Document 8RKrwQDwwb5zdB59bzEaeqpNe

II. #3500 LABORATORY EVALUATION B. Toluene and mixture of Toluene, isobutyl Alcohol, Acetone, and Perchloroethylene. The.Technical Research Center of Finland ran controlled laboratory comparisons of the #3500 vs. standard charcoal tubes and calibrated, low flow pumps. The compounds studied were Toluene, a mixture of Toluene and Iso Butyl Alcohol, Acetone, and Perchloroethylene. Chamber concentrations were checked by Miran IA infrared and by direct GC analysis. Agreement with the reference methods are excellent except in the case where exposure time was less than the recommended fifteen minutes or deliberate overload conditions were chosen. In Over load conditions, the 3500 more closely estimated the concentration than the tube. TOLUENE TECH. RESEARCH CENTER OF FINLAND SAMPLE TIME - 4 hrs. RESULTS - PPM MIRAN, IA GC 3500 #1 #2 *3 #4 #5 #6 #7 #8 TUBE #1 TUBE #2 TEST *1 288 289 297 287 - - 283 - TEST # 2 89 86 86 87 87 85 88 88 87 89 85 85 3M 107414 TOLUENE TECH RESEARCH CENTER OF FINLAND RESULTS - PPM SHORT TERM SAMPLING Sample Time (min.) Test #1 20 Test #2 15 Test #3 10 Miran, IA GC 3500 #1 //3500 // 2 560 560 560 544 544 544 550 600 503 54 2 530 529 MIXTURE TEST TECHNICAL RESEARCH CENTER OF FINLAND MONITOR - 3M #3500 RESULTS - PPM SAMPLE TIME: TUBES 1 & 2 - 66 min. TUBES 3 & 4 - 30 min. #3500 - 220 min. Miran, IA G. C. Tube 1 Tube 2 Tube 3 Tube 4 #3500 1 2 3 4 5 6 Tube #3500 ISO BUTYL ALCOHOL 144 143 150 138 138 130 138 148 144 142 146 144 139 + 8.2 144 + 3.5 TOLUENE 180 175 164 168 169 169 166 178 176 165 180 170 168 +2.4 173 + 6.4 3M 107415 acetone TECH RESEARCH CENTER OF FINLAND RESULTS - PPM Test #1 Sample time ( min.) Miran,IA GC #1 GC #2 #3500 #1 #3500 #2 65 1070 1066 - 1043 1063 Tube #1 Tube #2 - * Deliberate Overload Condition Test #2* 240 1555 1538 1560 1338 1210 1018 1007 Test #3 240 385 380 - 373 400 377 400 PERCHLOROETHYLENE TECH RESEARCH CENTER OF FINLAND SAMPLE TIME - 4 hrs. RESULTS - PPM TEST #1 Miran,IA GC #1 GC #2 120 121 125 #3500 #1 #3500 #2 124 121 Tube 122 TEST #2 42 40 42 41 42 - 3M 107416 I. LABORATORY EVALUATION C)t. Stability of Samples using Dichloromethane, 1, 2 ' Dichloroethane, Tetrachloroethylene, Heptane. The following data shows that the #3500 is gas tight. The data relate storage time to recovery coefficients (Desorption Efficiency). This data are valid only when the compound is stable. Recovery Coefficients may drop due to decomposition, polymer ization, etc. over extended storage time. Methyl Ethyl Ketone is an example of decaying recoveries over time. Reliability of samples are not assured if the captured volatile compound can escape during storage. 3M 107418 I. #3500 LABORATORY EVALUATION . Enflurane A study at an eastern U.S. university was carried out comparing the #3500 Miran 1A and Charcoal Tube (C.T.) to a known concentration in an exposure chamber. The Miran sampled continuously with readings taken at fifteen minute intervals. Velocity scans were taken along with pressure and temperature to insure uniformity of exposure conditions. The #3500 correlated well with the tank concentration and had excellent coefficient of variance. The regression slope of #3500 versus tank concentration was 0.95. ENFLURANCE LABORATORY STUDY CORRELATION COEFFICIENTS LINEAR REGRESSION 3500 - Tank 3500 - c. T. 3500 - Miran 0.960 0.970 0.963 LINEAR REGRESSION (FIT THROUGH ZERO) 3500 - Tank 0.990 3500 - C. T. 0.993 3500 - Miran 0.990 COEFFICIENTS OF VARIATION TANK CONC. (PPM) 3500 MIRAN 5 0.08 0.10 10 0.06 0.07 20 0.16 0.03 C.T. 0.11 0.27 0.25 3M 107419 enflurane LAB STUDY RESULTS - PPM TANK 5 10 20 3500 5.6 4.8 4.6 4.6 4.6 4.6 9.2 9.7 8.2 9.2 9.7 9.7 19.4 19.4 18.1 17.6 19.1 27.5 19.1 19.4 C.T. 5.2 5.2 4.6 4.6 5.0 5.0 10.4 10.4 8.5 8.5 8.9 8.9 19.7 19.7 19.3 19.3 20.9 20.9 19.2 19.2 MIRAN 4.4 4.4 4.2 4.2 4.2 4.2 8.8 8.8 7.8 7.8 8.0 8.0 15.9 15.9 15.3 15.3 15.9 15.9 15.9 15.9 1 I | j | 1 | 3M 107420 II. FIELD EVALUATION A. Complex Mixtures A U.S. chemical company ran a field comparison between a 600 mg. charcoal tube and the #3500 in a complex mixture. Agreement is good for all cases except acetone, where it appears the #3500 samples are high. In fact, the #3500 is more closely estimating the true valve. In these conditions, it is believed that acetone is subject to displacement and is more readily swept from the tube. A diffusional system has no flow of air, only organic vapor, and suffers less from displacement losses. MIXTURES SAMPLE TIME: 6 hrs. RESULTS - PPM TUBE - 600 mg. MIBK TOLUENE N-BUTYL ACETATE XYLENE MEK ACETONE METHYLEDE CHLORIDE TRICHLOROETHYLENE PERCHLOROETHYLENE ACETONE METHYLENE CHLORIDE TRICHLOROETHYLENE PERCHLOROETHYENE ACETONE #3500 8.5 16.5 2.3 0.4 ND 264 0.5 0.6 0.5 66.7 11.3 0.1 0.6 35.4 Pump & Tube 4.6 13.8 2.5 0.3 ND 115 0.3 0.4 0.3 45.2 9.2 0.1 0.4 17.0 3M 107421 II. FIELD'EVALUATION D. 1,1,2-Trichloroethylene 3M Company Industrial Hygiene ran comparison personal samples in a 3M plant. Samples were run for four hours and converted to TLV-TWA. Correlation was good between the two methods. 1,1,2 TRICHLOROETHYLENE 3M PLANT SAMPLE TIME: 4 hrs. RESULTS - PPM CORRELATION COEFFICIENT = 0.991 #3500 0.8 0.7 7.3 5.1 4.2 2.2 4.0 3.4 1.0 1.4 9.8 7.6 0.4 3.0 1.6 3.1 PUMP & TUBE 0.8 0.6 8.5 5.9 3.9 2.0 4.6 4.9 0.9 1.1 11.9 9.6 0.2 3.4 1.8 2.8 3M 107424