Document YrmmbB5nv3L2E0b9RdLmqor9O

3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Final Report Analysis of PFOA, PFBS, PFHS, and PFOS in Aqueous Samples from 3M Decatur, Alabama: Groundwater Treatment Plant GAC System March 2013 Laboratory Request Number: ISO13-07-03 Method Requirement: 3M Method ETS-8-044.1 Report Date - Date of Last Signature Testing Laboratory 3M EHS&S Operations 3M Environmental Laboratory Building 260-5N-17 Maplewood, MN 55144-1000 Requester Gary Hohenstein 3M Building 224-5W-03 Saint Paul, MN 55144-1000 Phone: (651) 737-3570 The testing reported herein meet the requirements of ANSI/ISO/IEC 17025:2005 "General Requirements for the Competence of Testing and Calibration Laboratories", in accordance with the A2LA Testing Certificate # 2052.01. Testing Cert #2052.01 PAGE 1 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 3M Environmental Laboratory 3M Environmental Laboratory Technical Director: William K. Reagen, Ph.D. 3M Principal Analytical Investigator: Susan Wolf Report Author: Chelsie Grochow Analytical Report IS13-07-03 3M Decatur Groundwater Treatment Plant System; March 2013 Sampling Report Date: Date of Last Signature 1 Introduction/Summary The 3M Environmental Laboratory analyzed water samples from the groundwater treatment plant GAC system at the 3M Decatur, AL facility. Water samples were collected by Weston Solutions personnel on March 27, 2013. Samples were returned to the 3M Environmental Laboratory for analysis of Perfluorooctanoate (PFOA), Perfluorobutane sulfonate (PFBS), Perfluorohexane sulfonate (PFHS), Perfluorooctane sulfonate (PFOS), and the surrogate recovery standards 13C4-PFOA and 13C4-PFOS under laboratory project number ISO13-07-03. Analysis of the samples was completed as specified under 3M Environmental Laboratory General Project Outline (GPO) number GPO11-1006. The 3M Environmental Laboratory provided sample containers for five sampling locations; ACI, ACM2, ACE, WW8R, and WW9R. The sample set consisted of a field sample, field sample duplicate and a field matrix spike. Each empty container was marked with a "fill to here" line that corresponded to a final volume of 200 mL. Containers reserved for field matrix spikes were fortified with an appropriate matrix spike solution containing all analytes prior to being sent to the field for sample collection. Sample bottles for sampling locations ACM2 and ACE, along with the Trip Blank sample and low spike, were fortified with internal standards and surrogates prior to being sent to the field for sample collection. Samples were prepared and analyzed by method ETS-8-044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis". Select internal standards were used to aid in the data quality objectives. Table 1 summarizes the sample results using the analytical method identified above. All results for quality control samples prepared and analyzed with the samples will be reported and discussed elsewhere in this report. The testing reported herein meet the requirements of ANSI/ISO/IEC 17025:2005 "General Requirements for the Competence of Testing and Calibration Laboratories", in accordance with the A2LA Testing Certificate # 2052.01. Testing Cert #2052.01 PAGE 2 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 1. Sample Results Summary 3M LIMS ID Sample Description ISO13-07-03-001 ISO13-07-03-002 ISO13-07-03-004 ISO13-07-03-005 ISO13-07-03-007 ISO13-07-03-008 ISO13-07-03-010 ISO13-07-03-011 ISO13-07-03-013 ISO13-07-03-014 DAL-GW-ACI; Sample DAL-GW-ACI; Sample Dup Average %RPD Sample/Sample Dup DAL-GW-ACM2; Sample DAL-GW-ACM2; Sample Dup Average %RPD Sample/Sample Dup DAL-GW-ACE; Sample DAL-GW-ACE; Sample Dup Average %RPD Sample/Sample Dup DAL-GW-WW8R; Sample DAL-GW-WW8R; Sample Dup Average %RPD Sample/Sample Dup DAL-GW-WW9R; Sample DAL-GW-WW9R; Sample Dup Average %RPD Sample/Sample Dup PFOA Concentration (ng/mL) 2190 2250 2220 (1) 2.7 52.9 50.9 51.9 (12) 3.9 0.0362 0.0398 0.0380 (3) 9.5 3660 3620 3640 (1) 1.1 3260 3400 3330 (1) 4.2 PFBS Concentration (ng/mL) 1060 1060 1060 (1) 0.0 2300 2330 2320 (1,2) 1.3 <0.0250 <0.0250 <0.0250 (3) NA 380 389 385 (1) 2.3 273 283 278 (1) 3.6 PFHS Concentration (ng/mL) 1080 1090 1090 (1) 0.92 170 171 171 (1,2) 