Document 0Jw6z3jBQz9z0jmwGNwnK51JO

Monsanto Chemical Company W. C, KRUMMRICH PLANT LABORATORY STANDARD ANALYTICAL METHOD Vnoouci ` ` ***"*..... ~ r'" Pyranol/Therminol (returned) ItlT Low Boilers by GLC _ THIS FORM [ J SUPERSEDES METHOD 1- , [ | REVISES INSTRUCTIONS DATE ISSUEO 11-19-70 Wkfr 1EN BY JHFrancis ATTACHMENT* FAOE I NUMB f. Ft I 1 Of 2 |l'-2286 APPHOVLU y I JlJHorncr - Sample Chromatogram ISSUED ON NEW SCOPE This procedure determines low toilers in returned pyranol or therminol. The area of the low boilers are normalized. The rest of the GLC curve is compared to a standard curve to t^ot irregularities in contaminated material. K^ymiENT 1. GLC, Varian-Aerograph Model 202-1C, dual-column linear temperature programmed, equipped with a 1 MV, 1 second response time (or better) recorder and a 120 rpm Disc, integrator (specify when ordering). Varian-Aerograph, 205 W. Touhy Ave., Park Ridge, 111. 60068, (or equivalent). 2. GLC columns, two *l-foot by 1/JI inch 0.1). stainless steel, ^Mctnl Goods Corp., St. Louis, Missouri (or equivalent). 5. Column Packing, y$> GE-SE30 on Chromosorb G., A/W, DMCS treated, 60/80 mesh, Analabs, Inc., F 0. Drawer 5,397, Hamden, Conn., 06318 (or equivalent). Jl, Helium Bupply and regulator to deliver 90// pressure at 10o ce/ minute. 9. Syringe, Hamilton, No. 701N with 2 inch needles. Fisher Scientific Co., 121)! Ambassador Blvd., St. Louis, Mo., 6313? RE10KNTS 1, Benzene, Reagent Grade, Cat. No. B-2'19, Fisher Scientific Co. SAFETY Benzene is flammable. Keep away from open flames and spark Bourccb. Pyranol/Pydraul/Thermii.ol may contain toxic contaminants when returned. Handle with care and wash contacts with soap arid hot water immediately, INS--u',NT CONDITIONS Injection Tort Detector Oven 330 -) 9c 3'I50C HONS 073500 Product: Pyranol/Therminol (returned) Test: low Boilers by GI..C Method Mo. T-??66 Page ? of ? pages Bridge Current Plow Hate . Column Oven (initial). Temperature Program 160 ma 100 cc Helium/rnin. So^o lOC/min. to J'IOC DETERMINATION Dilute 5 ml of sample with an equal volume of benzene. Stopper and shake to mix well. Bring the instrument to the conditions specified and inject 5 ul of samplc/solvent mix. Start tne temperature program immediately and turn on the integration device. An attenuation of 8:1 should keep all components on scale (except the benzene). 'I. After all components have eluted determine the area under all the peaks (excluding benzene). The low.toilers elute between the benzene and biphenyl. Determine' the counts in this area. ' .6 Compare the GLC curve to a standard curve and report extra peaks as "extra peak at_,__1___minutes". (Exclude low boilers when reporting extra peaks.) CAljCULATlON % I.ow HollerB Counts for Dow Boilers X 100 foaJ-Coun5~.... * 1*1)18 is a normalized calculation and area correction factors and precision have not been determined, ReproduSi'bility is expected be J rjf. ;f the amount ~tczz%:.':. MGNS 073501 HONS 0 7 3 5 0 2