Document 3J1zBYbBgXDzzkRv20YbnX9r3

Physical Chemistry Method No. 71-11 Job No. 1348522 GAS CHROMATOGRAPHIC DETERMINATION OF PENTA- AND HEXACHLOROBIPHENYLS IN AROCLOR 1242 AND MCS 1016 ____ PURPOSE OF ANALYSIS This is a rapid, simplified method to be used to determine the level of high boilers (mainly pentachlorobiphenyls) in MCS 1016 obtained from the fractionation of Aroclor 1142. It is not intended for general analysis of chlorinated biphenyl mixtures. Physical Chemistry Method No. 70-20 should be used for that purpose. CHROMATOGRAPHIC CONDITIONS Instrument: Perkin-Elmer 800, Hewlett-Packard 7624A or Equivalent Type Detector: Flame Ionization Column: 50' X 0.02" I.D. FFAP S.C.O.T. (Support-Coated Open Tubular) Capillary (Perkin-Elmer Part No. 008-0804) Column Temperature: 220C Detector Block Temperature: 300C Injection Port Temperature: 320C Sample Volume Injected: 0.005 ml Split 25 to 1 Flow Rates Column Helium: 8 ml/min (7 PSIG) Inlet Splitter: 210 ml/min Auxiliary Helium: 50 ml/min Hydrogen: 50 ml/min Air: 400 ml/min RETENTION TIMES (UNCORRECTED) OF COMPONENTS SEPARATED (CHROMATOGRAMS ATTACHED)' Biphenyl, Mono-, Di-, Tri- and Most Tetra-Chlorobiphenyls (1.0 to 5.0 min); High Boilers (Mostly Penta- and Hexa-Chlorobiphenyls)(5.5 to 11.2 min). CALCULATION TECHNIQUE Peak area normalization. The lack of pure knowns prevents the accurate determination of response factors (area correction factors) for the individual components. Therefore, the specification for MCS 1016 is simply based on area percent of total high boilers. PRECISION Reference samples* of MCS 1016 (Drum 2) and Aroclor 1242 (Lot KK 491) were analyzed in duplicate on the two instruments listed using both the Disc chart integrator and an electronic integrator (Infotronics CRS-100 on the Perkin-Elmer 800 chromatograph and Hewlett-Packard 3370A on the Hewlett-Packard 7624A chromatograph). The same column was used in both instruments. Results reported as Area % High Boilers: 0194709 MCS 1016 (Drum 2): Chromatograph Perkin-Elmer 800 Hewlett-Packard 7624A Disc Chart 0.30, 0.27 0.30, 0.30 Aroclor 1242 (Lot KK 491) Perkin Elmer 800 Hewlett-Packard 7624A 5.88, 5.78 6.14, 6.32 Method No. 71-11 Page 2 Method of Integration Infotronics CRS-100 Hewlett-Packard 3370A 0.23, 0.21 0.27, 0.28 6.22, 5.94 6.63, 6.50 `Portions of these samples are available to those Monsanto laboratories that are setting up this method for control of MCS 1016 manufacture. db Monsanto Company Organic Chemicals Division Applied Sciences Section St. Louis, Missouri 5/71 - G. M. Gasser, E. M. Emery 0194709.01 TOWOLDMON0057201 FVfcin-d vn&r ^60 oxql,l\0 TOWOLDMONOQ57202 l-f e.w/<rff' Rxcko/><1 7 ^ TOWOLDMONOQ57203 ~ E lv*e<* 60 \r =r TOWOLDMONOQ57204 Hewlett- - TacIaaJ 7CZH/\ A O 3" V * TOWOLDMONOQ57205