I To: OSHA File
Interoffice Communication
From: Dote:
Subject:
Randy Istre April 12,1990
Detected Areas in Blank Wafers
VISTA
In determining the 95% confidence limit for EDC, several added steps were taken because of detectedi areas appearing in blank Organic Vapor Monitor Wafers which have the same retention time as EDC. Ten blank monitor wafers were tested. An average wafer blank area of 1360 was determined for EDC. This determination was ihade by
measuring peak heights of other peaks ( in millimeters) and dividing these lengths into their own areas. The average of these, (82) was multiplied by the peak height for EDC in each monitor blank chromatograph to get ten corrected areas for EDC. The average of these is 1360.
Randy Istre Lab Analyst
Ik cc:
PLF
VVV 000017857
r To: OSHA File
IntwcHka Communication
From: Date:
Subject:
Randy istre
April 12,1990
Detected Areas in Blank Wafers
VISTA
In determining the 95% confidence limit for EDC, several added steps were taken because of detected areas appearing in blank Organic Vapor Monitor Wafers which have the same retention time as EDC. Ten blank monitor wafers were tested. An average wafer blank area of 1360
was determined for EDC. This determination was made by measuring peak heights of other peaks ( in millimeters) and dividing these lengths into their own areas. The average of these, (82) was multiplied by the peak height for EDC in each monitor blank chromatograph to get ten corrected areas for EDC. The average of these is 1360.