UTILITIES pH METHOD FOR O
E ft 920 METER
NOTE!!! WHEN pH METER IS NOT IN U$6 PLOCE IN STftNDBY BY PRESSING DOWN THE ELECTROi)0 BUTTON ftND HOLDING FOR
(4) SECONDS <TWO DEC I MOL TOINTS DISPLftYED MEftNS METER
IS IN STftNDBY) PRESSING ANV KEY RETURNS TO OPERATION.
ftLSO STORE ELECTRODE IN
17.0 BUFFER WHEN NOT IN USE.
I
MANUAL CALIBRATION WITH fiTC PROBE
2 BUFFERS
...1 SELECT pH MODE. ftLSO PROPER E^EC ERODE SET BY STATUS BUTTON.
2 SELECT STD 1 BY PRESSING FUNCTljS BUTTON 3 RINSE ELECTRODES WITH DISTILLED1WATER SHAKE OFF EXCESS WATER.
4. PLACE ELECTRODES AND ATC PROBE
pH BUFFER. WHEN READY
ILUMINATES, ADJUST DISPLAY TO BUFFER VALUE WITH ^ OR V KEY.
PRESS ENTER. THE METER STORES TH^ VALUE AND MOVES TO STD 2.
.5. RINSE ELECTRODES AND ATC PROBE AND PLACE IN SECOND BUFFER.
6 WHEN READY ILUMINATES ADJUST THEilDISPLAY TO THE CORRECT BUFFER VALUE AND PRESS ENTER. TH& METER STORES THE VALUE .
AND AUTOMATICALLY MOVES TO SAMPti$ MODE.
7. RINSE ELECTRODES AND ATC PROBE MD PLACE IN A STIRRED SAMPLE SOLUTION. WHEN READY ILUMINATES{i|RECORD pH READING.
NOTE!!!
USE MAGNETIC STIRRER Td S- R ALL SAMPLES AND BUFFERS WHEN CALIBRATING AND MEAsW ING' SAMPLE pH'S. RETURN TO STANDBY WHEN NOT IN USE.