Document Naz128aXZvayENBn8OEL8mv8
AR226-2727
LOW LEVEL DETERMINATION OF FC-143 IN WATER October 1991
rH?M HILLr INC. MONTGOMERY, ALABAMA. 36116
LOW LEVEL DETERMINATION FC-143 IN WATER
Scope and Application
1.1 This method uses capillary gas chromatography with electron capture detection (GC/ECD) to analyze water samples for the presence of FC-143. FC-143 is a technical grade of ammonium perfluorooctanoate whose most abundant component has CAS Number 3825-26-1. The use of this method should be restricted to only those analysts who can demonstrate acceptable precision and accuracy with four spike replicates into laboratory water at the 1 ppb level. Sample results are reported without blank subtraction and without correction for surrogate recoveries.
1.2 The estimated limit of detection for FC-143 Is lower than 0.1 ug/L.
Summary of the Method
2.1 A one-liter aliquot of sample is fortified with surrogates, acidified, and extracted by partitioning with an organic solvent. The extract is concentrated to small volume before adding ammonium hydroxide and further concentrating to dryness. Alcoholic HCl is added to the dry residue and heated for one hour to form the ethyl ester. Several cleanups are described as part of the overall procedure to remove interferences from the extract before analysis using GC/ECD.
Safety
3.1 The hazard of each reagent used in this method is not accurately known. Exposure to the chemicals should be held to the smallest practical level. Safety data sheets for all materials must be made available to the personnel involved in the chemical analysis.
3.2 Any unfamiliar water sample may offer dangerous native contents beyond the list of chemicals listed in this method.
Apparatus and Materials
4.1 Samples should be collected in the field in accordance with the quality assurance project plan. Sample containers should be amber glass or similar inert material which will not offer interferences for the analysis.
4.2 Glassware
4.2.1 Optional all glass continuous liquid/liquid extractor
1
designed to extract one liter of water sample using dichloromethane solvent.
4.2.2 Separatory funnel, 2-L size, PTFE stopcock.
4.2.3
Erlenmeyer flask, 1-L size with calibration mark
etched at the 1.0 liter volume.
4.2.4
Concentrator tube, Kuderna-Danish, 10-mL size,
graduated, ground glass joint.
425
Evaporative flask, Kuderna-Danish, 500--mL size,
purchased with a 24/40 ground glass stopper for the upper
joint.
4.2.5
Snyder column, Kuderna-Danish, three-ball macro,
ground glass joint, floodless type.
4.2.7 Snyder column, Kuderna-Danish, two-ball micro, ground glass joint, floodless type.
4.2.8 4.2.9
Pipets, various sizes and types. Autosampler vials, 2-mL size, PTFE-faced septum.
4.2.10 Screw-cap vials, 4-dram size, PTFE-lined cap.
4.2.11 Micro Reaction Vessels, graduated 5-mL size (Catalog #3-3299M, Supelco Inc., Beliefonte, PA).
4.2.12 Replacement PTFE-faced Rubber Septa for Micro Reaction Vessels, 20 mm size. (Catalog #2-3264M, Supelco Inc., Beliefonte, PA)
4.3 Analytical Balance.
4.4 Centrifuge, IEC HN-SII or eguivalent.
4.5 Heated water bath, steam delivery and temperature should be controllable, concentric ring covers for openings.
4.6 Nitrogen blowdown apparatus.
4.7 Gas Chromatographic System, ECD, split/splitless capillary injector and pneumatics, complete with a data system for collecting and processing chromatographic data.
4.8 Rtx--200 Primary column or eguivalent, 100% trifluoropropyl methyl polysiloxane stationary phase, 30 meters in length, 0.32 mm inside diameter, 0.5 urn film thickness. (Catalog #15039, Restek Corporation, Beliefonte, PA).
2
4.9 HP-5 Confirmation column or equivalent, 5% phenyl methyl polvsiloxane stationary phase, 25 meters in length, 0.32 m m inside diameter, 0.5 urn film thickness. (Hewlet Packard part number 10G91J-112, Telephone 1-800-227-9770).
Reagents
5.1 Blank water which does not produce an interference for parameters of interest.
5.2 Ether, pesticide quality or equivalent.
5.3 Hexane, pesticide quality or equivalent. Cleaned prior to use by briefly shaking 100 mL of hexane against 2 grams of Activity 1 Neutral Alumina.
5.4 Dichloromethane, pesticide quality or equivalent.
5.5 Acetone, pesticide quality or equivalent.
5.6 3% Ethanolic HC1. Prepared by adding 2.4 mL of Acetyl chloride (Instant Methanolic HCl Kit, Stock #18.053A, Alltech Associates, Deerfield, IL) dropwise to 56 mL of absolute ethanol (Florida Distillers, Lake Alfred, FL). Sealed tightly m an amber glass bottle, this reagent may be stored at room temperature for at least two weeks.
5.7 Sulfuric acid, ACS grade. Carefully dilute with an equal volume of laboratory blank water to prepare 1+1 sulfuric acrd for use.
