DIRECTIONS FOR TEST NO. Id-10-^2
SUBJECT: Speolflo Gravity of Liquid Aroolors
METHOD: Westphal Balance ; .
__ ____
iI
' General
.-
;
V Speolflo gravity; Is determined by means of the . . Westphal balanoe. The temperatures at which the specific
r gravities of liquid Aroolors are taken vary with the vis cosity Of the liquid. The specifications for eaoh Aroolor should State the temperature for both the Aroolor and the
f- Water to which comparison of gravity is made. In plant practice, specific gravities of Aroolors containing up to
54$ ohlorine are made at 65/65C while speolflo gravities of Aroolors containing 60-52$ chlorine are made at 9Q/90C.
> Specific gravities of any of the Aroolors may be made at 25/25 by use of the A.S.T.M. method for road tars, asphalt cements, soft tars pitches, serial designatlon D70-27, 1927 (described in Bureau of Standards Misoellaneous Publication No. 110, Standards and Speoiflcatlons fbr Nonmetallio Minerals and their products, page
125.)
. .--v Apparatus............:
.
Westpbal Balance; Similar to Elmer and Amend
Catalogue No. 16930. Plummet need not have theimometer
soale to 90C but should not be subjeot to Injury by
use at this temperature.
;^
Cylinder: Fyrex glass hydrometer jar 1.5" x 6*.
Thexmometer; Use thexmometer which has been checked for accuracy of scale in the range used.
Procedure
.
Place the westphal in such position that the
plummet is 7 - 9 inches above the work bench. By means of the leveling screw, adjust the Westphal balance to
show zero reading with a clean dry plummet In air. Suspend
the plummet during adjustment in a empty cylinder to pro tect the apparatus from air currents.
Heat the Aroolor to be tested to a temperature
about 15 higher than the temperature at which determin
ation is to be made. For example, if specific gravity Is
wanted at 65C, the Aroolor should be heated to about 8dC.
Then pour the heated Aroolor into a clean Pyrex cylinder
whloh has been warmed by setting at edge of hot plate.
. - .
. ..
'Ww?*;"'7
*
.
..
DSW 330747
v
STLCOPCB4077391
Page 2 DIRECTIONS FOR TEST NO. H-10-^2 SUBJECT; Speolflo Gravity of liquid Aroclors METHOD: Westphal Balance
The cylinder should be filled with the Aroolor to within
about 1 inch of the top. , \ v>s-V`
'
v. . . `
' /I*
' v v-'y
: >av.`,.'A>ihwv^`
. . . - :-.-',;.C;.-.
Hr .. ^'H^rjoiow 'tke' Aroolor n the cylinder to air cool
until the temperature is about 7 degrees above that de~
sired, then piaoe the cylinder in a hot water bath which
is at a temperature about 5 degrees above the desired tem
perature. A liter beaker containing sufficient water to
be equal to or above the level of the Aroolor in the cy
linder is a satisfactory water bath. In case the Specific
Gravity is to be taken at 90/90, a glycerin bath should
be substituted for the water bath to avoid the water vapor
with its condensation on the Westphal balance which is
present if water Is used at high temperatures such as 95C
Without disturbing the adjustment of the Westphql
whloh is set to read zero with plummet in air, place the cylinder containing the Aroolor in such position that the plummet is immersed in Aroolor end add weights to the beam of the Westphal until the beam is again level. The plummet should now be completely immersed as well as about 1/2" of the suspension wire and the plummet should not touch the cylinder at any poiht. Stir the Aroolor with a thermometer at frequent intervals until the desired temperature just is reached. At this point the weights are quickly adjusted un til equilibrium is obtained and the reading noted.
Saving obtained the reading for Aroolor at the desired temperature, the specific gravity is obtained by
dividing this figure by the reading when the plummet is
immersed in pure water at the same temperature. The read
ing fbr water is obtained In the same manner as for Arocloi
Sinoe the reading for water is constant, provided there
are no errors in weights or Westphal, this figure need not - be determined with each Aroolor test. The figure for watei
which Is used as a oonstant should, however, be checked
once amopth or whenever greatest accuracy is needed. The
readings fbr water at 65C and $Q9G are respectively 9o2o
and *972.
' .-
Calculation of Result
Example 1. Assume readings as follows: Westphal reading with plummet in air Westphal reading with plummet in Aroolor at 65C Westphal reading with plummet in water at 5*0 Then sp.gr. at 65/65C l.|02^ 1*5283
Zero 1*5025
*9o2o
''
^^
DSW330748
STLCOPCB4077392
.* I /= / *.
r... - '.
zz.&.-, . .
-
3Pag . DIRECTIONS FOR TEST NO, 14-10-52
SUBJECT; Specific Gravity of Liquid Aroolors METHOD: ffeatphal Balance
, Example 2* Assume readings as follows:
.
/' v> t> -
-. 4
-
tfestphal reading with plismaet In air S*-,"-iWestphal reading with plummet in Aroolor at gOC Westphal reading with plummet In water at 900 Then sp. gr. at 9O/9O0C 1.5435 * 1*5958
Zero
1*5435 *9of2
v..,
. ,
:
Notes
- / .. .-
- . , '
1. The weights used with the Westphal halanoe
should be checked periodically by weighing each on an
analytical balanoe.
. .
^
2. The adjustment of the liquid to the exact temperature desired is most important. Do not allow
to cool rapidly and stir frequently enough to maintain a uniform temperature from top to bottom in the cylinder.
4BG-hp