0.59 0.0433 0.0521 0.0477 (3) 18 1760 1760 1760 (1) 0.0 1410 1470 1440 (1) 4.2 PFOS Concentration (ng/mL) 2450 2520 2490 (1) 2.8 0.0509 0.0526 0.0518 (3) 3.3 <0.0232 <0.0232 <0.0232 (3) NA 5170 5070 5120 (1) 2.0 4530 4780 4660 (1) 5.4 NA = Not Applicable (1) Reported results were prepared and analyzed by ETS-8-044.1 using external standard calibration. The analytical method uncertainties associated with the reported results are as follows: PFOA 18%, PFBS 21 %, PFHS 15%, and PFOS 24%. (2) Sample did not have an appropriate FMS to access recovery. See section 4 for more details. (3) Reported results were prepared and analyzed by ETS-8-044.1 using internal standard calibration. The analytical method uncertainties associated with the reported results are as follows: PFOA 16%, PFBS 22%, PFHS 14%, and PFOS 24%. 2 Methods - Analytical and Preparatory 2.1 Methods Analysis was completed using 3M Environmental Laboratory Method ETS-8-044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis". Table 2 Target Analytes. Target Analytes Perfluorooctanoic Acid (C8 Acid) Perfluorobutanesulfonate (C4 Sulfonate ) Perfluorohexanesulfonate (C6 Sulfonate) Perfluorooctanesulfonate (C8 Sulfonate) Acronym PFOA PFBS PFHS PFOS Reference Material Structure Linear + Branched Linear Linear Linear + Branched PAGE 3 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 2.2 Sample Collection Samples were collected on March 27, 2013 in NalgeneTM (high-density polyethylene) bottles prepared at the 3M Environmental Laboratory. Prior to sample collection, bottles designated for field matrix spikes were spiked in the laboratory with a known volume of an appropriate matrix spiking solution containing PFOA, PFBS, PFHS, and PFOS. Sample bottles for sampling locations ACM2 and ACE, along with the Trip Blank sample and low spike, were fortified with a mixture of mass-labeled internal standards at a nominal concentration of 1 ng/mL and a mixture of surrogate recovery standards at a nominal concentration of 0.1 ng/mL. Collected sample bottles were received by the laboratory at ambient conditions on April 1,2013. 2.3 Sample Preparation Sample sets ACI, ACM2, WW8R, WW9R, and the Trip Blank high spike were diluted 1:100 and analyzed for all analytes. The 1:100 dilutions were prepared by removing a 0.1 mL aliquot of the well mixed sample and adding 9.8 mL of methanol and 0.1 mL of a surrogate spiking solution. Prepared samples were analyzed by external standard calibration. The ACE and ACM2 sample sets and the Trip Blank sample and low spike were prepared and analyzed for all analytes by transferring an aliquot of the well mixed sample into an autovial. Prepared samples were analyzed by internal standard calibration. During the preparation of the laboratory control samples for internal standard calibration, an aliquot of a separate internal standard spiking solution was added to the laboratory control samples (nominal concentration of 1 ng/mL). The sample bottles for ACE, ACM2, and Trip Blank were spiked with an internal standard mix at a nominal concentration of 1 ng/mL prior to being sent to the field for sample collection. The laboratory control samples were analyzed in the same manner as the samples. 2.4 Analysis All samples and quality control samples were analyzed for four target analytes using high performance liquid chromatography/tandem mass spectrometry (HPLC/MS/MS). Pertinent instrument parameters, the liquid chromatography gradient program, and the specific mass transitions analyzed are described in the tables below. Due to the nature of the sample, the wide range of concentrations found in the sample, and the environmental occurrence of multiple isomers of the laboratory's analytes of interest, the software used for processing the analytical results is not able to consistently integrate the analytical peak, manual integration of the analytical peak is necessary. All manual integrations are performed following the procedures outlined in method ETS-12-010. The consistency of the laboratory's integration is ensured through the training of laboratory personnel, the peer review process required for all manual integrations, the review of manual integrations by the QAU, and where necessary