposed to a 10. ppm VCM-air mixture. A
Tha calibration of the gas chromatograph was carried out only when the VCM-alc mixture in the dynamic vapor generator was i.O ppm. The resulting calibifitlon factor evidently does not apply as well when the VCM concentrat 1$ 5*0 or 10. ppm.
OTOZSi-Z
BFG20743
B.F. GOODRICH Rciurch nd
Development Center
Corporate Environmental Service -10 8504-76, RN, 10-14-76
DISCUSSION OF RESULTS
The response of both LSU badges to six-hour exposures to VCM concentrations of 1.0, 5.0, and 10 ppm in air Is linear. The VCM concentrations in air as measured via MSA tubes were 1.0, 3.4, and 8.3 ppm. The values obtained by MSA tube thus agreed exactly with the nominal value at 1.0 ppm but differed by 47$ and 20$ at 5.0 and 10 ppm, respectively. Thus, use of the nominal values gives 47$ and 20$ higher ppm values for the corresponding VCM weights. The nominal ppm values were calculated from the certified 508 ppm VCM value and the calibrated flow rates. Since the nominal and empirical values agreed exactly at 1.0 ppm, the nominal values at 5.0 and 10 ppm are probably correct. .They represent fivefold and tenfold higher VCM concentrations (namely 45 and 90 ml per minute) in the 4.6 1pm VCM-air stream (as measured by calibrated flowmeters). In any event, using the nominal ppm values gives the more conservative calibration factors.
The precision of better than 12$ of the average within each set of standardization values (Table IV)- was good. Note that this precision is about the same as that for the G.C. calibrations (Table III).
The similarity of weights of activated charcoal in the badges was good in five of the six cases (Table I). The one really low value (0.7923 g) came from the first badge that we filled. In effect we developed the technique during that first filling. The amount of charcoal in Badge No. 1 (excluding the 0.7923 9 value) averaged 88$ as much as was in Badge No. 2.
The calibration factors for both badges were'much lower (47 and 42 ppm/mg VCM) at 1.0 ppm than at 5.0 and 10 ppm (82 and 62 average ppm/mg VCM, respectively). See Table II. This marked difference in calibration factor results from a proportionately larger weight of VCM recovered at the 1.0 ppm level than at'the 5.0 and 10 ppm levels. The VCM weights at 10 ppm are about twice the weights at 5.0 ppm for both badges. But the VCM weights at 5.0 ppm are only about three times the weights recovered at the 1.0 ppm level. This reduced recovery efficiency at the 5.0 and 10 ppm exposure levels merits further investigation.
1 1O Z S frZ Z
BFG20744
B.F.GOODRICH Research and
Development Center
-ll-
Corporate Environmental Service 8504-76, RN, 10-14-76
REFERENCES
I. J. W. Born, P. M. Zakriskl, and R. D. Hardesty, Evaluation of Bendix Flasher Tubes and MSA Tubes as VCM Monitors, Corporate Environmental Services, Project $504-76, August 24, 1976.
CO N)
BFG20745
T
B.F. GOODRICH Reraarcfi and
Development Canter