5.8 Sodium hydroxide, ACS grade. Dissolve in laboratory blank water to prepare both 10 N NaOH and 1 N NaOH for use.
5.9 Ammonium hydroxide, ACS grade. Use as-received.
5.10 Florisil, Selective Adsorbent for Gas Chromatography, 100/200 mesh. (Baker Analyzed Catalog #M369-07). Activate at 130 C in a shallow tray overnight prior to use.
5.11 Alumina N (Neutral) Super I Activity. (Catalog #04580, Universal Scientific Inc., Atlanta, GA) . Use as received.
5.12 Boiling Chips, 10/40 mesh, heat at 400 C far 30 minutes before use to insure freedom from contamination.
5.13 Wide and narrow range of pH paper.
Standards
6 .1 Concentrated stock solutions may be prepared b y weighing the
primary neat materials and dissolving in ethanol.
Primary
3
standards may be secured from the following vendors.
6.1.1 PCR Incorporated, Gainesville, FL 1-800-331-6313
6.1.2 MTM Research Chemicals, Lancaster Synthesis, Windham, HH, 1-800-238-2324
6.1.3 ChexnService, West Chester, PA 215-692-3026
6.2 Surrogate spike solution. Prepare a three-component
surrogate spiking solution by diluting the concentrated stock of perfluorononanoic acid, perfluorodecanoic acid, ^ and ` h -
Eicosafluoroundecanoic acid. The final concentration of each
surrogate should be 10 ug/mL in ethanol. Note: Early work did not include the perfluorodecanoic
acid surrogate. '
.
6.3 Matrix spiking solution. Prepare an FC-143 spiking solution by diluting the concentrated stock to 10 ug/mL using ethanol.
6.4 Internal standard stock solution. Prepare a three-component Internal standard solution by diluting concentrated stocks of X, 2-Dibromoethane, 1,2--Dichlorobenzene, and 1,2,4-Trichlorobenzene. The final concentration of each internal standard should be 2
ug/mL, 40 ug/mL, and 4 ug/mL respectively in hexane. EXCEPTION - Recent work has discontinued 1,2Dibromoethane as an internal standard and substituted Ethyl perfluoroheptanoate at 20 ug/mL in this solution. A concentrated stock solution of the ester may be prepared by weighing approximately 10 mg of the acid directly into a reactor vial and following the derivatizing procedure beginning at procedure step 7.22.
6.5 Instrument calibration working standards may be prepared from either the acids or the salt primary materials by processing known masses through the derivatization steps of the sample preparation.
Sample Extraction - Separatory Funnel Technique
7.1 Set up a different 2-L separatory funnel for each sample and the blank. Using a graduated 1-L Erlenmeyer flask place 1 liter of sample into the separatory funnel.
7.2 Fortify each sample and the blank with 0.1 mL of the surrogate spiking solution (see 5.2). Fortify any matrix or blank spike aliquots with both surrogate and 0.1 mL of matrix spiking solution (see 6.3).
7.3 Adjust the sample to p H >12 with 10 N NaOH. Test the pH with pH paper.
4
7.4 Add IDO mL of dichloromethane to the separatory funnel and shake for two minutes to reach equilibrium. Vent the funnel frequently especially for the first few seconds of shaking to release the vapor pressure which will he produced. Allow the liquid phases to separate and discard the lower organic phase. If a significant emulsion develops, use a centrifuge to separate the
phases.
7.5 Adjust the sample to pH <1.5 with 1+1 sulfuric acid.
7.6 Rinse the 1-L Erlenmeyer flask used to aliquot the sample with 200 mL of ether before adding this ether to the separatory funnel.
7.7 Assemble a Kuderna-Danish (KD) apparatus by connecting a 10 m L concentrator tube to a 500-xnL evaporative flask. Place two or three boiling chips into the unit.
7.8 Shake the separatory funnel for two minutes and allow the
phases to separate before draining the sample (lower phase) back into its original Erlenmeyer. Drain the ether extract directly
into the assembled KD apparatus.
.
7.9 Return the aqueous sample to its separatory funnel. Rinse the Erlenmeyer flask with 100 mL of fresh ether before adding the ether to the separatory funnel for the last extraction. Repeat
step 7.8.
7.10 A d d 4.0 mL of 1 N NaOH and approximately 20 mL of fresh ether to the "empty" separatory funnel and shake (roll around)^ for several seconds. Allow at least two minutes for the liquid to drain down to the stopcock before adding the base and the ether to the KD apparatus.
7.11 Attach a three-ball macro Snyder column to the KD flask.
7.12 Concentrate the combined extract on the steam bath until boiling slows considerably and the KD begins to get warm. Only the aqueous liquid should be present in the receiver with ether
refluxing in the Snyder column.
7.13 With the KD still on the steam bath slowly add at least 5 mL
of dichloromethane into the macro Snyder column.