the review of manual integrations by laboratory management. PAGE 4 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 3. Instrument Parameters. Instrument Name Analysis Dates Analytical Method Liquid Chromatograph Guard column Analytical column Injection Volume Mass Spectrometer Ion Source Polarity Software ETS Kirk 4/3/13 - external standard calibration ETS-8-044.1 Agilent 1260 Betasil C18 (4.6 mm X 100 mm), 5 x Betasil C18 (4.6 mm X 100 mm), 5 x 2, 5, or 20 xL Applied Biosystems TripleQuad 5500 Turbo Spray Negative Analyst 1.6.1 ETS Kirk 4/4/13 - internal standard calibration ETS-8-044.1 Agilent 1260 Betasil C18 (4.6 mm X 100 mm), 5 x Betasil C18 (4.6 mm X 100 mm), 5 x 20 xL Applied Biosystems TripleQuad 5500 Turbo Spray Negative Analyst 1.6.1 Table 4. Liquid Chromatography Gradient Program. Step Number 0 1 2 3 4 5 6 7 8 9 Total Time (min) 0.00 0.50 4.00 6.00 11.0 13.0 13.5 16.0 16.5 19.0 Flow Rate (pL/min) 750 750 750 750 750 750 750 750 750 750 Percent A (2 m M ammonium acetate) 90.0 90.0 70.0 70.0 20.0 20.0 10.0 10.0 90.0 90.0 Percent B (Methanol) 10.0 10.0 30.0 30.0 80.0 80.0 90.0 90.0 10.0 10.0 Table 5. Mass Transitions Analyte Mass Transition Q1/Q3 Internal Standard (1) Mass Transition Q1/Q3 413/369 PFOA 413/219 [13Cs]PFOA 421/376 413/169 PFBS 299/80 299/99 [1 8 O2 ]PFBS 303/84 PFHS 399/99 399/80 [13Cs]PFHS 402/80 499/99 PFOS 499/80 [13Cs]PFOS 507/80 499/130 [13C4]PFOA surrogate 417/372 [13Cs]PFOA 421/376 [13C4]PFOS surrogate 503/80 [13Cs]PFOS 507/80 Dwell time was 50 msec for each transition. The individual transitions were summed to produce a "total ion chromatogram" (TIC), which was used for quantitation. (1) Internal standards were used for the analysis on 4/4/13. PAGE 5 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 3 Data Analysis 3.1 Calibration Internal Standard Calibration Analysis: Samples were analyzed against a matrix-matched stable isotope internal standard calibration curve. Calibration standards were prepared by spiking known amounts of stock solutions into 50 mL of laboratory prepared reagent water. The calibration standards contained an internal standard mix at a nominal concentration of 1 ng/mL. A total of thirteen calibration standards ranging from 0.025 ng/mL to 25 ng/mL (nominal) were analyzed with the prepared samples. Of these thirteen calibration standards, eleven contained the surrogates at concentrations ranging from 0.025 ng/mL to 10 ng/mL nominal. A quadratic, 1/x weighted, calibration curve of the ratio of the standard peak area counts over the internal standard peak area counts was used to fit the data for each analyte. The data were not forced through zero during the fitting process. Calculating the standard concentrations using the peak area ratios and the resultant calibration curve confirmed accuracy of each curve point. The reference standards of PFOA and PFOS used to prepare the calibration standards consisted of both linear and branched isomers. Each curve point was quantitated using the overall calibration curve and reviewed for accuracy. Method calibration accuracy requirements of 10025% (10030% for the lowest curve point) were met for all analytes. The correlation coefficient (r) was greater than 0.995 for all analytes. External Standard Calibration Analysis: Samples were analyzed against an external standard calibration curve. Calibration standards were prepared by spiking known amounts of the stock solution containing the target analytes into 90:10 methanol: Milli-Q laboratory water. A total of ten spiked standards ranging from 0.02 ng/mL to 25 ng/mL (nominal) were analyzed. The ten standards also contained the surrogates at concentrations ranging from 0.02 ng/mL to 25 ng/mL nominal. A quadratic, 1/x weighted, calibration curve of the standard peak area counts was used to fit the data for each analyte. The data were not forced through zero during the fitting process. Calculating the standard concentrations using the peak area counts and the resultant calibration curve confirmed accuracy of each curve point. Each curve point was quantitated using the overall calibration curve and reviewed for accuracy. Method calibration accuracy requirements of 10025% (10030% for the lowest curve point) were met for all analytes. The correlation coefficient (r) was greater than 0.995 for all analytes. 