( 3M Organic Vapor Monitor Sampling Guide For Organic Vapor Monitor #3500/3510 Organic Vapor Monitor With Back-Up Section #3520 ( February 1990 ( 70-0701-1274-6 (39.35) R2 CFD 260A. 282A 3M Occupational Health and Environmental Safety Division 3M Center Bldg. 220-3E-04 St. Paul, Minnesota 55144-1000 3H 005233 Lltho In U.S.A. Use f the Sampling Guide Tabl s: The following table summarizes OSHA standards, 3M monitor sampling information, and recommended sampling procedures for a variety of organic compounds for which the 3M Organic Vapor Monitors can be used to accurately determine the environmental exposures. The table is not exhaustive and will be updated periodically. A. OSHA Standards The OSHA TWA-PELs (Time Weighted Average), given as workshift time weighted averages, are taken from the Federal Register as found in 29 CRF 1910.1000 as of 19 Januaiy 1989 and are summarized in the "NIOSH/OSHA Pocket Guide to Chemical Hazards." Also included in parentheses are the current ACGIH (American Conference of Governmental Industrial Hygienist) values in cases where they differ from OSHA TWAs. These values are subject to change, and appropriate publications should be consulted for the most current information. B. Monitor Samples Information Sampling Rate All sampling rates have an accuracy of 5%. The (*) compounds in the Sampling Guide tables have been subjected to an extensive amount of laboratory work to verify the sampling rate. The sampling rates given for the remaining compounds in this table were determined from empirical relationships as outlined in a publication on Sampling Rate Validation. The sampling rates are tabulated as cubic centimeters/minute and micrograms/ppm-hour. The publication on Sampling Rate Validation Protocol can be obtained upon request from 3M. The top section of the 3520 (containing the primary adsorbent) has the same geometric dimension as the 3M Organic Vapor Monitor 3500. Therefore, the sampling rate is the same, and also has an accuracy of 5%. Capacity The capacity of the monitor for each individual compound is a function of molecular structure, vapor pressure, environmental conditions, etc. The capacity values are tabulated in Section II Analysis Guide, and are used to determine the length of a recommended sampling period. Because of the backup section, the effective sampling capacity of the Organic Vapor Monitor 3520 is four times greater than the values listed for the Organic Vapor Monitor 3500/3510. When sampling environments containing contaminant mixtures or environments with high relative humidity, it is difficult to accurately define the diffusional sampling capacity. Therefore, under these conditions, the weight (Ws) collected by the secondary adsorbent of the backup section can be compared with the weight (Wp) collected by the primary adsorbent to determine sample validity. The ratio WsAA/p must be equal to or less than 0.50. C. Length of Sampling Period 1. General 3500/3510 When sampling for organic contaminants, full workshift sampling periods are recommended as the most comprehensive measures of worker exposure. When sampling some organic contaminants, sampling periods shorter than a full workshift are required in order to sample within the recommended capacity of the monitor. Under these circumstances, sequential sampling with several monitors can be performed. 3520 For those compounds where the recommended length of the sampling period for the Organic Vapor Monitor 3500/3510 is less than a full workshift, the length of the sampling period can be increased by a factor of up to four when using the 3520 Organic Vapor Monitor. 3M 005234 1 2. Effect of Humidity Recommended sampling periods have been tabulated for concentration ranges from 0.1 to 0.5 times the PEL and from 0.5 to 3 times the PEL for relative humidities less than or greater than 70%. These recommended sampling periods should not be exceeded when using OVM 3500/3510. 3. Minimum Sampling Time To confirm quantitatively the presence and concentration of a contaminant in the atmosphere, most analysts must have a minimum of 10 micrograms forG.C. analysis. A sampling period of at least 15 minutes is recommended even when 10 micrograms of the contaminant could be collected in a shorter period. For a contaminant at a low concentration level, the sampling rate of micrograms/ppm-hr. should be used to verify a sampling period during which at least 10 micrograms of the contaminant would be collected. D. Short Term Exposure Limit (STEL) The ACGIH has recommended a short-term exposure limit (STEL) as a maximum concentration to which workers can be exposed for a period up to 15 minutes continuously. No more than four (4) such excursions per day are permitted, with at least 60 minutes between exposure periods, provided that the recommended ACGIH daily TLV-TWA also is not exceeded. The STEL values summarized in the following tables can be found in the "Threshold Limit Values and Biological Exposure Indices Adopted by ACGIH for 1989-1990." The monitor Is recommended for STEL sampling if, during the 15 minute sampling period, the monitor collects a sufficient quantity of the contaminant for analysis. E. Unsuitable Compounds The OVM is not recommended for the compounds listed below because of adverse or inadequate interactions with the sorbent material. This list is representative of classes of compounds not suitable for use with the OVM. Compounds not on this list or the Compound Guide should be handled by consultation with OH&ESD Technical Service at 1-800-243-4630. Compounds Ammonia Carbon Monixide Ethylene Oxide (1) Formaldehyde (2) Hydrogen Sulfide Isocyanates Methane, Ethane, Propane Methyl Alcohol (Methanol) Methyl Chloride Methyl, Dimethyl, Trimethyl Amines Methylene Chloride (3) Organic Solids Sulfur Dioxide (1) Ethylene Oxide can be monitored using 3M 3550/3551 Ethylene Oxide Monitor (2) Formaldehyde can be monitored using 3M 3720/3721 Formaldehyde Monitor. (3) Methylene Choride, see technical data bulletin #71. F. 3M Analytical Service Selected compounds listed in the OVM Sampling Guide will be analyzed by 3M using the 3M 3510 Monitor sold with a prepaid analysis. For more information, contact your 3M Sales Representative or your local 3M OH&ESD Safety Products Distributor. 2 3M 005235 Sampling Guide Compound Acetone* Acetonitrile* Acrylonitrile Allyl Alcohol Allyl Chloride n-Amyl Acetate* n-Amyl Alcohol (1-Pentanol) s-Amyl Acetate s-Amyl Alcohol (2-Pentanol) Benzene* Benzyl Chloride Bromoform Butadiene** t n-Butyl Acetate s-Butyl Acetate t-Butyl Acetate Butyl Acrylate n-Butyl Alcohol* Recommended Sampling Period (Hrs.) RH < 70% RH >70% 0.1-0.5 0.5-3 0.1-0.5 0.5-3 PEL PEL PEL PEL 1 0.5 82 82 88 88 88 88 88 88 88 88 88 0.4 NR 87 85 84 88 88 1 0.5 61 61 86 86 84 88 84 88 86 86 86 0.4 NR 63 63 63 86 84 OSHA Standard TWA-PEL ( ) ACGIH TLV (ppm) 750 40 2 2 1 100 -- 125 -- 1 1 0.5 (10) 150 200 200 10 -- C= Ceiling STEL 1000 60 10 4 2 5 200 C50 Sampling Rate (cc/mln.) 40.1 48.2 43.8 40.4 35.1 Microgram/ ppm-Hr. 5.71 4.85 5.70 5.75 6.63 26.0 30.9 27.2 31.15 35.5 7.36 6.68 8.68 6.73 6.78 27.2 8.48 29.3 18.19 42.8 5.67 31.6 9.00 28.6 8.14 29.4 8.37 28.7 9.01 34.3 5.55 s-Butyl Alcohol t-Butyl Alcohol Butyl Cellosolve (2-Butoxyethanol) Butyl Cellosolve Acetate Butyl Glycidyl Ether (BGE) p-tert-Butyltoluene* Camphor Carbon Disulfide Carbon Tetrabromide Carbon Tetrachloride* Cellosolve* (2-Ethoxyethanol) Cellosolve Acetate* (2-Ethoxyethyl Acetate) Chlorobenzene* o-Chlorostyrene o-Chlorotoluene Ch lorobromomethane * Chloroform* 1-Chloro-1-nitropropane Chloroprene (2-Chloro-l, 3-butadiene) Cumene* Cyclohexane* Cyclohexanol* Cyclohexanone* Cyclohexene* Cyclopentadiene Cyclopentane Diacetone Alcohol* o-Dichlorobenzene* p-Dichlorobenzene* 1,1-Dlchloroethane 1,2-Dichloroethylene* 1,1-Dichloro-1-nitroethane Dichloroethyl Ether Dicyclopentadiene Diethyl Ketone Diisobutyl Ketone* (DIBK) 86 87 88 88 88 88 88 84 88 88 85 88 63 63 86 88 88 86 86 62 88 88 83 84 88 88 88 8 1.5 88 88 88 88 83 88 88 83 31 0.5 NR 88 88 88 8 1.5 1 0.2 88 88 88 88 88 85 86 86 6 1.5 86 88 86 86 62 88 88 62 31 0.5 NR 86 86 86 4 1.5 1 0.2 88 86 88 88 88 100 100 25 -- 25 10 2 4 0.1 2 200 (5) 100 (5) 75 50 50 200 2 2 10 50 300 50 25 300 75 600 50 75 100 200 2 5 5 200 25 34.8 150 35.2 27.3 24.4 27.0 20 20.7 21.4 12 42.8 0.3 26.6 30.2 32.4 26.6 29.3 75 26.0 27.3 34.4 33.5 30.4 32.2 24.5 32.4 29.5 28.9 32.3 -- 39.5 -- 36.2 28.2 C50 27.8 110 27.8 33.2 35.2 28.5 10 26.1 -- 23.6 -- 32.8 24.6 6.32 6.39 7.92 9.59 8.61 7.51 7.11 7.88 21.6 11.41 7.14 8.60 8.12 8.87 8.51 10.89 9.78 9.25 7.03 8.82 6.67 7.24 6.94 6.49 6.41 6.22 7.15 10.03 10.03 8.07 8.38 10.07 9.16 7.67 6.92 8.56 * Compound has been subjected to extensive laboratory testing for sampling rate validation. **3520 only. f See Technical Data Bulletin No. 63. 