Physical Chemistry Method No. 71-11 Job No. 1348522
GAS CHROMATOGRAPHIC DETERMINATION OF PENTA- AND HEXACHLOROBIPHENYLS IN AROCLOR 1242 AND MCS 1016 ____
PURPOSE OF ANALYSIS
This is a rapid, simplified method to be used to determine the level of high boilers (mainly pentachlorobiphenyls) in MCS 1016 obtained
from the fractionation of Aroclor 1142. It is not intended for general analysis of chlorinated biphenyl mixtures. Physical Chemistry Method No. 70-20 should be used for that purpose.
CHROMATOGRAPHIC CONDITIONS
Instrument: Perkin-Elmer 800, Hewlett-Packard 7624A or Equivalent
Type Detector: Flame Ionization
Column: 50' X 0.02" I.D. FFAP S.C.O.T. (Support-Coated Open Tubular) Capillary (Perkin-Elmer Part No. 008-0804)
Column Temperature: 220C
Detector Block Temperature: 300C Injection Port Temperature: 320C
Sample Volume Injected: 0.005 ml Split 25 to 1
Flow Rates
Column Helium: 8 ml/min (7 PSIG) Inlet Splitter: 210 ml/min Auxiliary Helium: 50 ml/min Hydrogen: 50 ml/min Air: 400 ml/min
RETENTION TIMES (UNCORRECTED) OF COMPONENTS SEPARATED (CHROMATOGRAMS ATTACHED)'
Biphenyl, Mono-, Di-, Tri- and Most Tetra-Chlorobiphenyls (1.0 to 5.0 min); High Boilers (Mostly Penta- and Hexa-Chlorobiphenyls)(5.5 to 11.2 min).
CALCULATION TECHNIQUE
Peak area normalization. The lack of pure knowns prevents the accurate determination of response factors (area correction factors) for the individual components. Therefore, the specification for MCS 1016 is simply based on area percent of total high boilers.
PRECISION
Reference samples* of MCS 1016 (Drum 2) and Aroclor 1242 (Lot KK 491) were analyzed in duplicate on the two instruments listed using both the Disc chart integrator and an electronic integrator (Infotronics CRS-100
on the Perkin-Elmer 800 chromatograph and Hewlett-Packard 3370A on the Hewlett-Packard 7624A chromatograph). The same column was used in both instruments. Results reported as Area % High Boilers:
0194709
MCS 1016 (Drum 2):
Chromatograph
Perkin-Elmer 800 Hewlett-Packard 7624A
Disc Chart
0.30, 0.27 0.30, 0.30
Aroclor 1242 (Lot KK 491)
Perkin Elmer 800 Hewlett-Packard 7624A
5.88, 5.78 6.14, 6.32
Method No. 71-11 Page 2
Method of Integration
Infotronics CRS-100
Hewlett-Packard 3370A
0.23, 0.21
0.27, 0.28
6.22, 5.94
6.63, 6.50
`Portions of these samples are available to those Monsanto laboratories that are setting up this method for control of MCS 1016 manufacture.
db Monsanto Company Organic Chemicals Division Applied Sciences Section St. Louis, Missouri 5/71 - G. M. Gasser, E. M. Emery
0194709.01
TOWOLDMON0057201
FVfcin-d vn&r ^60
oxql,l\0 TOWOLDMONOQ57202
l-f e.w/<rff' Rxcko/><1 7 ^
TOWOLDMONOQ57203
~ E lv*e<* 60
\r =r
TOWOLDMONOQ57204
Hewlett- - TacIaaJ 7CZH/\
A
O
3"
V *
TOWOLDMONOQ57205