3.2 System Suitability A calibration standard was analyzed four times at the beginning of the analytical sequence to demonstrate overall system suitability. The acceptance criteria for system suitability samples is less than or equal to 5% relative standard deviation (RSD) for peak area/ratio and retention time criteria of less than or equal to 2% RSD. These criteria were met for all analytes. 3.3 Limit of Quantitation (LOQ) The LOQ as defined in method ETS-8-044.1 is the lowest non-zero calibration standard in the curve that meets linearity and accuracy requirements and for which the area counts are at least twice those of the appropriate blanks. The LOQs associated with the sample analysis are listed in Table 6. PAGE 6 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 6. LOQ Analyte PFOA PFBS PFHS PFOS PFOA PFBS PFHS PFOS Calibration Internal standard Internal standard Internal standard Internal standard External standard External standard External standard External standard LOQ, ng/mL 0.0240 0.0250 0.0250 0.0232 0.0192 (1) 0.0200 (1) 0.0200 (1) 0.0185 (1) (1) The dilution factor was not applied to the LOQ. 3.4 Continuing Calibration During the course of the analytical sequence, several continuing calibration verification samples (CCVs) were analyzed to confirm that the instrument response and the initial calibration curve were still in control. The method acceptance criteria of 100% 25% was met for all analytes. 3.5 Blanks Two types of blanks were prepared and analyzed with the samples: method blanks/solvent blanks and field/trip blanks. Each blank result was reviewed and used to evaluate method performance to determine the LOQ for each analyte. 3.6 Lab Control Spikes (LCSs) Low, mid, and high lab control spikes were prepared for the target analytes and analyzed in triplicate, while only low and high lab control spikes were prepared for the surrogates. LCSs were prepared by spiking known amounts of the analytes into either 10 mL of laboratory-prepared matrix matched water or 1 mL of Milli-Q water to produce the desired concentration. When applicable, the LCSs were then diluted with methanol in the same manner as the samples. Method ETS-8-044.1 states that the average recovery of LCSs at each spiking level be within 80%-120% with a RSD <20%. All LCS samples met method accuracy criteria except where noted below. External standard analysis: Mid level LCSs did not meet average recovery criteria for PFBS (121%) and PFOS (122%). High level LCSs did not meet average recovery criteria PFOS (124%). Internal standard analysis: Mid level LCSs did not meet average recovery criteria for PFOS (124%). High level LCSs did not meet average recovery criteria for PFBS (122%) and PFOS (122%). A method deviation is included with the raw data. All LCSs were used in the determination of the analytical method uncertainty in section 3.9. The following calculations were used to generate data in Table 7. Calculated Concentration * ____ _ LCS Percent Recovery -------------------------------------* 100% Spike Concentration LCS% RSD = standard deviation LCS replicates * 1QQ% average LCS recovery PAGE 7 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 7. Lab Control Spike Results. ETS-8-044.1 External Standard Calibration Analyzed 4/3/13 Lab ID PFOA (Linear + Branched) Spiked Calculated Concentration Concentration (ng/mL) (ng/mL) %Recovery LCS-130403-1 LCS-130403-2 LCS-130403-3 Average %RSD LCS-130403-4 LCS-130403-5 LCS-130403-6 Average %RSD 0.191 0.191 0.191 1.91 1.91 1.91 0.212 0.201 0.211 109% 3.0% 2.29 2.26 2.33 120% 1.7% 111 105 110 120 118 122 LCS-130403-7 LCS-130403-8 LCS-130403-9 Average %RSD 19.1 19.1 19.1 22.1 21.0 22.1 114% 3.0% 116 110 116 PFOS (Linear + Branched) Spiked Concentration (ng/mL) Calculated Concentration (ng/mL) %Recovery 0.185 0.185 0.185 0.198 0.199 0.200 108% 0.54% 107 108 108 1.85 2.24 1.85 2.24 1.85 2.28 122% 0.95% (1) 121 121 123 18.5 23.0 18.5 22.3 18.5 23.5 124% 2.4% (1) 124 121 127 ETS-8-044.1 External Standard Calibration Analyzed 4/3/13 Lab ID Spiked Concentration (ng/mL) PFBS Calculated Concentration (ng/mL) LCS-130403-1 LCS-130403-2 