COMMENTS ON PCB METHODS
/
Method I. Gas Chromatographic Determination of Higher Boiling Homoloaues in Aroclor 1016 and Aroclor 114 2.
1. There should be a clearer definition of how the chromatogram is divided and over what range the elution is recorded.
Recommendations:
' a) Describe the division of the chromatogram in terms of a retention index (ASTM's preferred retention parameter
E 355-68, p. 1128); eg., all components eluting with . . a -retention index greater than 2400 will be calculated
as a "High Boiler."
b) In a similar manner the range over which compounds are
measured could be described using a retention index window: e.g., all compounds eluting with retention index 100 <1.. <2400 are "Low Boilers," those with 2400 <1^ <36t)0 are "High Boilers."
.2 Although pure known materials are generally not available, Dr. Ronald G. Webb, Southeast Environmental Research Labor atory, National Environmental Research Center - Corvallis, Environmental Protection Agency, Athens, Georgia 30601 can provide samples of materials whose compositions have been determined by a combination of elemental analysis, . . -gas chromatography with a Coulson conductivity detector
and combined gas chromatography - mass spectrometry. Using these as standards, response factors can readily be determined.
.3 The method should provide information on sensitivity and minimum detectable amounts. Refer to ASTM E260-69, p. 748.
.4 For precision, five rather than two measurements should be
made.
.
Method II. Polychlorinated Biphenyls in Petrochemicals.
1. In Note 1, there appears to be some confusion of the terms "sensitivity" and "minimum detectable limit."
.2 Method Summary:
a) Is documentation available to show that compounds elute in boiling point order? It is generally accepted that the free energy of solution determines elution order in gas-liquid partition chromatography, rather than the
' boiling points. 0079782
TOWOLDMONOQ56998
b) Many hydrocarbons with functional groups show strong electron capture response.
3. It seems that 100 yl is an awfully large sample to inject on to a 3.5% loaded column. That much material would probably kill the response of a Nib electron capture cell.
.4 Retention parameters should probaoly be expressed in terms of retention indices. Refer to ASTM E 355-68. .
.5 Provide some information on precision, accuracy, and sensitivity measurements as per ASTM E 260-69. 6. We have found that PCBs are adsorbed readily on untreated
glass surfaces. Serial dilutions should not be made in - glass containers unless the glass is properly pre-treated.