The Measurement of the Exposure of Workers to Respirable Crystalline Silica
(RCS)
Peter Stacey Health and Safety Laboratory
Sheffield S3 7HQ
HEA^^^tfEIY LABORATORY
Topics
National and international pressure to reduce the occupational exposure limits for Respirable Crystalline Silica (RCS)
The work of the organisation for standardisation working group ISO/TC146/SC2/WG 7
-Air quality/workplace air/silica
HEA^^^tfEIY LABORATORY
Occupational Exposure to Respirable Crystalline Silica
Crystalline silica is abundant and used in a wide range of materials
Workers working with silica containing materials represent the largest potentially exposed groups - 1.7 million workers in the USA - 100,000 workers in the UK
HEALTH & SAFETY LABORATORY
Silica Exposure
HEALTH & SAFETY LABORATORY
Selection of current occupational exposure standards mg.m-3
County Germany Austria Netherlands Australia France Belgium Italy Ireland
Quartz 0.15 0.15 0.075 0.1 0.1 0.1 0.05 0.05
Cristobalite 0.15 0.15 0.075 0.05 0.05 0.05 0.05 0.05
HEALTH ASAFETY LABORATORY
1
03/10/2005
Legislative Position in UK
Control of Substances Hazardous to Health Regulations (COSHH) Workplace Exposure Limit (WEL)
0.3 mg.m-3 (8 hours TWA) HSE are considering reducing to
0.1 or 0.05 mg.m-3 Medical surveillance recommended at 0.075
mg/m3
HEALTH & SAFETY LABORATORY
Percentage
After 1
m
u_ o i.
Risk of Silicosis
25 o^
20 152
0.04 0.08 0.12 0.16 0.2 0.24 0.28 xposure to Air Concentration of RCS for 15 Years
-3
mg.m
HEALTH A SAFETY LABORATORY
Goals
Elimination of silicosis - Despite improvements a persistent levels remains
In Europe - it is proposed that to eliminate silicosis the occupational exposure standard should be below 0.05 mg.irr3
HEALTH a WOT LABORATORY
Silica Analysis Methods
. Gravimetric analysis for respirable dust X-ray diffraction - Direct on-filter analysis - Redeposition onto silver filters Infrared - Direct on-filter analysis - Redeposition onto filter - KBr Pellet
HEALTH A SAFETY LABORATORY
Infrared spectra
A = Amorphous C = Cristobalite K = Kaolin (clay) Q = Quartz
Source: Madsen, et.al. Appl Occ. Environ. Hyg. 10(12): _____________________ 991-1002 (1995)________ ____________________iMaumn
Methods in SA
NIOSH 7602 - Infrared KBr disc (6 laboratories)
NIOSH 7500 - X-ray diffraction - deposition (1 laboratory)
MDHS 101 - X-ray diffraction and infrared - Direct on air sample filter analysis (1 laboratory)
HEALTH A SAFETY LABORATORY
2
03/10/2005
Limits of detection for XRD
Source NIOSH 7500 NIOSH 7602 MDHS 101
Method
Limit
Re-deposition Re-deposition
5 pg 5 pg
Direct analysis 3-6 pg
HEALTH & SAFETY LABORATORY
Quality of Measurement
Primary instrument to ensure compliance with the law
To enforce the law the precision of measurement must be sufficiently reliable - To reduce the incidence of dispute -To avoid incorrect interpretation
HEALTH & SAFETY LABORATORY
Standards for Analytical Performance
EN482 - Uncertainty of occupational hygiene measurements including bias and sampling Within 30 % (0.5 - 2 the exposure limit)
NIOSH - Accuracy including sampling Within 25 %
HEALTH & SAFETY LABORATORY
Precision between laboratories in WASP
Mass Measured On-Filter
. Exposure Limit mg/m3
0.3
8- hour sample
Mg 316
4 - hour sample
________ MS 156 '
0.1
106
53 .
,
0.05
53
26
0.025
26
13
WASP Analytical Range 60 - 460 pg
HEALTH & AFETY
Predicted Between Laboratory Precision
3
03/10/2005
The Challenge
Some laboratories are not likely to achieve the requirements of EN482 - especially if interferences are present
To measure samples of 4 hours duration or less from an air concentration of 0.05 mg.irr3
HEALTH X SAFER' LABORATORY
ISO Working Group TC146/SC2/WG7 Silica
Guidance to ensure the quality of respirable crystalline silica measurements - Targeted at analyst
. Underdevelopment
.
. Guidance for the selection of methods
'
'
. - Targeted at end those that commission work
'
- May include advice about the interpretation of results .....
Not yet proposed as a new work item
.
HEALTH & SAFETY LABORATORY
Ensuring quality
Comparing the crystalline content and particle size distribution of the calibration standard against a certified material
Encouraging the measurement of a significant number of calibration standards
Applying recovery and quality control checks and correcting for differences with the expected result
Applying instrumental checks for drift, peak and resolution.
Interpretation
HEALTH & SAFETY LABORATORY
Proficiency testing
Analysis at low levels involves interpretation
- Interferences
.'
. - Peak shape
- Movement of peak position
' .
This adds to the differences of results between . laboratories
.
Some countries may demand participation in PT schemes and/or accreditation
- 'The asbestos precedent'-why not for other
disabling lung diseases?
HEALTH & SAFETY LABORATORY
Conclusion
Precision of measurement is not sufficient to reliably measure at 0.05 mg.irr3
A multi disciplined approach is needed to help eliminate silicosis
HEA^^^tfETY LABORATORY
4
Want to help improve silica measurement ?
Contact your standards organisation and encourage their participation in our ISO working group - TC146/SC2/WG7 - Air Quality\Workplace Air\Silica
HEALTH & SAFETY LABORATORY