3 3M 005236 Sampling Guide Compound Dimethyl Formamide(DMF) p-Dioxane Dipropyl Ketone (4-hepanone) Dipropylene Glycol Methyl Ether Divinyl Benzene Enflurane (2-Chloro-1,1,2-trifluoroethyl difluoromethyl ether) Epichlorohydrin (1-Ch loro-2,3-epoxy-propane) Ethyl Acetate* Ethyl Acrylate Ethyl Alcohol Ethyl Amyl Ketone (5-menthyl -3-heptanone) Ethyl Benzene Ethyl Bromide* Ethyl Butyl Ketone (3-Heptanone) Ethyl Ether Ethyl Formate Ethyl Mercaptan (Ethanethiol) Ethylene Chlorohydrin (2-Chloroethanol) Ethylene Dibromide* (1,2-Dlbromoethane) Ethylene Dichlorlde* (1,2-Dichloroethane) Furfural Furfuryl Alcohol Forane (1-Chloro -2,2,2, Trifluro Ethyl Dlfluro Methyl Ether) Glycidol (2,3-Epoxy-1-propanol) Halothane (2-Bromo-2-chloro,1,1,1-trifluoroethane) n-Heptane* Hexachlorobutadiene Hexachloroethane Hexachloropentadiene Hexane Isomers* n-Hexane s-Hexyl Acetate Isoamyl Acetate Isoamyl Alcohol* Isobutyl Acetate* Isobutyl Alcohol* Isooctyl Alcohol Isophorone* Isopropoxyethanol (Ethylene Glycol Isopropyl Ether) Isopropyl Acetate Isopropyl Alcohol Isopropyl Ether Isopropyl Glycidyl Ether Mesityl Oxide* Mesitylene* (Trimethyl Benzene) Methyl Acetate* Methyl Acrylate Methylal (Dimethyoxymethane) Methyl n-Amyl Ketone (2-Heptanone) Methyl Bromide** Recommended Sampling Period (Hrs.) RH < 70% RH >70% 0.1-0.5 0.5-3 0.1-0.5 0.5-3 PEL PEL PEL PEL 88 86 88 88 88 88 88 84 88 88 88 86 OSHA Standard TWA-PEL ( )ACGIH TLV (ppm) C= Celling STEL 10 25 50 100 150 10 .. (2) Sampling Rate (cc/mln.) 32.4 34.5 27.8 25.3 22.2 28.3 Microgram/ ppm-Hr. 5.8 7.45 7.79 9.19 7.09 12 81 88 88 2 29.6 6.76 71 41 400 34.5 88 86 5 25 32.2 4 0.5 4 0.5 1000 51.2 88 88 _25 26.0 88 84 100 125 27.3 1 0.2 1 0.2 200 250 36.4 88 86 50 28.0 0.3 NR 0.3 NR 400 500 36.8 3 0.5 3 0.5 100 88 88 0.5 88 86 Cl 88 86 20 88 88 12 38.8 41.1 33.9 29.6 33.2 88 88 88 88 88 86 2 34.3 10 15 32.6 28.4 7.45 7.90 5.78 8.18 7.10 9 74 7.83 6.68 7.04 6.27 6.66 13.66 8.07 8.08 6.98 12.8 88 88 88 64 2_5 (2) 82 42 400 500 88 88 0.02 88 88 1 88 88 0.01 _. 8 1.5 4 1.5 500 1000 84 84 88 84 88 84 88 84 88 84 50 50 100 100 125 150 88 86 88 88 88 88 88 88 50 50 ____ 4 25 82 62 250 82 62 400 5 0.8 4 0.8 500 88 86 50 88 86 15 88 86 25 3 0.5 3 0.5 200 87 63 10 0.3 NR 0.3 NR 1000 88 84 100 (50) 4 0.5 3 0.5 5 310 500 75 25 250 37.1 30.2 28.9 23.1 26.7 21 32.0 32.0 28.1 27.2 32.3 31.0 35.9 25.3 21.7 29.4 31.7 39.4 31.2 29.1 31.2 26.3 37.9 35.8 37.9 27.9 46.5 6.74 14.63 7.08 148 15 53 14.1 6.74 6.74 9.93 8.68 6.98 8.82 5.81 8.07 7.34 7.51 7.93 5.17 7.81 8.28 7.49 7.73 6.72 7.56 7.07 7.81 10.84 * Compound has been subjected to extensive laboratory testing for sampling rate validation. **3520 only. 4 3M 005237 Sampling Guide Compound Recommended Sampling Period (Hrs.) RH <70% RH >70% 0.1-0.5 0.5-3 0.1 -0.5 0.5-3 PEL PEL PEL PEL OSHA Standard TWA-PEL ( ) ACGIH TLV (ppm) Methyl n-Butyl Ketone* (2-Hexanone) Methyl Cellosolve* (2-Methoxyethanol) Methyl Cellosolve Acetate* (Ethylene Glycol Methyl Ether Acetate) Methyl Cyclohexane* Methyl Cyclohexanol 8 8 88 5 8 886 25 8 886 (5) 25 (5) 8 2.5 6 2 400 8 886 50 Methyl Ethyl Ketone* (2-Butanone) Methyl Formate 5-Methyl-3-heptanone (Ethyl Amyl Ketone) Methyl Iodide Methyl Isoamyl Ketone 8 3 6 2 200 0.7 NR 0.7 NR 100 8 886 25 6 1 41 8 884 2 50 Methyl Isobutyl Carbinol (Methyl Amyl Alcohol) Methyl Isobutyl Ketone* (Hexanone) Methyl Isopropyl Ketone Methyl Methacrylate Alpha Methyl Styrene* 8 8 8 8 8 8 86 8 88 8 88 8 84 8 86 25 50 200 100 50 Methylene Chloride* (Dichloromethane)** Morpholine Naphtha (VM&P)* Naphthalene Nonane* t t t t 500 (100) 8 8 88 20 8 2 6 2 300 8 8 86 10 8 5 6 3 200 n-Octane* Pentane* 2-Pentanone* (Methyl Propyl Ketone) Perchloroethylene(Tetrachloroethylene) Phenyl Ether 8 3 6 2 300 2.5 0.4 2.5 0.4 600 8 5 6 3 200 8 8 86 25 8 8 86 1 Phenyl Glycidyl Ether n-Propyl Acetate* n-Propyl Alcohol* Propylene Dlchloride* (1,2-Dichloropropane) Propylene Glycol Monomethyl Ether 8 8 86 1 8 4 8 3 200 8 8 6 4 200 8 8 64 75 8 8 8 4 100 Propylene Oxide n-Propyl Nitrate Resorcinol Stoddard Solvent Styrene* 6 342 20 8 884 25 8 8 88 10 8 8 8 6 100 8 8 86 50 1,1,2,2-Tetrachloroethane 1,1,1,2-Tetrachloro-2,2-difluoroethane 1,1,2,2-Tetrachloro-1,2-difluoroethane Tetrahydrofuran Toluene* 8 8 88 1 5 0.75 4 0.75 500 5 0.75 4 0.75 500 8 2 4 2 200 8 8 8 6 100 1,1,1-Trichloroethane 1,1,2-Trlchloroethane* Trichloroethylene* 1,1,2-Trichloro-1,2,2-trifluoroethane (Freon 113) 1,2,3-Trichloropropane 6 8 8 0.5 8 1 41 350 8 66 10 8 88 50 NR 0.5 NR 1000 8 86 10 Vinyl Acetate Vinyl Bromide Vinyl Chloride** Vinylidene Chloride Vinyl Toluene Xylene* 8 8 64 10 2 2 22 5 2 2 2 1.5 1 2 2 22 (5) 1 8 8 6 4 100 8 8 6 4 100 STEL 300 150 40 75 -- 100 30 400 15 375 750 250 250 250 110 150 _ 40 20 100 250 150 450 200 1250 20 150 Sampling Rate (ccfmln.) 29.7 36.3 Microgram/ ppm-Hr. 7.29 6.76 29.0 8.38 28.9 6.94 28.8 8.06 36.3 6.41 45.0 6.33 26.4 7.39 36.7 12.79 28.0 7.83 29.2 6.51 30.0 7.35 32.8 6.94 31.8 7.80 25.0 7.24 37.9 7.91 33.1 7.08 33.2 8.15 24.6 6.88 24.6 7.71 26.6 7.43 35.3 5.56 33.0 6.95 28.3 11.53 24.1 10.05 20.8 7.71 30.1 7.53 39.7 5.21 30.6 8.49 32.4 6.09 42.5 6.06 33.3 7.27 27.0 7.28 24.3 8.59 26.8 6.83 28.4 11.71 29.7 14.87 28.2 14.12 37.2 6.57 31.4 7.08 30.9 10.09 29.7 9.69 31.1 10.00 31.4 14.41 27.4 9.88 35.8 7.56 41.0 10.77 40.9 6.22 38.6 26.8 27.3 9.19 7.75 7.09 `Compound has been subjected to extensive laboratory testing for sampling rate validation. "3520 Only tSee Technical Data Bulletin #71. 5 3H 005238 3M Organic Vapor Monitor Sampling Guide For Organic Vapor Monitor #3500/3510 Organic Vapor Monitor With Back-Up Section #3520 May 1989 700701-1274-6 (59 5) R2 CFD 260A, 282A 3M Occupati nal Health and Environmental Safety Division 3M Center Bldg. 220-3E-04 St. Paul, Minnesota 55144-1000 3M 005239 Litho in U S A Use of the Sampling Guide Tables: The following table summarizes OSHA standards, 3M monitor sampling information, and recommended sampling procedures for a variety of organic compounds for which the 3M Organic Vapor Monitors can be used to accurately determine the environmental exposures. The table is not exhaustive and will be updated periodically. A. OSHA Standards The OSHA TWA-PELs (Time Weighted Average), given as workshift time weighted averages, are taken from the Federal Register as found in 29 CRF 1910.1000 as of 1 January 1977 and are summarized in the "NIOSH/OSHA Pocket Guide to Chemical Hazards." Also included in parentheses are the current ACGIH (American Conference of Governmental Industrial Hygienist) values in cases where they differ from OSHA TWAs. These values are subject to change, and appropriate publications should be consulted for the most current information. B. Monitor Samples Information Sampling Rate All sampling rates have an accuracy of 5%. The (*) compounds in the Sampling Guide tables have been subjected to an extensive amount of laboratory work to verify the sampling rate. The sampling rates given for the remaining compounds in this table were determined from empirical relationships as outlined in a publication on Sampling Rate Validation. The sampling rates are tabulated as cubic centimeters/minute and micrograms/ppm-hour. The publication on Sampling Rate Validation Protocol can be obtained upon request from 3M. The top section of the 3520 (containing the primary adsorbent) has the same geometric dimension as the 3M Organic Vapor Monitor 3500. Therefore, the sampling rate is the same, and also has an accuracy of 5%. Capacity The capacity of the monitor for each individual compound is a function of molecular structure, vapor pressure, environmental conditions, etc. The capacity values are tabulated in Section II Analysis Guide, and are used to determine the length of a recommended sampling period. Because of the backup section, the effective sampling capacity of the Organic Vapor Monitor 3520 is four times greater than the values listed for the Organic Vapor Monitor 3500/3510. When sampling environments containing contaminant mixtures on environments with high relative humidity, it is difficult to accurately define the diffusional sampling capacity. Therefore, under these conditions, the weight (Ws) collected by the secondary adsorbent of the backup section can be compared with the weight (Wp) collected by the primary adsorbent to determine sample validity. The ratio Ws/Wp must be equal to or less than 0.50. C. Length of Sampling Period 1. General 3500/3510 When sampling for organic contaminants, full workshift sampling periods are recommended as the most comprehensive measures of worker exposure. When sampling some organic contaminants, sampling periods shorter than a full workshift are required in order to sample within the recommended capacity of the monitor. Under these circumstances, sequential sampling with several monitors can be performed. 