LCS-130403-3 Average %RSD LCS-130403-4 LCS-130403-5 LCS-130403-6 Average %RSD LCS-130403-7 LCS-130403-8 LCS-130403-9 Average %RSD 0.199 0.199 0.199 1.99 1.99 1.99 20.0 20.0 20.0 0.221 0.219 0.223 111% 0.90% 2.39 2.41 2.45 121% 1.3% (1) 22.9 22.1 23.8 114% 3.9% %Recovery 111 110 112 120 121 123 114 110 119 Spiked Concentration (ng/mL) 0.199 0.199 0.199 1.99 1.99 1.99 20.0 20.0 20.0 PFHS Calculated Concentration (ng/mL) 0.216 0.214 0.223 109% 2.1% 2.32 2.36 2.37 118% 0.98% 22.1 21.1 21.9 108% 2.4% (1) LCSs did not meet acceptance criteria of 100 20%. %Recovery 108 108 112 117 119 119 110 105 109 PAGE 8 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 7 continued. Lab Control Spike Results. ETS-8-044.1 External Standard Calibration Analyzed 4/3/13 Lab ID Spiked Concentration (ng/mL) PFOA surrogate Calculated Concentration (ng/mL) LCS-130403-1 LCS-130403-2 LCS-130403-3 Average %RSD LCS-130403-4 LCS-130403-5 LCS-130403-6 Average %RSD 0.199 0.199 0.199 1.99 1.99 1.99 0.202 0.206 0.208 103% 1.7% 2.09 2.13 2.12 106% 0.94% %Recovery 101 104 104 105 107 106 Spiked Concentration (ng/mL) PFOS surrogate Calculated Concentration (ng/mL) 0.190 0.190 0.190 0.196 0.197 0.195 103% 0.0% 1.90 2.06 1.90 2.05 1.90 2.05 108% 0.53% %Recovery 103 103 103 109 108 108 ETS-8-044.1 Internal Standard Calibration Analyzed 4/4/13 Lab ID PFOA (Linear + Branched) Spiked Calculated Concentration Concentration (ng/mL) (ng/mL) %Recovery LCS-130404-1 LCS-130404-2 LCS-130404-3 Average %RSD LCS-130404-4 LCS-130404-5 LCS-130404-6 Average %RSD 0.190 0.190 0.190 1.90 1.90 1.90 0.205 0.202 0.197 106% 1.9% 2.25 2.21 2.23 117% 0.85% 108 106 104 118 116 117 LCS-130404-7 LCS-130404-8 LCS-130404-9 Average %RSD 9.52 9.52 9.52 10.5 11.8 12.0 120% 6.8% 111 124 126 PFOS (Linear + Branched) Spiked Concentration (ng/mL) Calculated Concentration (ng/mL) %Recovery 0.184 0.184 0.184 0.217 0.215 0.218 118% 0.85% 118 117 119 1.84 2.35 1.84 2.17 1.84 2.32 124% 4.3% (1) 128 118 126 9.21 10.4 9.21 11.4 9.21 11.9 122% 6.7% (1) 113 124 129 (1) LCSs did not meet acceptance criteria of 100 20%. PAGE 9 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 7 continued. Lab Control Spike Results. ETS-8-044.1 Internal Standard Calibration Analyzed 4/4/13 Lab ID Spiked Concentration (ng/mL) PFBS Calculated Concentration (ng/mL) LCS-130404-1 LCS-130404-2 LCS-130404-3 Average %RSD LCS-130404-4 LCS-130404-5 LCS-130404-6 Average %RSD 0.198 0.198 0.198 1.98 1.98 1.98 0.215 0.221 0.217 110% 1.8% 2.34 2.34 2.44 120% 2.4% LCS-130404-7 LCS-130404-8 LCS-130404-9 Average %RSD 9.94 9.94 9.94 11.3 12.5 12.6 122% 5.9% (1) %Recovery 108 112 110 118 118 123 114 126 127 Spiked Concentration (ng/mL) 0.198 0.198 0.198 1.98 1.98 1.98 9.94 9.94 9.94 PFHS Calculated Concentration (ng/mL) 0.219 0.216 0.213 109% 1.4% 2.32 2.29 2.38 118% 1.8% 11.0 11.8 12.2 118% 5.2% %Recovery 111 109 108 117 116 120 111 119 123 ETS-8-044.1 Internal Standard Calibration Analyzed 4/4/13 Lab ID Spiked Concentration (ng/mL) PFOA surrogate Calculated Concentration (ng/mL) LCS-130404-1 LCS-130404-2 LCS-130404-3 Average %RSD 0.198 0.198 0.198 0.191 0.189 0.186 95.3% 1.3% LCS-130404-4 LCS-130404-5 LCS-130404-6 Average %RSD 1.98 1.98 1.98 1.95 2.01 2.00 101% 1.8% %Recovery 96.4 95.5 93.9 98.5 102 101 Spiked Concentration (ng/mL) PFOS surrogate Calculated Concentration (ng/mL) 0.189 0.189 0.189 0.186 0.186 0.190 99.3% 1.5% 1.89 1.95 1.89 1.87 1.89 1.92 101% 2.2% %Recovery 98.4 98.4 101 103 98.8 102 (1) LCSs did not meet acceptance criteria of 100 20%. PAGE 10 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 3.7 Analytical Method Uncertainty Analytical uncertainty is based on historical QC data that is control charted and used to evaluate method accuracy and precision. The method uncertainty is calculated following ETS-12-012.2. The standard deviation is calculated for the set of accuracy results (in %) obtained for the QC samples. The expanded uncertainty is calculated by multiplying the standard deviation by a factor of 2, which corresponds to a confidence level of 95%. The method uncertainty was determined to be 17% for PFBS and 16% for PFOS using internal calibration and 15% for PFBS and 20% for PFOS using external calibration, but based on the percent