Document 6amVg5yxgDKkvryBZdMXnN03
ABDOO140408
OXYGEN_MONITORS ABERDEEN CHEMICAL PLANT
FEBRUARY, 8 1 988
WORK BY: W. S. Fox, Jr. Contract Procesi
Engineer
ABDOO140409
OXYGEN MONITORS ABERDEEN CHEMICAL PLANT
TABLE OF CONTENTS
I. PURPOSE OF DESIGN
II. DESCRIPTION
III.
INSTRUMENT LIST
ABD00140410
OXYOEN_MONITORS ABERDEEN_CHEMICAL_PLANT
The purpose of this design is to provide continuous monitoring of ambient air oxygen concentrations. The areas to be monitored were identified in the inert gas/nitrogen system surveys. The survey report said that the following areas will be monitored:
1.- P-1 east side second floor. 2.- The inert gas generator room. 3.- Development Lab near liquid nitrogen storage room 4.- The Development Lab storage room. 5.- Old Module GC room. 6.- V-ll dryer bldg. GC room.
ABDOO140411
OXYGEN-MONITORS ABERDEEN CHEMICAL PLANT
II. DESCRIPTION
The Safety and the Mechanical Department recommend the oxygen analyzer manufactured by Teledyne Analytical Instruments. Several of these instruments have already been purchased as part of the RQIP program for this same purpose. This type of instrument uses a Micro-Fuel oxygen sensor which provides a linear output that is proportional to oxygen concentration. This sensor is an electrochemical design that at the end of its useful life is disposed of like a flashlight battery.
alarm levels are provided,
at 19.5 % oxygen* infli-ioiie
, per the recommendation of the Safety^
Department. At each entry into the monitored area 0 alarm
lights will be x>ropided. A green light will indicate a safe
condition. A
light will indicate an oxygen content of
Also in each monitored area an audible alarm will be provided.
'is The instrument used will be a TAI Model 322/13^-2,
identical to those purchased for RQIP, except that
racks only hold two instruments instead of three.
which these
is
The areas in which the oxygen sensors will be installed are as follows:
1.-
P~1 east side second floor.- Used to detect inert gas leakage when area is shut up due to cold weather.
2.- Inert gas generator room.- Used to detect leaks.
The rack with the instruments for the above two areas will be located in the P-1 Bldg, control room.
3.-
Development Lab near liquid nitrogen system.- The liquid nitrogen cylinder is stored inside the lab. Relocating the cylinder and/or routing the relief valve discharge outside is not practical and in any case would not eliminate the need for the monitor. Release of nitrogen gas in the past has already occurred.
ABD00140412
4.-
Development Lab storage room.- This room has been
provided with a carbon dioxide fog blanket fire
protection system.
It also contains a compressed
nitrogen system.
Both system can be a source of
inert gases.
The rack with the instruments for the above two areas will be located in the Development Laboratory.
5. - Old Module GC room.- Nitrogen is used by the GC.
6. - V-ll dryer bldg. GC room.- Nitrogen is used by the GC.
The rack with the instruments for the above two areas will be located in the Vinyl Control Room.
ABD00140413
OXYGEN.MONITORS ABERDEEN CHEMICAL PLANT
III. INSTRUMENT LIST
_ITEM_ AN-101 AN-102 AN-103
LOCATION P-1 Bldg. Control Room Development Laboratory Vinyl Control Room