3520 For those compounds where the recommended length of the sampling period for the Organic Vapor Monitor 3500/3510 is less than a full workshift, the length of the sampling period can be increased by a factor of four when using the 3520 Organic Vapor Monitor. 3H 005240 i 4 2, Effect of Humidity Recommended sampling periods have been tabulated for concentration ranges from 0.1 to 0.5 times the PEL and from 0.5 to 3 times the PEL for relative humidities less than or greater than 70%. These recommended sampling periods should not be exceeded when using OVM 3500/3510. 3. Minimum Sampling Time To confirm quantitatively the presence and concentration of a contaminant in the atmosphere, most analysts must have a minimum of 10 micrograms for G.C. analysis. A sampling period of at least 15 minutes is recommended even when 10 micrograms of the contaminant could be collected in a shorter period. For a contaminant at a low concentration level, the sampling rate of micrograms/ppm*hr. should be used to verify a sampling period during which at least 10 micrograms of the contaminant would be collected. D. Short Term Exposure Limit (STEL) The ACGIH has recommended a short-term exposure limit (STEL) as a maximum concentration to which workers can be exposed for a period up to 15 minutes continuously. No more than four (4) such excursions per day are permitted, with at least 60 minutes between exposure periods, provided that the recommended ACGIH daily TLV-TWA also is not exceeded. The STEL values summarized in the following tables can be found in the "Threshold Limit Values for Chemical Substances in Workroom Air Adopted by ACGIH for 1980." The monitor is recommended for STEL sampling if, during the 15 minute sampling period, the monitor will collect a minimum of 10 micrograms of the contaminant when sampling at the STEL concentrations. E. Unsuitable Compounds The OVM is not recommended for the compounds listed below because of adverse or inadequate interactions with the sorbent material. This list is representative of classes of compounds not suitable for use with the OVM. Compounds not on this list or the Compound Guide should be handled by consultation with OH&ESD Technical Service at 1-800-243-4630. Compounds Ammonia Carbon Monixide Ethylene Oxide (1) Formaldehyde (2) Hydrogen Sulfide Isocyanates Methane, Ethane, Propane Methyl Alcohol (Methanol) Methyl Chloride Methyl, Dimethyl, Trimethyl Amines Methylene Chloride Organic Solids Sulfur Dioxide (1) Ethylene Oxide can be monitored using 3M 3550/3551 Ethylene Oxide Monitor (2) Formaldehyde can be monitored using 3M 3720/3721 Formaldehyde Monitor. F. 3M Analytical Service Selected compounds listed in the OVM Sampling Guide will be analyzed by 3M using the 3M 3510 Monitor sold with a prepaid analysis. For more information, contact your 3M Sales Representative or your local 3M OH&ESD Safety Products Distributor. 3M 005241 2 Sampling Guide Compound Acetone* Acetonitrile* Acrylonitrile Allyl Alcohol Allyl Chloride n-Amyl Acetate* s-Amyl Acetate Benzene* Benzyl Chloride Bromoform Butadiene** t n-Butyl Acetate s-Butyl Acetate t-Butyl Acetate Butyl Acrylate n-Butyl Alcohol* s-Butyl Alcohol t-Butyl Alcohol Butyl Cellosolve* (2*Butoxyethanol) Butyl Cellosolve Acetate Butyl Glycidyl Ether (BGE) p-tert-Butyltoluene* Camphor Carbon Disulfide Carbon Tetrachloride* Cellosolve* (2-Ethoxyethanol) Cellosolve Acetate* (2-Ethoxyethyl Acetate) Chlorobenzene* o-Chlorostyrene o-Chlorotoluene Chlorobromomethane* Chloroform* 1-Chloro-l-nitropropane Chloroprene (2-Chloro-1,3-butadiene) Cumene* Cyclohexane* Cyclohexanol* Cyclohexanone* Cyclohexene* Diacetone Alcohol* o-Dichlorobenzene* p-Dichlorobenzene* 1,1-Dichloroethane 1,2-Dichloroethylene* 1,1-Dichloro-1-nitroethane Dichloroethyl Ether Diisobutyl Ketone* (DIBK) Recommended Sampling Period (Hrs.) RH < 70% RH >70% 0.1-0.5 0.5-3 0.1 -0.5 0.5-3 PEL PEL PEL PEL 1 0.5 82 82 88 88 88 88 88 88 88 0,4 NR 87 85 84 88 88 1 0.5 61 61 86 86 84 84 86 86 86 0.4 NR 63 63 63 86 84 86 87 88 88 88 88 88 84 88 85 63 63 86 88 88 86 86 62 88 83 88 84 88 88 88 85 86 86 8 1.5 6 1.5 88 86 88 88 88 86 88 86 83 88 88 83 88 62 88 88 62 86 88 86 88 86 8 1.5 4 1.5 1 0.2 1 0.2 88 88 88 88 86 88 OSHA Standard TWA-PEL ( ) ACQIH TLV (ppm) 750 40 2 2 1 100 125 1 1 0.5 (10) 150 200 200 10 100 (50) 100 100 25 -- 25 10 2 4 2 200 (5) 100 (5) 75 50 50 200 2 2 10 50 300 50 25 300 50 75 100 200 2 5 25 STEL 1000 60 10 4 2 5 200 50 150 12 75 50 100 10 Sampling Rate (cc/mln.) 40.1 48.2 43.8 40.4 35.1 26.0 27.2 35.5 27.2 29.3 42.8 31.6 28.6 29.4 28.7 34.3 Microgram ppm/hr. 5.71 4.85 5.70 5.75 6.63 7.36 8.68 6.78 8.48 18.19 5.67 9.00 8.14 8.37 9.01 5.55 34.8 6.32 35.2 6.39 27.3 7,92 24.4 9.59 27.0 8.61 20.7 7.51 21.4 7.11 42.8 7.88 30.2 11.41 32.4 7.14 26.6 8.60 29.3 8.12 26.0 8.87 27.3 8.51 34.4 10.89 33.5 9.78 30.4 9.25 32.2 7.03 24.5 8.82 32.4 6.67 29.5 7.24 28.9 6.94 32.3 6.49 28.2 7.15 27.8 10.03 27.8 10.03 33.2 8.07 35.2 8.38 28.5 10.07 26.1 9.16 24.6 8.56 * Compound has been subjected to extensive laboratory testing for sampling rate validation. **3520 only. t See Technical Data Bulletin No. 63. 3 3M 005242 Sampling Guide Compound Dimethyl Formamide(DMF) p-Dioxane Dipropylene Glycol Methyl Ether Enflurane (2-Chloro-1,1,2-trifluoroethyl difluoromethyl ether) Epichlorohydrin (1-Chloro-2,3-epoxy-propane) Ethyl Acetate* Ethyl Acrylate Ethyl Alcohol Ethyl Benzene Ethyl Bromide* Ethyl Butyl Ketone (3-Heptanone) Ethyl Ether Ethyl Formate Ethylene Chlorohydrin (2-Chloroethanol) Ethylene Dibromide* (1,2-Dibromomethane) Ethylene Dichloride* (1,2-Dichloroethane) Furfural Furfuryl Alcohol Glycldol (2,3-Epoxy-1-propanol) Halothane (2-Bromo-2-chloro,1,1,1-trifluoroethane) n-Heptane* Hexachloroethane Hexane* n-Hexane s-Hexyl Acetate Isoamyl Acetate Isoamyl Alcohol* Isobutyl Acetate* Isobutyl Alcohol* Isophorone* Isopropyl Acetate Isopropyl Alcohol Isopropyl Ether Isopropyl Glycidyl Ether Mesityl Oxide* Mesitylene* (Trimethyl Benzene) Methyl Acetate* Methyl Acrylate Methylal (Dimethyoxymethane) Methyl n-Amyl Ketone (2-Heptanone) Methyl Bromide** Recommended Sampling Period (Hrs.) RH < 70% RH >70% 0.1 -0.5 0.5-3 0.1 -0.5 o.s-3 PEL PEL PEL PEL 88 86 88 88 88 84 88 86 88 88 OSHA Standard TWA-PEL ( )ACGIH TLV (ppm) 10 25 100 STEL (2) 2 Sampling Rate (cc/min.) 32.4 34.5 25.3 28.3 Microgram ppm/hr. 5,8 7.45 9.19 12.81 29.6 6.76 71 41 400 34.5 88 86 5 25 32.2 4 0.5 4 0.5 1000 51.2 88 84 100 125 27.3 1 0.2 1 0.2 200 250 36.4 88 86 50 28.0 0.3 NR 0.3 NR 400 500 36.8 3 0.5 3 0.5 100 38.8 88 86 1 33.9 88 86 20 29.6 7.45 7.90 5.78 7.10 9.74 7.83 6.68 7.04 6.66 13.66 8888 12 88 88 88 88 88 88 88 64 2 10 15 25 (2) 82 42 400 500 88 88 1 8 1.5 4 1.5 500 1000 8484 50 88 84 50 88 84 100 88 84 88 84 8886 8888 82 62 100 125 150 50 4 250 310 8262 400 500 5 0.8 4 0.8 500 8886 50 75 88 86 15 25 8886 25 3 0.5 3 0.5 200 87 63 10 0.3 NR 0.3 NR 1000 88 84 100 (50) 4 0.5 3 0.5 5 250 33.2 34.3 32.6 37.1 30.2 28.9 26.7 32.0 32.0 28.1 27.2 32.3 31.0 35.9 21.7 31.7 39.4 31.2 29.1 31.2 26.3 37.9 35.8 37.9 27.9 46.5 8.07 8.08 6.98 6.74 14.63 7.08 15.53 6.74 6.74 9.93 8.68 6.98 8.82 5.81 7.34 7.93 5.17 7.81 8.28 7.49 7.73 6.72 7.56 7.07 7.81 10.84 `Compound has been subjected to extensive laboratory testing for sampling rate validation. **3520 only. 3M 005243 4 Sampling Guide Compound Methyl n-Butyl Ketone* (2-Hexanone) Methyl Isobutyl Ketone* (Hexanone) Methyl Cellosolve* (2-Methoxyethanol) Methyl Cellosolve Acetate* (Ethylene Glycol Methyl Ether Acetate) Methyl Chloroform* Methyl Cyclohexane* Methyl Cyclohexanol Methyl Ethyl Ketone* (2-Butanone) Methyl Formate 5-Methyl-3-heptanone (Ethyl Amyl Ketone) Recommended Sampling Period (Hra.) RH < 70% RH >70% 0.1-0.5 0.5-3 0.1 -0.5 0.5-3 PEL PEL PEL PEL 8 888 8 886 8 886 8 886 6 1 41 8 2.5 6 2 8 886 8 362 0.7 NR 0.7 NR 8 886 OSHA Standard TWA-PEL ( ) ACGIH TLV (ppm) 5 50 25 (5) 25 (5) 350 400 50 200 100 25 STEL 75 450 300 150 Methyl Iodide Methyl Isobutyl Carbinol (Methyl Amyl Alcohol) Methyl Isoamyl Ketone Methyl Methacrylate Alpha Methyl Styrene* Methylene Chloride* (Dichloromethane)** Naphtha (VM&P)* Naphthalene Nonane* n-Octane* Pentane* 2-Pentanone* (Methyl Propyl Ketone) Perch loroethylene(Tetrachloroethylene) Phenyl Ether Phenyl Glycidyl Ether n-Propyl Acetate* n-Propyl Alcohol* Propylene Dichloride* (1,2-Dichloropropane) Propylene Glycol Monomethyl Ether n-Propyl Nitrate Stoddard Solvent Styrene* 1,1,2,2-Tetrachloroethane 1,1,1,2-Tetrachloro-2,2-difluoroethane 1,1,2,2-Tetrachloro-1,2-difluoroethane 6 1 41 8 886 2 25 40 8 884 50 8 8 8 4 100 8 886 50 100 t + t t 500 (100) 8 2 6 2 300 400 8 886 10 15 8 5 6 3 200 8 3 6 2 300 375 2.5 0.4 2.5 0.4 600 750 8 5 6 3 200 250 8 886 25 8 886 1 8 886 1 8 4 8 3 200 250 8 8 6 4 200 250 8 864 75 110 8 8 8 4 100 150 8 884 25 40 8 8 8 6 100 8 8 86 50 100 8 888 1 5 0.75 4 0.75 500 5 0.75 4 0.75 500 Tetrahydrofuran Toluene* 1,1,1-Trichloroethane 1,1,2-Trichloroethane* Trichloroethylene* 1,1,2-Trichloro-1,2,2-trifluoroethane 1,2,3-Trichloropropane Vinyl Acetate Vinyl Bromide Vinyl Chloride** Vinylidene Chloride Vinyl Toluene Xylene* 8 242 200 250 8 886 100 150 6 1 41 350 450 8 866 10 8 888 50 200 0.5 NR 0.5 NR 1000 1250 8 886 10 8 864 10 20 2 222 5 2 2 2 1.5 1 2 222 (5) 1 8 864 8 8 64 100 100 150 `Compound has been subjected to extensive laboratory testing for sampling rate validation. **3520 Only tSee Technical Data Bulletin #71. Sampling Rate (cc/mln.) 29.7 30.0 36.3 Mlcrogram ppm/hr. 7.29 7.35 6.76 29.0 8.38 30.9 10.09 28.9 6.94 28.8 8.06 36.3 6.41 45.0 6.33 26.4 7.39 36.7 29.2 28.0 31.8 25.0 37.9 33.2 24.6 24.6 26.6 35.3 33.0 28.3 24.1 20.8 30.1 39.7 30.6 32.4 33.3 24.3 26.8 28.4 29.7 28.2 37.2 31.4 28.9 29.7 31.1 31.4 27.4 35.8 41.0 40.9 38.6 26.8 27.3 12.79 6.51 7.83 7.80 7.24 7.91 8.15 6.88 7.71 7.43 5.56 6.95 11.53 10.05 7.71 7.53 5.21 8.49 6.09 7.27 8.59 6.83 11.71 14.87 14.12 6.57 7.08 9.46 9.69 10.00 14.41 9.88 7.56 10.77 6.22 9.19 7.75 7.09 3M 005244 5 3M Organic Vapor Monitor Sampling Guide For Organic Vapor Monitor #3500/3510 Organic Vapor Monitor With Back-Up Section #3520 ANOTHER STEP TOWARDS BETTER OCCUPATIONAL HEALTH AND SAFETY 70*0701-1274-6 (68.3) R2 CFO 280A. 282A 3MOccupati nal Health and Safety Products Division Building 220-3E, 3M Center St. Paul, Minnesota 55144-1000 3H 005245 Lltho In U.S.A. Use of the Sampling Guide Tables: The following table summarizes OSHA standards, 3M monitor sampling information, and recommended sampling procedures for a variety of organic compounds for which the 3M Organic Vapor Monitors can be used to accurately determine the environmental exposures. The table is not exhaustive and will be updated periodically. A. OSHA Standards The OSHA TWA-PELs (Time Weighted Average), given as workshift time weighted averages, are taken from the Federal Register as found in 29 CRF 1910.1000 as of 1 January 1977 and are summarized in the "NIOSH/OSHA Pocket Guide to Chemical Hazards." Also included in parentheses are the current ACGIH (American Conference of Governmental Industrial Hygienist) values in cases where they differ from OSHA TWAs. These values are subject to change, and appropriate publications should be consulted for the most current information, B. Monitor Samples Information Sampling Rate All sampling rates have an accuracy of 5%. The (*) compounds in the Sampling Guide tables have been subjected to an extensive amount of laboratory work to verify the sampling rate. The sampling rates given for the remaining compounds in this table were determined from empirical relationships as outlined in a publication on Sampling Rate Validation. The sampling rates are tabulated as cubic centimeters/minute and micrograms/ppm-hour. The publication on Sampling Rate Validation Protocol can be obtained upon request from 3M. The top section of the 3520 (containing the primary adsorbent) has the same geometric dimension as the 3M Organic Vapor Monitor 3500. Therefore, the sampling rate is the same, and also has an accuracy of 5%. Capacity The capacity of the monitor for each individual compound is a function of molecular structure, vapor pressure, environmental conditions, etc. The capacity values are tabulated in Section II Analysis Guide, and are used to determine the length of a recommended sampling period. Because of the backup section, the effective sampling capacity of the Organic Vapor Monitor 3520 is four times greater than the values listed for the Organic Vapor Monitor 3500/3510. When sampling environments containing contaminant mixtures on environments with high relative humidity, it is difficult to accurately define the diffusional sampling capacity. Therefore, under these conditions, the weight (Ws) collected by the secondary adsorbent of the backup section can be compared with the weight (Wp) collected by the primary adsorbent to determine sample validity. The ratio Ws/Wp must be equal to or less than 0.50. C. Length of Sampling Period 1. General 3500/3510 When sampling for organic contaminants, full workshift sampling periods are recommended as the most comprehensive measures of worker exposure. When sampling some organic contaminants, sampling periods shorter than a full workshift are required in order to sample within the recommended capacity of the monitor. Under these circumstances, sequential sampling with several monitors can be performed. 3520 For those compounds where the recommended length of the sampling period for the Organic Vapor Monitor 3500/3510 is less than a full workshift, the length of the sampling period can be increased by a factor of four when using the 3520 Organic Vapor Monitor. 1 3H 005246 2. Effect of Humidity Recommended sampling periods have been tabulated for concentration ranges from 0.1 to 0.5 times the PEL and from 0.5 to 3 times the PEL for relative humidities less than or greater than 70%. These recommended sampling periods should not be exceeded when using OVM 3500/3510. 3. Minimum Sampling Time To confirm quantitatively the presence and concentration of a contaminant in the atmosphere, most analysts must have a minimum of 10 micrograms for G.C. analysis. A sampling period of at least 15 minutes is recommended even when 10 micrograms of the contaminant could be collected in a shorter period. For a contaminant at a low concentration level, the sampling rate of micrograms/ppm-hr. should be used to verify a sampling period during which at least 10 micrograms of the contaminant would be collected. D. Short Term Exposure Limit (STEL) The ACGIH has recommended a short-term exposure limit (STEL) as a maximum concentration to which workers can be exposed for a period up to 15 minutes continuously. No more than four (4) such excursions per day are permitted, with at least 60 minutes between exposure periods, provided that the recommended ACGIH daily TLV-TWA also is not exceeded. The STEL values summarized in the following tables can be found in the "Threshold Limit Values for Chemical Substances in Workroom Air Adopted by ACGIH for 1980." The monitor is recommended for STEL sampling if, during the 15 minute sampling period, the monitor will collect a minimum of 10 micrograms of the contaminant when sampling at the STEL concentrations. E. Unsuitable Compounds The OVM is not recommended for the compounds listed below because of adverse or inadequate interactions with the sorbent material. This list is representative of classes of compounds not suitable for use with the OVM. Compounds not on this list or the Compound Guide should be handled by consultation with OH&SP Technical Service. Compounds Ammonia Carbon Monixide Ethylene Oxide (1) Formaldehyde (2) Hydrogen Sulfide Isocyanates Methane, Ethane, Propane Methyl Alcohol (Methanol) Methyl Chloride Methyl, Dimethyl, Trimethyl Amines Methylene Chloride Organic Solids Sulfur Dioxide (1) Ethylene Oxide can be monitored using 3M 3550/3551 Ethylene Oxide Monitor (2) Formaldehyde can be monitored using 3M 3720/3721 Formaldehyde Monitor. F. 3M Analytical Service Selected compounds listed in the OVM Sampling Guide will be analyzed by 3M using the 3M 3510 Monitor sold with a prepaid analysis. For more information, contact your 3M Sales Representative or your local 3M OH&SP Safety Products Distributor. 