bias of the LCSs, the method uncertainty has been expanded for PFBS and PFOS. Table 8. Analytical Method Uncertainty. Analyte Calibration Method PFOA External Standard PFBS External Standard PFHS External Standard PFOS External Standard PFOA Internal Standard PFBS Internal Standard PFHS Internal Standard PFOS NA= Not Applicable Internal Standard Standard Deviation 8.86 NA 7.71 NA 8.02 NA 7.02 NA Method Uncertainty 18% 21% 15% 24% 16% 22% 14% 24% 3.8 Field Matrix Spikes (FMS) A field matrix spike was collected at each sampling point to verify that the analytical method is applicable to the collected matrix. Field matrix spikes are generated by adding a measured volume of field sample to a container spiked by the laboratory with the target analytes prior to shipping sample containers for sample collection. Field matrix spike recoveries within method acceptance criteria of 10030% confirm that "unknown" components in the sample matrix do not significantly interfere with the extraction and analysis of the analytes of interest. The standards used for the preparation of the field matrix spiking solutions contained reference materials comprised of both linear and branched isomers for PFOA and PFOS. Field matrix spikes are presented in section 4 of this report. FMS Recovery - (Sample Concentration of FMS - Average Concentration: Field Sample & Field Sample Dup.) * 1 0 0 % Spike Concentraton Table 9. Field Matrix Spike Concentrations Location ACI ACM2, ACE WW8R, WW9R Trip Blank Description Field Matrix Spike Field Matrix Spike Field Matrix Spike Low Field Matrix Spike High Field Matrix Spike Final Concentration (ng/mL PFOA PFBS PFHS PFOS 2110 1990 1970 2000 10.5 9.96 9.86 10.0 4220 3980 3940 4000 10.5 9.96 9.86 10.0 4220 3980 3940 4000 PAGE 11 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 4 Data Summary and Discussion The tables below summarize the sample results and field matrix spike recoveries for the three locations as well as the Trip Blank. Each table provides the average concentration and the relative percent difference (%RPD) of the sample and sample duplicate. Results and average values are rounded to three significant figures. Percent relative difference (%RPD) values are rounded to two significant figures. Because of rounding, values may vary slightly from those listed in the raw data. Field matrix spikes meeting the method acceptance criteria of 30%, demonstrate that the method is appropriate for the given matrix. All field matrix spike samples and surrogate recovery standards met method acceptance criteria except where noted below. DAL-GW-ACM2; the FMS sample (10 ng/mL) was not appropriate for PFOA, PFBS, or PFHS; however, the FMS sample for PFOS and the surrogate recovery standards 13C4-PFOA and 13C4-PFOS, met method acceptance criteria. No additional QC samples were prepared. PAGE 12 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 10. DAL GW ACI 130327 (1) 3M LIMS ID Description ISO13-07-03-001 DAL-GW-ACI-0-130327 ISO13-07-03-002 DAL-GW-ACI-DB-130327 ISO13-07-03-003 DAL-GW-ACI-FMS-130327 Average Concentration (ng/mL) %RPD PFOA PFBS Concentration (ng/mL) %Recovery 2190 NA 2250 NA 4210 94.3 2220 ng/mL 2.7% Concentration (ng/mL) %Recovery 1060 NA 1060 NA 2930 94.0 1060 ng/mL 0.0% 3M LIMS ID Description ISO13-07-03-001 DAL-GW-ACI-0-130327 ISO13-07-03-002 DAL-GW-ACI-DB-130327 ISO13-07-03-003 DAL-GW-ACI-FMS-130327 Average Concentration (ng/mL) %RPD PFHS PFOS Concentration (ng/mL) %Recovery 1080 NA 1090 NA 2950 94.7 1090 ng/mL 0.92% Concentration (ng/mL) %Recovery 2450 NA 2520 NA 4590 105 2490 ng/mL 2.8% 13C4-PFOA 13C4-PFOS 3M LIMS ID Description ISO13-07-03-001 DAL-GW-ACI-0-130327 ISO13-07-03-002 DAL-GW-ACI-DB-130327 ISO13-07-03-003 DAL-GW-ACI-FMS-130327 Average Concentration (ng/mL) %RSD %Recovery 100 104 96.8 100% 3.7% %Recovery 101 102 103 102% 0.91% NA = Not Applicable (1) Samples were analyzed by external standard calibration. The surrogate recovery standards 13C4-PFOA and 13C4-PFOS were added to the samples during sample preparation. PAGE 13 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 11. DAL GW ACM2 130327 3M LIMS ID Description ISO13-07-03-004 DAL-GW-ACM2-0-130327 ISO13-07-03-005 DAL-GW-ACM2-DB-130327 ISO13-07-03-006 