Document 2j8L3do1nNwVo95Qgrzj1G7Ng
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN CHEMICAL PLANT JUNE 30, 1987
VAB.0001153903
7: - ^ '
LEAK DETECTION AND ELIMINATION PROGRAM
' ABERDEEN CHEMICAL PLANT
( -^
TABLE OF CONTENTS
A. GENERAL
B. VINYL CHLORIDE CONTINUOUS MONITORING SYSTEM
, C. HNU PORTABLE HYDROCARBON DETECTORS
T
1. INTRODUCTION 2. OPERATION 3. DETECTION PRINCIPLE AND THEORY
"Um .
D. TLV PORTABLE HYDROCARBON DETECTORS
1. INTRODUCTION AND DETECTION PRINCIPLE 2. OPERATION
v 'V. -C . n
E. CONTINUOUS MONITORING SYSTEM - CALIBRATION AND MAINTENANCE PROCEDURES
1. CALIBRATION 2. MAINTENANCE
F. HNU PORTABLE HYDROCARBON DETECTOR - CALIBRATION AND MAINTENANCE PROCEURES
1. HNU CALIBRATION 2. HNU MAINTENANCE
G. TLV PORTABLE HYDROCARBON DETECTOR - CALIBRATION AND MAINTENANCE PROCEDURES
1. TLV CALIBRATION 2. TLV MAINTENANCE
H. ROUTINE LEAK PATROL - DATA RECORDING AND HANDLING
I. LOCATION OF CONTINUOUS MONITORING POINTS
J. ACTION PLAN FOR LEAK ELIMINATION
K. DEFINITION OF THE FIXED POINT MONITORING SYSTEM ACTION LEVEL AND A LEAK
4
VAB.0001153904
r
LEAK DETECTION AND ELIMINATION PROGRAM ABERDEEN CHEMICAL PLANT LIST OF TABLES
TABLE 1 TABLE 2
TABLE 3 TABLE 4 TABLE 5
TABLE 6
TABLE 7
TABLE 8
TABLE 9
TABLE 10
TABLE 11 TABLE 12
TABLE 13 TABLE 14 TABLE 15 TABLE 16 TABLE 17 TABLE 18
ANALYZER SECTION, OPERATING CONTROLS AND INDICATORS MONITOR MODULE, OPERATING CONTROLS AND INDICATORS SEQUENCER MODULE, OPERATING CONTROLS AND INDICATORS AMPLIFIER MODULE, OPERATING CONTROLS AND INDICATORS COMPONENT MODULE, OPERATING CONTROLS AND INDICATORS VALVE MODULE, OPERATING CONTROLS AND INDICATORS DUAL PREAMP MODULE, OPERATING CONTROLS AND INDICATORS SPECIFICATIONS FOR MODEL PI - 101 PHOTOIONIZATION ANALYZER BRIEF DESCRIPTION OF INSTRUMENT CONTROLS AND FUNCTIONS VERIFICATION OF ELECTRONIC ZERO FOR PHOTOIONIZATION ANALYZER RELATIVE PHOTOIONIZATION SENSITIVITIES FOR VARIOUS GASES SOME DERIVATIVES OF OLEFINS RELATIVE SENSIVITIES FOR VARIOUS GASES SPECTRUM ANALYSIS TROUBLESHOOTING CHART OVEN HEATING SYSTEM TROUBLESHOOTING CHART TLV TROUBLESHOOTING PROCEDURES LOCATION OF CONTINUOUS MONITORING POINTS ACTION LEVELS FOR LEAK DETECTION REPORTING AND ELIMINATION PROGRAM
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN CHEMICAL PLANT LIST OF FIGURES
FIGURE 1 FIGURE 2
FIGURE 3 FIGURE 4 FIGURE 5
FIGURE 6
FIGURE 7
FIGURE 8
FIGURE 9
FIGURE 10 FIGURE 11 FIGURE 12
FIGURE 13
CONTROL PANEL FUNCTIONS OF PHOTOIONIZATION ANALYZER TIME RESPONSE FOR THE PHOTOIONIZATION ANALYZER TYPICAL CALIBRATION CURVE FOR PHOTOIONIZATION ANALYZER BLOCK DIAGRAM OF PHOTOIONIZATION ANALYZER ELECTRICAL BLOCK DIAGRAM OF PHOTOIONIZATION ANALYZER CALIBRATION PROCEDURE FOR PHOTOIONIZATION ANALYZER
COMPONENT PARTS OF PROBE COMPONENT PARTS OF LAMP HOUSING COMPONENT PARTS OF READOUT UNIT POWER SUPPLY PC BOARD TLV CIRCUITRY ADJUSTMENT POTENTIOMETER LOCATIONS LOCATION AND ADJUSTING R-13 TLV SNIFFER CHASSIS SUBASSEMBLY
A
VAB.0001153906
*
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN CHEMICAL PLANT
y
A
GENERAL
The emission standard for vinyl chloride requires the plant to have a formal leak detection program.
The Federal Register describes specific criteria which must be met within the program. This document describes the leak detection program. Each of the specific requirements will be met. Namely, this program describes:
The continuous VCM detection system which is used by the plant.
The portable hydrocarbon detectors which are used by the plant.
The calibration and maintenance procedures for the above equipment.
The Routine Leak Patrols and procedure when a leak is found by the portable hydrocarbon detector.
The location of the continuous monitoring points which are used for leak detection.
The criteria which will be used to identify the presence of an action level and a leak.
The action plan which will be used to eliminate a leak after it has been detected.
VAB.0001153907
4
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN CHEMICAL PLANT
B. VINYL CHLORIDE CONTINUOUS MONITORING SYSTEM
Introduction
The concentration of vinyl chloride in the atmosphere is continuously monitored by four flame ionization chromatographs, each of which is operated continuously. Each chromatograph is a Honeywell Model 1000 Process Gas Chromatograph manufactured by:
Honeywell Systems, Inc. 6625 McGrew Street
t
Houston, Texas
Each chromatograph sequentially measures, on a continuous basis, the vinyl chloride content in air samples taken from ten different locations. A total of forty (40) locations are monitored throughout the plant continuously and the concentrations at each point are recorded.
Analytical Technique
The Flame Ionization Detector used in the Honeywell 1000 Process Gas Chromatographs has a lower detection limit of 0.1 part per million.
The special-purpose elements of the FID chromatograph are a flame cell, an FID Electrometer Assembly, a Temperature Control/Inverter Assembly, and a plug-in FID Amplifier Module which mounts in the Control Section.
The flame cell consists of an ignitor wire, a hydrogen-fed burner, and an electron collector. Within the cell, hydrogen combustion occurs in an electrical field created by the voltage from the ionization inverter. Because of the carbon-free property of hydrogen, ionization occurs only when a carbon-bearing sample is injected into the flame. Ionization causes current to flow from the collector in direct proportion to the amount of the hydrocarbon elements present in the sample. The output current from the flame cell is routed through the Electrometer (Preamp) in the Analyzer, and is applied to the input of the FID Amplifier in the Control Section. The gain is made available for external recording.
The FID Electrometer Assembly is physically divided into two major subassemblies: the Power Supply-Electrometer P.C. Card Assembly, and the Electrometer P.C. Card Assembly. The electrometer power supply accepts +_ 20 volt inputs from the Control Section power supply, and produces regulated plus and minus 15 volt outputs for operation of the electrometer circuits and the inverter located on the Temperature Control/Inverter P.C. Card Assembly. The inverter output is a -100 volt ionization potential which is applied to the base (burner) of the flame cell.
VAB.0001153908