2 3M 005247 Compound Acetone* Acetonitrile* Acrylonitrile Allyl Alcohol Allyl Chloride n-Amyl Acetate* s-Amyl Acetate n-Amyl Alcohol* i-Amyl Alcohol* s-Amyl Alcohol Benzene* Benzyl Chloride Bromoform Butadiene** n-Butyl Acetate s-Butyl Acetate t-Butyl Acetate Butyl Acrylate Butyl Alcohol* s-Butyl Alcohol t-Butyl Alcohol Butyl Cellosolve* (2-Butoxyethanol) Butyl Glycidyl Ether (BGE) p-tert-Butyltoluene* Camphor Carbon Disulfide Carbon Tetrachloride* Cellosolve* (2-Ethoxyethanol) Cellosolve Acetate* (2-Ethoxyethyl Acetate) Chlorobenzene* o-Chlorostyrene o-Chlorotoluene Chlorobromomethane* Chloroform* 1 -Chloro-1 -nitropropane Chloroprene (2-Chloro-1,3-butadiene) Cumene* Cyclohexane* Cyclohexanol* Cyclohexanone* Cyclohexene* Diacetone Alcohol* o-Dichlorobenzene* p-Dichlorobenzene* 1,1-Dichloroethane 1,2-Dichloroethylene* 1,1-Dichloro-1-nitroethane Dichloroethyl Ether Diisobutyl Ketone* (DIBK) * * 3520 only. Sampling Guide Recommended Sampling Period (Hrs.) RH < 70% RH >70% 0.1 -0.5 0.5-3 0.1 -0.5 0.5-3 PEL PEL PEL PEL i 0.5 1 0.5 82 82 61 61 88 88 88 88 88 88 88 88 88 88 0.4 NR 87 85 84 88 88 86 86 84 84 84 84 84 86 86 86 0.4 NR 63 63 63 86 84 86 87 88 88 88 88 82 88 85 63 63 86 86 86 86 61 86 83 88 84 88 85 88 86 88 86 8 1.5 6 1.5 84 63 88 86 88 64 88 83 88 88 86 62 86 86 83 88 88 88 62 86 86 86 8 1.5 4 1.5 1 0.2 1 0.2 88 86 88 86 88 86 OSHA Standard TWA-PEL ( ) ACQIH TLV C-Ceillng (ppm) 1000 (750) 40 2 (2) 2 1 100 125 (100) 100 (100) 1 1 0.5 (10) 150 200 200 (10) 100 (50) 150 100 50 50 10 2 20 10 200 (5) 100 (5) 75 (75) (50) (50) 200 50 (10) 20 (2) 25 (10) 50 300 500 50 300 50 50 75 100 (200) 200 10 15 (5) 50 (25) 3 Sampling Rate (cc/min.) 40.1 Microgram ppm/hr. 5.71 48.2 4.85 43.8 5.70 40.4 5.75 35.1 6.63 26.0 7.36 27.2 8.68 31.4 6.78 32.3 6.98 32.3 6.98 35.5 6.78 27.2 8.48 29.3 18.19 42.8 5.67 31.6 9.00 28.6 8.14 29.4 8.37 28.7 9.01 34.3 5.55 34.8 6.32 35.2 6.39 28.2 8.15 27.0 8.61 20.7 7.51 21.4 7.11 42.8 7.88 30.2 11.41 32.4 7.14 26.6 8.60 29.3 8.12 26.0 8.87 27.3 8.51 34.4 10.89 33.5 9.78 30.4 9.25 32.2 7.03 24.5 8.82 32.4 6.67 29.5 7.24 28.9 6.94 32.3 6.49 28.2 7.15 27.8 10.03 27.8 10.03 33.2 8.07 35.2 8.38 28.5 10.07 26.1 9.16 24.6 8.56 3M 005248 Sampling Guide Compound Dimethyl Formamide (DMF) p-Dioxane Dipropylene Glycol Methyl Ether Enflurane (2-Chloro-1,1,2 trifluoroethyl difluoromethyl ether) Epichlorohydrin (1-Chloro-2,3-epoxy-propane) Ethyl Acetate* Ethyl Acrylate Ethyl Alcohol Ethyl Benzene Ethyl Bromide* Ethyl Butyl Ketone (3-Heptanone) Ethyl Ether Ethyl Formate Ethylene Chlorohydrin (2-Chloroethanol) Ethylene Dibromide* (1,2-Dibromomethane) Ethylene Dichloride* (1,2-Dichloroethane) Furfural Furfuryl Alcohol Glycidol (2,3-Epoxy-1-propanol) Halothane (2-Bromo-2-chloro,1,1,1 Heptane* trifluoroethane) Hexachloroethane Hexane* s-Hexyl Acetate Isoamyl Acetate Isoamyl Alcohol* Isobutyl Acetate* Isobutyl Alcohol* Isophorone* Isopropyl Acetate Isopropyl Alcohol Isopropyl Ether Isopropyl Glycidyl Ether Mesityl Oxide* Mesitylene* (Trimethyl Benzene) Methyl Acetate* Methyl Acrylate Methylal (Dimethyoxymethane) Methyl Amyl Ketone (2-Heptanone) Methyl Bromide Recommended Sampling Period (Hrs.) RH <70% RH >70% 0.1-0.5 0.5-3 0.1-0.5 0.5-3 PEL PEL PEL PEL 88 86 88 84 88 84 88 86 88 86 71 41 88 64 4 0.5 4 0.5 88 84 1 0.2 1 0.2 88 86 0.3 NR 0.3 NR 3 0.5 3 0.5 88 86 88 86 88 64 8888 88 86 88 86 88 64 82 42 88 88 8 1.5 4 1.5 8884 88 84 8884 88 84 8884 88 86 82 62 82 62 5 0.8 4 0.8 88 86 88 86 88 86 3 0.5 3 0.5 8763 0.3 NR 0.3 NR 88 84 4 0.5 3 0.5 OSHA Standard TWA-PEL ( ) ACGIH TLV C-Ceiling (ppm) 10 100 (25) 100 (2) 5 (2) 400 25 (5) 1000 100 200 50 400 100 5 20 50 (10) 5 (2) 50 (5) 50 (25) (2) 500 (400) 1 (10) 500 (50) 50 100 100 150 100 (50) 25 (5) 250 400 500 (250) 50 25 (35) 25 200 10 1000 100 (50) 20 (5) Sampling Rate (cc/min.) 32.4 34.5 Mlcrogram ppm/hr. 5.8 7.45 25.3 9.19 28.3 12.81 29.6 34.5 32.2 51.2 27.3 36.4 28.0 36.8 38.8 33.9 29.6 33.2 34.3 32.6 37.1 30.2 28.9 26.7 32.0 28.1 27.2 32.3 31.0 35.9 21.7 31.7 39.4 31.2 29.1 31.2 26.3 37.9 35.8 37.9 27.9 46.5 6.76 7.45 7.90 5.78 7.10 9.74 7.83 6.68 7.04 6.66 13.66 8.07 8.08 6.98 6.74 14.63 7.08 15.53 6.74 9.93 8.68 6.98 8.82 5.81 7.34 7.93 5.17 7.81 8.28 7.49 7.73 6.72 7.56 7.07 7.81 10.84 3M 005249 4 Compound Methyl Butyl Ketone* (2-Hexanone) Methyl Isobutyl Ketone* (Hexanone) Methyl Cellosolve* (2-Methoxyethanol) Methyl Cellosolve Acetate* (Ethylene Glycol Methyl Ether Acetate) Methyl Chloroform* Methyl Cyclohexane* Methyl Cyclohexano! Methyl Ethyl Ketone* (2-Butanone) Methyl Formate 5-Methyl-3-heptanone (Ethyl Amyl Ketone) Methyl Iodide Myethyl Isobutyl Carbinol (Methyl Amyl Alcohol) Methyl Isoamyl Ketone Methyl Methacrylate Alpha Methyl Styrene* Methylene Chloride* (Dichloromethane)** Naphtha (VM&P)* Naphthalene Nonane* Octane* Pentane* 2-Pentanone* (Methyl Propyl Ketone) Perchloroethylene(Tetrachloroethylene) Phenyl Ether Phenyl Glycidyl Ether n-Propyl Acetate* n-Propyl Alcohol* Propylene Dichloride* (1,2-Dichloropropane) Propylene Glycol Monomethyl Ether n-Propyl Nitrate Stoddard Solvent Styrene* 1,1,2,2-Tetrachloroethane 1,1,1,2-Tetrachloro-2,2-difluoroethane 1,1,2,2-Tetrachloro-1,2-difluoroethane **3520 Only fSee Technical Data Bulletin #71. Sampling Guid Recommended Sampling Period (Hra.) RH <70% RH >70% 0.1-0.5 0.5-3 0.1-0.5 0.5-3 PEL PEL PEL PEL 8 884 8 884 8 886 8 886 6 1 41 8 2.5 6 2 8 884 8 362 0.7 NR 0.7 NR 8 8 86 OSHA Standard TWA-PEL ( ) ACGIH TIV C-Ceillng (ppm) 100 (5) 100 (50) 25 (5) 25 (5) 350 500 (400) 100 (50) 200 100 25 6 1 41 8 886 5 (2) 25 8 884 8 884 8 884 50 100 100 (50) t t tt 500 (100) 8 2 6 2 (300) 8 886 10 8 563 (200) 8 362 500 (300) 2.5 0.4 2.5 0.4 1000 (600) 8 563 200 8 763 100 (50) 8 886 1 8 886 10 (D 8 483 8 864 8 8 64 200 200 75 8 884 (100) 8 884 25 8 463 500 (100) 8 884 100 (50) 8 886 5 (D 5 0.75 4 0.75 500 5 0.75 4 0.75 500 Sampling Rate (cc/min.) 29.7 Microgram ppmihr. 7.29 30.0 7.35 36.3 6.76 29.0 8.38 30.9 10.09 28.9 6.94 28.8 8.06 36.3 6.41 45.0 26.4 6 33 7.39 36.7 12.79 29.2 6.51 28.0 7.83 31.8 780 25.0 7.24 37.9 7.91 33.2 8.15 24.6 6.88 24.6 7.71 26.6 7.43 35.3 5.56 33.0 6.95 28.3 11.53 24.1 10 05 20.8 7.71 3b.1 7.53 39.7 5.21 30.6 8.49 32.4 6.09 33.3 7.27 24.3 8.59 26.8 6.83 28.4 11.71 29.7 14 87 28.2 14.12 5 3M 005250 Compound Tetrahydrofuran Toluene* 1,1,2-Trichloroethane* Trichloroethylene* 1,1,2-Trichloro -1,2,2-trifluoroethane 1,2,3-Trichloropropane Vinyl Acetate Vinyl Bromide Vinyl Chloride** Vinylidene Chloride Vinyl Toluene Xylene* "3520 Only Sampling Guide Recommended Sampling Period (Hre.) RH <70% RH >70% 0.1-o.s 0.5-3 0.1-0.5 0.5-3 PEL PEL PEL PEL 8 242 8 6 63 8 866 8 886 0.5 NR 0.5 NR 8 886 8 864 2 222 2 2 2 1.5 2 2 22 8 864 8 8 64 OSHA Standard TWA-PEL ( )ACQIH TLV C-Ceillng (ppm) 200 200 (100) 10 100 (50) 1000 50 (10) (5) 1 (5) (10) 100 (50) 100 Sampling Rate (cc/mfn.) 37.2 31.4 29.7 31.1 31.4 27.4 35.8 41.0 40.9 38.6 26.8 Mierogram ppm/hr. 6.57 7.08 9 69 10.00 14.41 9.88 7.56 10.77 6.22 9.19 7.75 27.3 7.09 6 3M 005251 3M Organic Vapor Monitor Sampling Guide For Organic Vapor Monitor #3500/3510 Organic Vapor Monitor With Back-Up Section #3520 December, 1982 ANOTHER STEP TOWARDS BETTER OCCUPATIONAL HEALTH AND SAFETY R35SGA (1221) R2 Occupati nal Health and Safety Products Division/3M 220-7W 3M Center St. Paul, Minnesota 55144 3H 005252 Litho in U-S A, 4* Use of the Sampling Guide Tables: The following table summarizes OSHA standards, 3M monitor sampling information and recommended sampling procedures for a variety of organic compounds for which the 3M Organic Vapor Monitors can be used to acccurately determine the environmental exposures. The table is not exhaustive and will be updated periodically. To obtain periodic updates, return the registration card contained in every box of monitors to 3M Company. A. OSHA Standards The OSHA TWA-PEL's (Time Weighted Average) given as workshift time weighted averages are taken from the Federal Register as found in 29 CRF 1910.1000 as of 1 January 1977 and are summarized in the "NIOSH/OSHA Pocket Guide to Chemical Hazards." Also included in parentheses are the current ACGIH (American Conference of Governmental Industrial Hygienist) values in cases where they differ from OSHA TWA's. These values are subject to change and appropriate publications should be consulted for the most current information. B. Monitor Samples Information * Sampling Rate All sampling rates have