DAL-GW-ACM2-FMS-130327 Average Concentration (ng/mL) %RPD 3M LIMS ID Description ISO13-07-03-004 DAL-GW-ACM2-0-130327 ISO13-07-03-005 DAL-GW-ACM2-DB-130327 ISO13-07-03-006 DAL-GW-ACM2-FMS-130327 Average Concentration (ng/mL) %RPD PFOA (1) PFBS (1) Concentration (ng/mL) %Recovery 52.9 NA 50.9 NA 61.9 NC 51.9 ng/mL 3.9% Concentration (ng/mL) %Recovery 2300 NA 2330 NA 2320 NC 2320 ng/mL 1.3% PFHS (1) PFOS (2) Concentration (ng/mL) %Recovery 170 NA 171 NA 176 NC 171 ng/mL 0.59% Concentration (ng/mL) %Recovery 0.0509 NA 0.0526 NA 10.1 100 0.0518 ng/mL 3.3% 13c 4-p f o a (1) 13c 4-p f o s (2) 3M LIMS ID Description ISO13-07-03-004 DAL-GW-ACM2-0-130327 ISO13-07-03-005 DAL-GW-ACM2-DB-130327 ISO13-07-03-006 DAL-GW-ACM2-FMS-130327 Average Concentration (ng/mL) %RSD %Recovery 103 103 103 103% 0.0% %Recovery 94.1 98.5 94.7 95.8% 2.5% NA = Not Applicable NC = Not Calculated; Spike level was less than 0.5x the endogenous sample concentration. (1) Samples were analyzed by external standard calibration. The surrogate recovery standards 13C4-PFOA and 13C4-PFOS were added to the samples during sample preparation. (2) Samples were analyzed by internal standard calibration. PAGE 14 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 12. DAL GW ACE 130327 (1) 3M LIMS ID Description ISO13-07-03-007 DAL-GW-ACE-0-130327 ISO13-07-03-008 DAL-GW-ACE-DB-130327 ISO13-07-03-009 DAL-GW-ACE-FMS-130327 Average Concentration (ng/mL) %RPD 3M LIMS ID Description ISO13-07-03-007 DAL-GW-ACE-0-130327 ISO13-07-03-008 DAL-GW-ACE-DB-130327 ISO13-07-03-009 DAL-GW-ACE-FMS-130327 Average Concentration (ng/mL) %RPD PFOA PFBS Concentration (ng/mL) %Recovery 0.0362 NA 0.0398 NA 10.2 96.4 0.0380 ng/mL 9.5% Concentration (ng/mL) %Recovery <0.0250 NA <0.0250 NA 9.85 98.9 <0.0250 ng/mL PFHS PFOS Concentration (ng/mL) %Recovery 0.0433 NA 0.0521 NA 9.31 93.9 0.0477 ng/mL 18% Concentration (ng/mL) %Recovery <0.0232 NA <0.0232 NA 10.4 104 <0.0232 ng/mL 13C4-PFOA 13C4-PFOS 3M LIMS ID Description ISO13-07-03-007 DAL-GW-ACE-0-130327 ISO13-07-03-008 DAL-GW-ACE-DB-130327 ISO13-07-03-009 DAL-GW-ACE-FMS-130327 Average Concentration (ng/mL) %RSD %Recovery 102 105 104 104% 1.5% %Recovery 112 110 107 109% 2.4% NA = Not Applicable (1) Samples were analyzed by internal standard calibration. PAGE 15 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 13. DAL GW WW8R 130327 (1) 3M LIMS ID Description ISO13-07-03-010 DAL-GW-WW8R-0-130327 ISO13-07-03-011 DAL-GW-WW8R-DB-130327 ISO13-07-03-012 DAL-GW-WW8R-FMS-130327 Average Concentration (ng/mL) %RPD 3M LIMS ID Description ISO13-07-03-010 DAL-GW-WW8R-0-130327 ISO13-07-03-011 DAL-GW-WW8R-DB-130327 ISO13-07-03-012 DAL-GW-WW8R-FMS-130327 Average Concentration (ng/mL) %RPD PFOA PFBS Concentration (ng/mL) %Recovery 3660 NA 3620 NA 7620 94.3 3640 ng/mL 1.1% Concentration (ng/mL) %Recovery 380 NA 389 NA 4100 93.4 385 ng/mL 2.3% PFHS PFOS Concentration (ng/mL) %Recovery 1760 NA 1760 NA 5520 95.4 1760 ng/mL 0.0% Concentration (ng/mL) %Recovery 5170 NA 5070 NA 9670 114 5120 ng/mL 2.0% 13C4-PFOA 13C4-PFOS 3M LIMS ID Description ISO13-07-03-010 DAL-GW-WW8R-0-130327 ISO13-07-03-011 DAL-GW-WW8R-DB-130327 ISO13-07-03-012 DAL-GW-WW8R-FMS-130327 Average Concentration (ng/mL) %RSD %Recovery 96.0 94.6 93.4 94.7% 1.4% %Recovery 101 101 96.0 99.3% 2.9% NA = Not Applicable (1) Samples were analyzed by external standard calibration. The surrogate recovery standards 13C4-PFOA and 13C4-PFOS were added to the samples during sample preparation. PAGE 16 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 14. DAL GW WW9R 130327 (1) 3M LIMS ID Description ISO13-07-03-013 DAL-GW-WW9R-0-130327 ISO13-07-03-014 DAL-GW-WW9R-DB-130327 ISO13-07-03-015 DAL-GW-WW9R-FMS-130327 Average Concentration (ng/mL) %RPD 3M LIMS ID Description ISO13-07-03-013 DAL-GW-WW9R-0-130327 ISO13-07-03-014 DAL-GW-WW9R-DB-130327 ISO13-07-03-015 DAL-GW-WW9R-FMS-130327 Average Concentration (ng/mL) %RPD PFOA PFBS Concentration (ng/mL) %Recovery 3260 NA 3400 NA 7570 100 3330 ng/mL 4.2% Concentration (ng/mL) %Recovery 273 NA 283 NA 4260 100 278 ng/mL 3.6% PFHS PFOS Concentration (ng/mL) %Recovery 1410 NA 1470 NA 5470 102 1440 ng/mL 4.2% Concentration (ng/mL) %Recovery 4530 NA 4780 NA 9070 110 4660 ng/mL 5.4% 13C4 -PFOA 13C4 -PFOS 3M LIMS ID Description ISO13-07-03-013 DAL-GW-WW9R-0-130327 ISO13-07-03-014 DAL-GW-WW9R-DB-130327 ISO13-07-03-015 DAL-GW-WW9R-FMS-130327 Average Concentration (ng/mL) %RSD %Recovery 102 105 96.3 101% 4.5% %Recovery 103 106 101 103% 2.2% NA = Not Applicable (1) Samples were analyzed by external standard calibration. The surrogate recovery standards 13C4-PFOA and 13C4-PFOS were added to the samples during sample preparation. PAGE 17 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Table 15. Trip Blank Results 3M LIMS ID ISO13-07-03-016 ISO13-07-03-017 ISO13-07-03-018 Description DAL-GW-TRIP01 -0 DAL-GW-TRIP01 -LS DAL-GW-TRI P01-HS 3M LIMS ID ISO13-07-03-016 ISO13-07-03-017 ISO13-07-03-018 Description DAL-GW-TRI P01-0 DAL-GW-TRI P01-LS DAL-GW-TRI P01-HS PFOA PFBS Concentration (ng/mL) <0.0240 (1) 10.2 (1) 4360 (2) %Recovery NA 96.4 103 Concentration (ng/mL) <0.0250 (1) 9.60 (1) 4070 (2) %Recovery NA 96.4 102 PFHS PFOS Concentration (ng/mL) <0.0250 (1) 9.10 (1) 4020 (2) %Recovery NA 92.3 102 Concentration (ng/mL) <0.0232 (1) 9.44 (1) 4650 (2) %Recovery NA 94.4 116 13C4 -PFOA 13C4 -PFOS 3M LIMS ID ISO13-07-03-016 ISO13-07-03-017 ISO13-07-03-018 Description DAL-GW-TRI P01-0 DAL-GW-TRI P01-LS DAL-GW-TRI P01-HS %Recovery 94.9 (1) 91.0 (1) 102 (2) NA = Not Applicable (1) Samples were analyzed by internal standard calibration. (2) Samples were analyzed by external standard calibration. %Recovery 98.5 (1) 94.0 (1) 105 (2) PAGE 18 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. IS013-07-03 5 Conclusion Laboratory control spikes were used to determine the analytical method accuracy and precision for all analytes. The accuracy and precision were then used to estimate the method uncertainty for the results. Field matrix spike recoveries demonstrated that the analytical method was appropriate for the given sample matrix. Analysis was completed using 3M Environmental Laboratory Method ETS-8044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis". Analytical results are reported in Table 1 and 10-15 of this report. 6 Data / Sample Retention All remaining sample and associated project data (hardcopy and electronic) will be archived according to 3M Environmental Laboratory standard operating procedures. 7 Attachments Attachment A: Analyte Trend Data for select sampling points at the Decatur Ground W ater Treatment Plant (GWTP) GAC System. 8 Signatures Susan T. Wolf, 3M Qjncipal Analytical Investigator Date William K. Reagen, Ph.D., 3M Environmental Laboratory Technical Director Date The 3M Environmental Laboratory's Quality Assurance Unit has audited the data and report for this project. Ol^ J Quality Assurance Represent!ative / Date This test report shall not be reproduced except in full, without written approval of the 3M Environmental Laboratory. PAGE 19 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Attachment A: Analyte Trend Data for select sampling points at the Decatur Ground Water Treatment Plant (GWTP) GAC System ACI (Activated Carbon Influent) - Sample Results in ng/mL ACI PFOA PFBS PFHS PFOS 6/17/2011 1410 1110 983 691 2/8/2012 1180 792 1100 782 6/29/2012 1650 1410 1460 964 8/28/2012 1410 1830 1390 172 12/14/2012 2580 723 1380 2810 1/17/2013 1910 588 895 1810 2/13/2013 2300 830 1580 2940 3/27/2013 2220 1060 1090 2490 ACE (Activated Carbon Effluent) - Sample Results in ng/mL ACE PFOA PFBS PFHS PFOS 6/17/2011 2.57 11.9 2.50 2.07 2/8/2012 1250 818 1240 741 8/28/2012 0.387 0.256 0.552 0.103 The ACE location was not sampled on 6/29/12. 12/14/2012 0.102 <0.0250 0.117 0.0376 1/17/2013 <0.0240 <0.0250 <0.0250 <0.0232 2/13/2013 0.0124 0.0360 0.178 0.0329 3/27/2013 0.0380 <0.0250 0.0477 <0.0232 PAGE 20 OF 21 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-07-03 Attachment A: Analyte Trend Data for select sampling points at the Decatur Ground Water Treatment Plant (GWTP) GAC System WW8R - Sample Results in ng/mL W W 8R PFOA PFBS PFHS PFOS 12/13/2012 3630 441 1740 4920 1/17/2013 4350 410 1930 5710 2/13/2013 3250 354 1900 4850 WW9R - Sample Results in ng/mL 3/27/2013 3640 385 1760 5120 W W 9R PFOA PFBS PFHS PFOS 12/13/2012 3360 321 1480 4080 1/17/2013 2710 184 954 3690 2/13/2013 2850 236 1400 4100 3/27/2013 3330 278 1440 4660 PAGE 21 OF 21