Document 1gOn54X65E3j11RwMLQ7qpgnK

Physical Chemistry Method 74-4 Job No. 1630022 GAS CHROMATOGRAPHIC DETERMINATION OF AREA PERCENT OF 4,4'-DICHLOROBIPHENYL IN AROCLOR PRODUCTS PURPOSE OF ANALYSIS To measure the 4,4'-dichlorobiphenyl content of Aroclor products. This method was developed because Physical Chemistry Method 70-20, which is used for complete isomer analysis of Aroclor products, does not always adequately resolve the 4,4*-isomer. Method 70-20 is still the best method for overall isomer resolution. It should be noted that this method only provides the area percent and not the true weight percent of 4,4'-dichlorobiphenyl. CHROMATOGRAPHIC CONDITIONS Instrument: Varian-Aerograph 1200 Type Detector: Flame Ionization / Column: 50* X 0.02" I.D. Apiezon L SCOT {support coated open tubular) Capillary (Perkin-Elmer Part No. 008-0804) Column Temperature: 220*C Flow Rates Detector Block Temperature: Helium: 6'ml/min. {10 psig) 250*C Auxiliary Gas: 30 ml/min. Injection Port Temperature: 300*C Hydrogen: 40 ml/min. Sample Volume Injected: Air: 400 ml/min. 0.0005 ml {split about 1 to 30) RETENTION TIMES (UNCORRECTED) OF COMPONENTS OF INTEREST (CHROMATOGRAM ATT'AC HEP) ' 4,4'-Dichlorobiphenyl (7.O'). CALCULATION TECHNIQUE The area of the I,4'-dichlorobiphenyl peak is multiplied by 100 ana then is divided by the sum of the areas of all the peaks in the chromatogram. PRECISION No determination of precision was made. The variability of a single G analysis at the 95% confidence limits is usually +1 to 2% relative. db Monsanto Industrial Chemicals Co. Applied Sciences St. Louis, Mo. r 7-74 - G. M. Gasser, E. M. Emery HONS 021716 Lit- C M A h lfl HONS 021717