an accuracy of + 5%. The (*) compounds in the Sampling Guide tables have been subjected to an extensive amount of laboratory work to verify the sampling rate. The sampling rates given for the remaining compounds in this table were determined from empirical relationships as outlines in a publication on Sampling Rate Validation. The sampling rates are tabulated as cubic centimeters/minute and micrograms/ppm-hour. The publication on Sampling Rate Validation Protocol can be obtained upon request from 3M Company. The top section of the #3520 (containing the primary absorbent) has the same geometric dimension as the 3M Organic Vapor Monitor #3500. Therefore, the sampling rates are the same, and also have an accuracy of 5%. * Capacity The capacity of the monitor for each individual compound is a function of molecular structure, vapor pressure, environmental conditions, etc. The capacity values are tabulated in Section II - Analysis Guide, and are used to determine the length of a recommended sampling period. Because of the backup section, the effective sampling capacity of the Organic Vapor Monitor #3520 is four times greater than the values listed for the Organic Vapor Monitor #3500/3510. When sampling environments containing contaminant mixtures on environments with high relative humidity, it is difficult to accurately define the diffusional sampling capacity. Therefore, under these conditions, the weight (Ws) collected by the secondary absorbent of the backup section can be compared with the weight (Wp) collected by the primary absorbent to determine sample validity. The ration Ws/Wp must be equal to or less than 0.50. C. Length of Sampling Period 1. General 3500/3510 When sampling for organic contaminants, full workshift sampling periods are recommended as the most comprehensive measures of worker exposure. When sampling some organic contaminants, sampling periods shorter than a full workshift are required in order to sample within the recommended capacity of the monitor. Under these circumstances, sequential sampling with several monitors can be performed. 3520 For those compounds where the recommended length of the sampling period for the Organic Vapor Monitor 3500/3510 is less than a full workshift, the length of the sampling period can be increased by a factor of four when using the Organic Vapor Monitor #3520. Because of the increased effective capacity of the Organic Vapor Monitor #3520, sampling periods longer than a full workshift are possible. The preferred recommendation is for full workshift sampling periods. 1 3M 005253 2. Effect on Humidity Recommended sampling periods have been tabulated for concentration ranges from .1 to .5 times the PEL and from .5 to 3 times the PEL for relative humidities less than or greater than 70%. These recommended sampling periods should not be exceeded when using OVM #3500/3510. 3. Minimum Sampling Time To confirm quantitatively the presence and concentration of a contaminant in the atmosphere, most analysts must have a minimum of 10 micrograms for G.C. analysis. A sampling period of at least 15 minutes is recommended even when 10 micrograms of the contaminant could be collected in a shorter period. For a contaminant at a low concentration level, the sampling rate of micrograms/ppm-hr. should be used to verify a sampling period during which at least 10 micrograms of the contaminant would be collected. D. Short Term Exposure Limit (STEL) The ACGIH has recommended a short-term exposure limit (STEL) as a maximum concentration to which workers can be exposed for a period up to 15 minutes continuously. No more than four (4) such excursions per day are permitted, with at least 60 minutes between exposure periods, provided that the recommended ACGIH daily TLV-TWA also is not exceeded. The STEL values summarized in the following tables can be found in the "Threshold Limit Values for Chemical Substances in Workroom Air Adopted by ACGIH for 1980." The monitor is recommended for STEL sampling if, during the 15 minute sampling period, the monitor will collect a minimum of 10 micorgrams of the contaminant when sampling at the STEL concentrations. E Unsuitable Compounds The OVM is not recommended for the compounds listed below because of adverse or inadequate interactions with the sorbent material. This list is representative of classes of.compounds not suitable for use with the OVM. Compounds not on this list or the Compound Guide should be handled by consultation with OH&SP Technical Service. Compounds Ammonia Carbon Monixide (1) Ethylene Oxide (2) Formaldehyde (3) Hydrogen Sulfide Isocyanates Methane, Ethane, Propane Methyl Alcohol (Methanol) Methyl Chloride Methyl, Dimethyl, Trimethyl Amines Organic Solids Sulfur Dioxide (1) Carbon Monixide can be monitored using 3M Monitor #3400 (2) Ethylene Oxide can be monitored using 3M Ethylene Oxide Monitor #3550/3551 (3) Formaldehyde can be monitored using 3M Formaldehyde Monitor #375CV3751 F. Compounds Printed in Bold Type Ail compounds listed in bold type in the OVM Sampling Guide will be analyzed by 3M for the OVM #3510. fFor more information contact your 3M Sales Representative or your local OH&SP Safety Products Distributor. tPre-paid analysis for up to three compounds. 2 3M 005254 4 Sampling Guide COMPOUND Acetone* Acetonitrile* Acrylonitrile Allyl Alcohol Ally! Chloride n-Amyl Acetate* s-Amyl Acetate n-Amyl Alcohol* 1-Amyl Alcohol* s-Amyl Alcohol Benzene* Benzyl Chloride Bromoform Butadiene n-Butyl Acetate s-Butyl Acetate t-Butyl Acetate Butyl Acrylate Butyl Alcohol* s-Butyl Alcohol t-Butyl Alcohol Butyl Cellosolve* (2-Butoxyethanol) Butyl Glycidyl Ether (BGE) p-tert-Butyltoluene* Camphor Carbon Disulfide Carbon Tetrachloride* Cellosolve* (2-Ethoxyethanol) Cellosolve Acetate* (2-Ethoxyethyl Acetate) Chlorobenzene* o-Chlorostyrene o-Chlorotoluene Chlorobromomethane* Chloroform* 1-Chloro-1-nitropropane Chloroprene (2-Chloro-1,3-butadiene) Cumene* Cyclohexane* Cyclohexanol* Cyclohexanone* Cyclohexene* Dlacetone Alcohol* o-DIchlorobenzene* p-DIchlorobenzene* 1,1-Dichloroethane 1,2-Dichloroethylene* 1,1-Dichloro-1-nltroethane Dlchloroethyl Ether Diisobutyl Ketone* (DIBK) RECOMMENDED SAMPLING PERIOD (Hrs.) RH< 70% RH > 70% SHORTTERM EXPOSURE LIMIT (15 min.) .1.5 .5-3 1-.5 .5-3 STEL PEL PEL PEL PEL (ppm) 1 0.5 1 0.5 82 6 1 82 6 1 88 8 6 88 8 6 88 88 88 88 88 84 84 84 84 84 88 8 6 88 86 88 8 6 0.4 NR 0.4 NR 87 63 85 84 88 63 63 86 1250 60 -- 2 2 150 125 -- 25 _ -- 1250 200 250 250 -- OVM Usage Yes Yes No No Yes Yes Yes Yes Yes Yes Yes 88 8 4 -- 86 87 88 88 88 88 82 88 85 63 63 86 86 86 86 61 86 83 -- 150 Yes 150 Yes __ 20 Yes 3 No -- 20 Yes 150 Yes 88 8 4 150 Yes 88 88 85 86 -- 75 Yes 88 8 6 75 Yes 8 1.5 6 1.5 84 6 3 250 50 Yes Yes 88 88 86 64 -- _ 88 83 88 88 86 62 86 86 83 88 88 88 62 86 86 86 8 1.5 4 1.5 75 375 ___ -- _ 75 _ 110 250 Yes Yes Yes Yes Yes 1 0.2 1 0.2 88 86 88 8 6 250 10 10 Yes Yes Yes 88 8 6 -- OSHA Standard TWA-PEL ( ) ACGIH TLV C-Celllng (ppm) SAMPLING RATE (ccfmln.) Mlcrogram ppm/hr. 1000 40 2 2 1 100 125 (100) 100 (100) 10 1 0.5 1000 150 200 200 -- (10) 100 (50) 150 100 50 50 10 2 20 10 200 (100) 100 40.1 48.2 43.8 40.4 35.1 26.0 27.2 31.4 32.3 32.3 35.5 27.2 29.3 42.8 31.6 28.6 29.4 28.7 34.3 34.8 35.2 28.2 27.0 20.7 21.4 42.8 30.2 32.4 26.6 5.71 4.85 5.70 5.75 6.63 7.36 8.68 6.78 6.98 6.98 6.78 8.48 18.19 5.67 9.00 8.14 8.37 9.01 5.55 6.32 6.39 8.15 8.61 7.51 7,11 7.88 11.41 7.14 8.60 75 29.3 8.12 26.0 8.87 (50) 27.3 8.51 (50) 200 34.4 10.89 50 33.5 9.78 (10) 20 30.4 9.25 25 32.2 (10) 50 24.5 300 32.4 SO 29.5 50 28.9 7.03 8.82 6.67 7.24 6.94 300 32.3 6.49 50 28.2 7.15 50 27.8 10.03 75 27.8 10.03 100 33.2 8.07 (200) 200 35.2 8.38 10 28.5 10.07 15 26.1 9.16 (5) 50 24.6 8.56 3 3M 005255 Sampling Guide COMPOUND RECOMMENDED SAMPLING PERIOD (Hrs.) RH< 70% RH > 70% .1-.5 .5-3 1-.5 .5-3 PEL PEL PEL PEL Dimethyl Formamide (DMF) p-Dloxane Dlpropylene Glycol Methyl Ether Enflurane (2-Chloro-1,1,2 trltluoroethyl difluoromethyl ether) Epichlorohydrln (1 -Chloro-2,3-epoxy-propane) Ethyl Acetate* Ethyl Acrylate Ethyl Alcohol Ethyl Benzene Ethyl Bromide* . Ethyl Butyl Ketone (3-Heptanone) Ethyl Ether Ethyl Formate Ethylene Chlorohydrin (2-Chloroethanol) Ethylene Dlbromide* (1,2-Dlbromomethane) Ethylene Dlchloride* (1,2-Dichloroethane) Furfural Furturyl Alcohol 88 88 88 88 86 84 84 86 88 86 71 41 88 6 4 4 0.5 4 0.5 88 8 4 1 0.2 1 0.2 88 86 0.3 NR 0.3 NR 3 0.5 3 0.5 88 8 6 88 86 88 88 88 64 88 86 Glycidol (2,3-Epoxy-1 -propanol) Halothane (2-Bromo-2-chloro-1,1,1 trlfluoroethane) Heptane* 8 a 8 8 8 2 86 64 42 Hexachloroethane Hexane* 88 88 8 1.5 4 1.5 s-Hexyl Acetate Isoamyl Acetate Isoamyl Alcohol* Isobutyl Acetate* Isobutyl Alcohol* Isophorone* Isopropyl Acetate Isopropyl Alcohol 88 88 88 88 88 88 82 82 84 84 84 84 84 86 62 62 Isopropyl Ether 5 0.8 4 0.8 Isopropyl Glycldyl Ether Mesityl Oxide* Mesltylene* (Trimethyl Benzene) Methyl Acetate* Methyl Acrylate Methylal (Dimethyoxymethane) Methyl Amyl Ketone (2-Heptanone) Methyl Bromide 88 88 88 86 86 86 3 0.5 3 0.5 87 63 0.3 NR 0.3 NR 88 8 4 4 0.5 3 0.5 SHORTTERM EXPOSURE LIMIT (15 min.) STEL (ppm) OVM Usage 20 Yes -- 150 Yes -- 5 No ___ -- -- 125 Yes 250 Yes 75 Yes 500 Yes 150 Yes 15 Yes 15 Yes 10 Yes 75 Yes 500 3 125 125 125 187 75 -- 310 500 310 75 -- 35 250 ___ 1250 150 Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes OSHA Standard TWA-PEL ( ) ACGIH TLV C-Celllng (ppm) SAMPLING RATE (ec/min.) Microaram ppm/hr. 10 32.4 5.8 100 34.5 7.45 (50) 100 25.3 9.19 28.3 12.81 (2) 5 (2) 400 25 1000 100 200 50 400 100 5 20 29.6 34.5 .32.2 51.2 27.3 36.4 28.0 36.8 38.8 33.9 29.6 6.76 7.45 7.90 5.78 7.10 9.74 7.83 6.68 7.04 6.66 13.66 50 33.2 8.07 (10) 5 34.3 8.08 50 32.6 6.98 (5) 50 37.1 6.74 30.2 14.63 (2) 500 28.9 7.08 (400) 1 26.7 15.53 500 32.0 6.74 (100) 50 28.1 9.93 100 27.2 8.68 100 32.3 150 31.0 100 35.9 (50) 25 21.7 (5) 250 31.7 6.98 8.82 5.81 7.34 7.93 400 (250) 500 (250) 50 25 25 39.4 31.2 29.1 31.2 26.3 5.17 7.81 8.28 7.49 7.73 200 37.9 6.72 10 35.8 7.56 1000 37.9 7.07 100 27.9 7.81 20 46.5 10.84 (15) 4 3M 005256 Sampling Guide COMPOUND RECOMMENDED SAMPLING PERIOD (Hrs.) RH< 70% RH >70% SHORTTERM EXPOSURE LIMIT (15 min.) 1-.5 .5-3 1-.5 .5-3 STEL OVM PEL PEL PEL PEL (ppm) Usage Methyl Butyl Ketone* (2-Hexanone) Methyl Isobutyl Ketone* (Hexanone) Methyl Cellosolve* (2-Methoxyethanol) Methyl Cellosolve Acetate* (Ethylene Glycol Methyl Ether Acetate) Methyl Chloroform* 8 8 8 8 6 8 8 8 8 1 84 84 86 86 41 50 Yes 125 Yes 35 Yes 35 Yes 450 Yes Methyl Cyclohexane* Methyl Cyclohexanol 8 2.5 6 88 8 2 4 500 Yes 75 Yes Methyl Ethyl Ketone* (2-Butanone) Methyl Formate 5-Methyl-3-heptanone (Ethyl Amyl Ketone) Methyl Iodide Myethyl Isobutyl Carbinol (Methyl Amyl Alcohol) Methyl Isoamyl Ketone Methyl Methacrylate Alpha Methyl Styrene* Methylene Chloride* (Dlchloromethane) Naphtha (VM&P)* Naphthalene Nonane* Octane* Pentane* 2-Pentanone* (Methyl Propyl Ketone) Perchloroethylene (Tetrachloroethylene) Phenyl Ether Phenyl Glycidyl Ether n-Propyl Acetate* n-Propyl Alcohol* Propylene Dlchloride* (1,2-Dlchloropropane) Propylene Glycol Monomethyl Ether n-Propyl Nitrate Stoddard Solvent Styrene* 1,1,2,2-Tetrachloroethane 1,1,1,2-Tetrachloro*2,2-dlfluoroethane 1,1,2,2-Tetrachloro-1,2-difluoroethane Tetrahydrofuran Toluene* 1,1,2-Trlchloroethane* Trichloroethylene* 1,1,2-Trlchloro-l ,2,2-trifluoroethane 1,2,3-Trlchloropropane Vinyl Acetate Vinyl Bromide Vinyl Chloride* Vinylldene Chloride Vinyl Toluene Xylene* 83 6 2 0.7 NR 0.7 NR 88 8 6 300 150 -- 61 88 41 86 10 40 a8 88 88 84 84 84 1 0.2 1 0.2 150 125 -- 250 82 88 85 62 86 63 400 15 250 83 62 2.5 0.4 2.5 0.4 375 750 85 87 88 88 84 88 88 63 63 86 86 83 84 64 250 150 2 15 250 250 110 88 8 4 150 88 84 84 63 40 125 88 8 4 125 88 8 6 10 5 0.75 4 0.75 625 5 0.75 4 0.75 625 82 86 88 88 42 63 68 86 250 150 20 150 0.5 NR 0.5 NR 1250 88 88 86 64 75 20 22 22 22 88 88 22 2 1.5 22 64 64 _ -- 20 150 150 Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes No Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes OSHA Standard TWA-PEL ( ) ACGIH TLV C-Celllng (Ppm) SAMPLING RATE (cc/mln.) Mlcrogram ppm/hr. 100 29.7 100 30.0 25 36.3 25 29.0 7.29 7.35 6.76 8.38 350 30.9 10.09 500 28.9 100 28.8 (50) 200 36.3 100 45.0 25 26.4 6.94 8.06 6.41 6.33 7.39 5 25 100 100 100 500 (200) 300 10 ___ (200) 500 1000 (600) 200 100 1 10 200 200 75 (100) 25 500 (100) 100 5 500 500 200 200 10 100 (50) 1000 50 -- (10) (5) 1 (5) (10) 100 100 36.7 29.2 28.0 31.8 25.0 37.9 33.2 24.6 24.6 26.6 35.3 33.0 28.3 24.1 20.8 30.1 39.7 30.6 32.4 33.3 24.3 26.8 28.4 29.7 28.2 37.2 31.4 29.7 31.1 31.4 27.4 35.8 41.0 40.9 38.6 26.8 27.3 12.79 6.51 7.83 7.80 7.24 7.91 8.15 6.88 7.71 7.43 5.56 6.95 11.53 10.05 7.71 7.53 5.21 8.49 6.09 7.27 8.59 6.83 11.71 14.87 14.12 6.57 7.08 9.69 10.00 14.41 9.88 7.56 10.77 6.22 9.19 7.75 7.09 5 3M 005257 i 3M Organic Vapor Monitor Sampling Guide May, 1981 ANOTHER STEP TOWARDS BETTER OCCUPATIONAL HEALTH AND SAFETY R-3SSG(811.)R Occupational H alth and Safety Products Division/3M 220-7W 3M Center St. Paul, MN 55144 3H 005258 Litho in U.S.A. T 3M Organic Vapor Monitor Sampling Guide USE OF THE SAMPLING GUIDE TABLES: The following table summarizes OSHA standards, 3M monitor sampling information and recommended sampling procedures for a variety of organic compounds for which the 3M #3500 Organic Vapor Monitor can be used to accurately determine the environmental exposures. The table is not exhaustive and will be updated periodically. To obtain periodic updates, return the registration card contained in every box of monitors to 3M Company. A. OSHA Standards The OSHA TWA-PEL's (Time Weighted Average) given as workshift time weighted averages are taken from the Federal Register as found in 29 CFR 1910.1000 as of 1 January 1977 and are summarized in the "NIOSH/OSHA Pocket Guide to Chemical Hazards." Also included in parentheses are the current ACGIH (American Conference of Governmental Industrial Hygienist) values in cases where they differ from OSHA TWA's. These values are subject to change and appropriate publications should be consulted for the most current information. B. Monitor Sampling Information Sampling Rate All sampling rates have an accuracy of 5%. The (*) compounds in the Sampling Guide tables have been subjected to an extensive amount of laboratory work to verify the sampling rate. The sampling rates given for the remaining compounds in this table were determined from empirical relationships as outlined in a publication on Sampling Rate Validation. The sampling rates are tabulated as cubic centimeters/minute and micrograms/ ppm-hour. The publication on Sampling Rate Validation Protocol can be obtained upon request from 3M Company. Capacity The capacity of the monitor for each individual compound is a function of molecular structure, vapor pressure, environmental conditions, etc. The capacity values are tabulated in Section II -- Analysis Guide, and are used to determine the length of a recommended sampling period. C. Length of Sampling Period When sampling for organic contaminants, full workshift sampling periods are recommended as the most comprehensive measure of worker exposure. When sampling some organic contaminants, sampling periods shorter than a full workshift are required in order to sample within the recommended capacity of the monitor. Under these circumstances, sequential sampling with several monitors is recommended. Recommended sampling periods have been tabulated for concentration ranges from .1 to .5 times the PEL and from .5 to 3 times the PEL for relative humidities less than or greater than 70%. These recommended sampling periods should not be exceeded. To confirm quantitatively the presence and concentration of a contaminant in the atmosphere, most analysts must have a minimum of 10 micrograms for G.C. analysis. A sampling period of at least 15 minutes is recommended even when 10 micrograms of the contaminant could be collected in a shorter period. For a contaminant at a low concentra tion level, the sampling rate of micrograms/ppm-hr. should be used to verify a sampling period during which at least 10 micrograms of the contaminant would be collected. 3M 005259 1 D. Short Term Exposure Limit (STEL) The ACGIH has recommended a short-term exposure limit (STEL) as a maximum concentra tion to which workers can be exposed for a period up to 15 minutes continuously. No more than four (4) such excursions per day are permitted, with at least 60 minutes between exposure periods, provided that the recommended ACGIH daily TLV-TWA also is not exceeded. The STEL values summarized in the following tables can be found in the "Threshold Limit Values for Chemical Substances in Workroom Air Adopted by ACGIH for 1980." The monitor is recommended for STEL sampling if, during the 15 minute sampling period, the monitor will collect a minimum of 10 micrograms of the contaminant when sampling at the STEL concentrations. E. Unsuitable Compounds The OVM is not recommended for the compounds listed below because of adverse or inadequate interactions with the sorbent material. This list is representative of classes of compounds not suitable for use with the OVM. Compounds not on this list or the Compound Guide should be handled by consultation with OH&SP Technical Service. Compounds Ammonia Carbon Monoxide Ethylene Oxide Formaldehyde Hydrogen Sulfide Isocyanates Methane, Ethane, Propane Methyl Alcohol (Methanol) Methyl Chloride Methyl, Dimethyl, Trimethyl Amines Organic Solids Sulfur Dioxide * All Compounds listed in bold type in the 3500 Sampling Guide will be analyzed by the 3M Specialty Analysis Service. For more information contact your 3M Sales Representative or your local OH&SP Safety Products Distributor. 3H 005260 2 1 1 Sampling Guide COMPOUND Acetone* Acetonitrile Acrylonitrile Allyl Alcohol Allyl Chlorido n-Amyl Acetate* Amyl Acetate n-Amyl Alcohol* 1-Amyl Alcohol* a-Amyl Alcohol Benzene* Benzyl Chloride Bromoform Butadiene n-Butyl Acetate -Butyl Acetate t-Butyl Acetate Butyl Acrylate Butyl Alcohol* -Butyl Alcohol t-Butyl Alcohol Butyl Celloaolve* (2-Butoxyethanol) Butyl Glycldyl Ether (BGE) p-tert-Butyltoluene* Camphor Carbon Disulfide Carbon Tetrachloride* Celloaolve* (2-Ethoxyethanol) Celloaolve Acetate* (2-Ethoxyethyl Acetate) Chlorobenzene* o-Chlorostyrene o-Chlorotoluene Chlorobromomethane* Chloroform 1 -Chioro-i -nltropropane Chloroprene(2-Chloro-l,3-butadiene) Cumene* Cyclohexane* Cyclohexanol* Cyclohexanone* Cyclohexene* Dlacetone Alcohol* o-Dlehlorobenzene* p-Dlchlorobenzene* 1,1-Dichloroethane 1,2-Dlchloroethylene* 1,1-Dichloro-1-nltroethane Dichloroethyl Ether Dllsobutyl Ketone* RECOMMENDED SAMPLING PERIOD (Hrs.) RH< 70% RH >70% SHORTTERM EXPOSURE LIMIT (15 min.) .1.5 .5-3 1-.5 .5-3 STEL PEL PEL PEL PEL (ppm) OVM Usage 1 0.5 1 0.5 82 6 1 82 6 1 88 8 6 88 8 6 88 88 88 88 88 84 84 84 84 84 88 88 88 86 86 86 0.4 NR 0.4 NR 87 63 85 84 88 83 63 86 88 8 4 1250 60 -- 2 2 150 -- 125 -- 25 -- -- 1250 200 250 250 -- -----. Yes Yes No No Yes Yes Yes Yes Yes Yes Yes 86 87 88 88 88 88 82 88 85 63 63 86 86 86 86 61 86 83 -- 150 Yes 150 Yes -- 20 Yes 3 No -- 20 Yes 150 Yes 88 84 150 Yes 88 88 85 86 88 8 6 8 1.5 6 1.5 84 6 3 -- 75 75 250 50 Yes Yes Yes Yes 88 88 86 64 -- -- 88 83 88 88 86 62 86 86 83 88 88 88 62 86 86 86 8 1,5 4 1.5 75 375 -- -- -- 75 -- 110 250 Yes Yes Yes Yes Yes 1 0.2 1 0.2 88 86 88 86 250 10 10 Yes Yes Yes 88 86 -- OSHA Standard TWA-PEL ( ) ACGIH TLV C-Celling (ppm) SAMPLING RATE (cc/mln.) Microqram ppm/nr. 1000 40 2 2 1 100 125 (100) 100 (100) 10 1 0.5 1000 150 200 200 -- (10) 100 (50) 150 100 50 50 10 2 20 10 200 (100) 100 40.1 48.2 43.8 40.4 35.1 26.0 27.2 31.4 32.3 32.3 35.5 27.2 29.3 42.8 31.6 28.8 29.4 28.7 34.3 34.8 35.2 28.2 27.0 20.7 21.4 42.8 30.2 32.4 26.6 5.71 4.85 5.70 5.75 6.63 7.36 8.68 6.78 6.98 6.98 6.78 8.48 18.19 5.67 9.00 8.14 8.37 9.01 5.55 6.32 6.39 8.15 8.61 7.51 7.11 7.88 11.41 7.14 8.60 75 -- (50) (50) 200 50 (10) 20 25 (10) 50 300 50 50 300 50 50 75 100 (200) 200 10 15 (5) 50 29.3 26.0 27.3 34.4 33.5 30.4 32.2 24.5 32.4 29.5 28.9 32.3 28.2 27.8 27.8 33.2 35.2 28.5 26.1 24.6 8.12 8.87 8.51 10.89 9.78 9.25 7.03 8.82 6.67 7.24 6.94 6.49 7.15 10.03 10.03 8.07 8.38 10.07 9.16 8.56 3 3H 005261 Sampling Guide COMPOUND RECOMMENDED SAMPLING PERIOD (Hrs.) RH<70% RH >70% .1.5 .5-3 1-.5 .5-3 PEL PEL PEL PEL Dimethyl Formamlde p-Dioxane 88 88 86 84 Dipropylene Glycol Methyl Ether Enflurane (2-Chloro-1,1,2 trlfluoroethy1 dlfluoromethyl ether) Epichlorohydrin (1 -Chloro-2,3-epoxy-propane) 88 88 84 86 88 8 6 Ethyl Acetate* Ethyl Acrylate Ethyl Alcohol Ethyl Benzene Ethyl Bromide* 71 41 88 6 4 4 0.5 4 0.5 88 8 4 1 0.2 1 0.2 Ethyl Butyl Ketone (3-Heptanone) 88 8 6 Ethyl Ether 0.3 NR 0.3 NR Ethyl Formate 3 0.5 3 0.5 Ethylene Chlorohydrln (2-Chloroethanol) 8 8 8 6 Ethylene Dlbromide* 88 8 6 (1,2-Dlbromomethane) Ethylene Dlchlorlde* (1,2-Dlchloroethane) Furfural Furfuryl Alcohol 88 88 88 64 88 86 Glycidol (2,3-Epoxy-1-propanol) Halothane (2-Bromo-2-chloro-1,1,1 trlfiuoroethane) Heptane* 8 8 8 8 8 2 86 64 42 Hexachloroethane Hexane* 88 88 8 1.5 4 1.5 s-Hexyl Acetate Isoamyl Acetate Isoamyl Alcohol* Isobutyl Acetate* Isobutyl Alcohol* 88 88 88 88 88 84 84 84 84 84 Isophorone* 88 8 6 Isopropyl Acetate Isopropyl Alcohol 82 82 62 62 Isopropyl Ether 5 0.8 4 0.8 Isopropyl Glycldyl Ether Mesityl Oxide* Mesltylene* (Trlmethyl Benzene) Methyl Acetate* Methyl Acrylate Methylal (Dimethyoxymethane) Methyl Amyl Ketone (2-Heptanone) Methyl Bromide 88 88 88 86 86 86 3 0.5 3 0.5 87 6 3 0.3 NR 0.3 NR 88 8 4 4 0.5 3 0.5 SHORTTERM EXPOSURE LIMIT (15 min.) STEL (ppm) OVM Usage 20 Yes -- 150 Yes -- 5 No -- -- -- 125 Yes 250 Yes 75 Yes 500 Yes 150 Yes -- -- 15 Yes 15 Yes 10 Yes 75 Yes 500 3 125 *-- 125 125 187 75 -- 310 500 310 75 -- 35 250 -- 1250 150 Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes OSHA Standard TWAPEL ( ) ACGIH TLV C-Celling (ppm) SAMPLING RATE (cc/mln.) Micronram ppm/hr. 10 32.4 5.8 100 34.5 7.45 (50) 100 25.3 9.19 28.3 12.81 (2) 5 29.6 (2) 6.76 400 25 1000 100 200 34.5 32.2 51.2 27.3 36.4 7.45 7.90 5.78 7.10 9.74 50 28.0 7.83 400 36.8 6.68 100 38.8 7.04 5 33.9 6.66 20 29.6 13.66 50 33.2 8.07 (10) 5 34.3 8.08 50 32.6 6.98 (5) 50 37.1 6.74 30.2 14.63 (2) 500 28.9 7.08 (400) 1 26.7 15.53 500 32.0 6.74 (100) 50 28.1 9.93 100 27.2 8.68 100 32.3 150 31.0 100 35.9 (50) 25 21.7 (5) 250 31.7 6.98 8.82 5.81 7.34 7.93 400 (250) 500 (250) 50 25 25 39.4 31.2 29.1 31.2 26.3 5.17 7.81 8.28 7.49 7.73 200 37.9 6.72 10 35.8 7.56 1000 37.9 7.07 100 27.9 7.81 20 46.5 10.84 (15) 3M 005262 4 J a Sampling Guide COMPOUND RECOMMENDED SAMPLING PERIOD (Hrs.) RH < 70% RH >70% SHORTTERM EXPOSURE LIMIT (15 min.) -1-.5 .5-3 .1.5 .5-3 STEL OVM PEL PEL PEL PEL (ppm) Usage Methyl Butyl Ketone* (2-Hexanone) Methyl leobutyl Ketone* (Hexenone) Methyl Celloeolve* (2-Methoxyethanol) Methyl Celloeolve Aeetate* (Ethylene Glycol Methyl Ether Aeetate) Methyl Chloroform* Methyl Cyclohexane* Methyl Cyclohexanol 8 8 8 8 6 8 8 88 88 88 88 14 2.5 6 88 4 4 6 6 1 2 4 50 Yes 125 Yes 35 Yes 35 Yes 450 Yes 500 Yes 75 Yes Methyl Ethyl Ketone* (2-Butanone) Methyl Formate 5-Methyl-3-heptanone (Ethyl Amyl Ketone) 83 6 2 0.7 NR 0,7 NR 88 8 6 Methyl Iodide Methyl Isobutyl Carbinol (Methyl Amyl Alcohol) Methyl Isoamyl Ketone Methyl Methacrylate Alpha Methyl Styrene* 61 88 88 88 88 41 86 84 84 84 Methylene Chloride* (Dlchloromethane) 1 0.2 1 0.2 300 150 -- 10 40 150 125 -- 250 Yes Yes Yes Yes Yes Yes Yes Naphtha (VM&P)* Naphthalene Nonane* 8 2 6 2 400 Yes 88 86 15 Yes 8 5 6 3 250 Yes Octane* Pentane* 83 62 2.5 0.4 2.5 0.4 375 750 Yes Yes 2-Pentanone* (Methyl Propyl Ketone) 8 5 6 3 250 Yes Perehk>roethylene(TetrachloiDethylene) 8 7 6 3 150 Yes Phenyl Ether 88 8 6 2 No Phenyl Glycidyl Ether 88 8 6 15 Yes n-Propyl Acetate* 8 4 8 3 250 Yes n-Propyl Alcohol* 8 8 6 4 250 Yes Propylene Bichloride* 8 8 6 4 110 Yes (1,2-Dlchloropropane) Propylene Glycol Monomethyl Ether 8 8 8 4 150 Yes n-Propyl Nitrate Stoddard Solvent 88 8 4 40 Yes 8 4 6 3 125 Yes Styrene* 1,1,2,2-Tetrch loroethane 1,1,1,2-Tetrachloro-2,2-difluoroethane 1,1,2,2-Tetrchloro-1,2-dlfluoroethane Tetrahydrofuran Toluene* 1,1,2-Trlehloroethane* Trichloroethylene* 88 8 4 125 88 86 10 5 0.75 4 0.75 625 5 0.75 4 0.75 625 82 86 88 88 42 63 66 86 250 150 20 150 Yes Yes Yes Yes Yes Yes Yes Yes 1,1,2-Trlchloro-1,2,2-trifluoroethane 1,2,3-Trichloropropane Vinyl Acetate 0.5 NR 0.5 NR 1250 88 88 86 64 75 20 Yes Yes Yes Vinyl Bromide Vinyl Chloride* 22 22 22 2 1.5 -- -- Vlnylidene Chloride Vinyl Toluene Xylene* 22 2 2 20 Yes 8 8 6 4 150 Yes 8 8 6 4 150 Yes OSHA Standard TWA-PEL ( ) ACGIH TLV C-Ceillng (ppm) SAMPLING RATE (cc/mln.) Microgram ppm/hr. 100 29.7 100 30.0 25 36.3 25 29.0 7.29 7.35 6.76 8.38 350 30.9 10.09 500 28.9 100 28.8 (50) 200 36.3 100 45.0 25 26.4 6.94 8.06 6.41 6.33 7.39 5 25 100 100 100 500 (200) 300 10 -- (200) 500 1000 (600) 200 100 1 10 200 200 75 (100) 25 500 (100) 100 5 500 500 200 200 10 100 (50) 1000 50 -- (10) (5) 1 (5) (10) 100 100 36.7 29.2 28.0 31.8 25.0 37.9 33.2 24.6 24.6 26.6 35.3 33.0 28.3 24.1 20.8 30.1 39.7 30.6 32.4 33.3 24.3 26.8 28.4 29.7 28.2 37.2 31.4 29.7 31.1 31.4 27.4 35.8 41.0 40.9 38.6 26.8 27.3 12.79 6.51 7.83 7.80 7.24 7.91 8.15 6.88 7.71 7.43 5.56 6.95 11.53 10.05 7.71 7.53 5.21 8.49 6.09 7.27 8.59 6.83 11.71 14.87 14.12 6.57 7.08 9.69 10.00 14.41 9.88 7.56 10.77 6.22 9.19 7.75 7.09 3M 005263 5