The
dichloromethane should be added slowly to avoided superheating with
any contact against the warm receiver region. Continue heating for
approximately one minute to exchange most of the residual ether.
7.14 Remove the KD from the heat and cool for two minutes or more before pouring 5 mL of dichloromethane down the macro Snyder column. Remove the Snyder column and add two or three new boiling chips. Replace the Snyder column and return to the steam bath. Heat the KD until only the aqueous phase appears to be present in
5
the receiver and the KD begins to warm. This procedure should remove the dissolved ether without floating the aqueous keeper out of the receiver into the KD regions.
7.15 Remove the KD from the heat for the last time and allow it to cool. Replace the Snyder column with a 24/4D stopper. Disassemble the receiver and transfer all liquid to a 4-dram vial. Add three mL of 1 N NaOH and three mL of fresh dichloromethane to the receiver before reconnecting it to the KD flask. Gently rotate the apparatus to rinse the interior walls. Vapor pressure will be produced which should be released frequently at the stoppered joint. After several seconds of rotation, return the apparatus to the upright position and allow a few minutes for the liquids to drain into the receiver. Remove the receiver from the KD and transfer the rinse liquids to the 4-dram vial. Additional fresh dichloromethane may be required to establish equal volumes of the
two liquid phases in the vial.
Note: ' The stopper, the KD, the Snyder, and the empty receiver
should be rinsed with water without delay to avoid future problems
with this reusable glassware- Do not allow the basic water to dry
on the glass surfaces.
.
7.16 Secure the cap on the 4-dram vial. Vigorously shake the vial for 30 seconds to reach equilibrium. Remember to vent any pressure which may be produced. Briefly centrifuge the vial to secure a good phase separation. Use a disposable 5-ml, pipet to remove the dichloromethane phase (lower) and discard it to waste.
7.17 Adjust the aqueous sample extract to pH <1.5 using 1+1 sulfuric acid. Verify the pH by placing a small drop of the adjusted sample onto p H indicator paper.
7.18 Place 5 mL of fresh ether into the vial containing the acidified aqueous extract. Recap the vial and shake for 30 seconds remembering to vent any pressure which may develop. After equilibration and phase separation, transfer the upper ether phase to a clean 10-mL receiver. CAUTION: Do not approach the liquid/liquid interface too closely. Severe problems may result
from the transfer of any aqueous phase.
7.19 Repeat step 7.18 for the last extraction. To the combined ether extract add 0.1 m L of concentrated ammonium hydroxide. Add two or three boiling chips to the receiver and attach a micro twoball Snyder column. Take the extract to the steam bath and concentrate it to approximately 0.5 mL volume. Remove the sample
from the heat and allow it to cool.
7.20 Transfer the extract to a clean 5-mi reactor vial. Complete the transfer quantitatively using' 0.5 mL of acetone.
7.21 Further concentrate the sample to dryness using a gentle
6
stream of clean nitrogen at room temperature. This process should take approximately one hour due to the small amount of water in the sample. Many samples will offer a visible residue at dryness. Each sample will require careful inspection to determine that all water has been removed. Some samples may have tiny droplets of water covered by a thin membrane of solid phase which can be broken by adjusting the stream of nitrogen. Experience will dictate the best way of judging the dryness of the sample. Nitrogen blowdown may be continued as long as one hour past dryness without loss of
target compounds.
7.22 Add 1.0 mL of ethanolic HCi reagent to the dry sample residue. Secure the screw cap to the vial and shake it for several seconds. Place the sample into an oven at 65 C for one hour to derivatize the compounds of interest to their respective ethyl
esters.
7.23 Remove the samples from the oven and allow the reactor vials to cool. Remove the cap and add 1.0 mL of clean hexane and 1.0 mL of laboratory blank water. Recap and shake vigorously for several seconds. Allow the phases to separate and discard most of the lower liquid phase keeping the upper "hexane" phase..
7.24 Add Florisil to a clean empty 2-mL autosampler vial. The amount of Florisil should provide approximately 1 m m of depth in
the vial. Transfer the "hexane" phase from the jreactor vial in
procedure step 7.23 to the vial containing Florisil. Be careful
not to transfer any residual aqueous phase which may still be
present. Shake the Florisil vigorously for a several seconds and
then allow it to settle.
.
IMPORTANT NOTE: The extract is stable against the Florxsil for
less than one hour. Leaving the extract; against Florisil for more
than several minutes will result in gradual loss of the compounds of interest. The "hexane" phase from procedure step 7.23 contains
a significant amount of ethanol and should not be diluted with pure
hexane before performing the Florisil adsorption step. If any
dilutions are needed, then they should be performed after removing
the extract from the Florisil vial.
7.25 Transfer a measured portion of the final extract away from . the Florisil and place it into an autosampler vial already containing internal standards. This final extract should be stable for at least forty days. The vial should be clearly labeled and is
ready for analysis.
Sample Extraction - Continuous Extractor Technique
8 .0 An alternative to separatory funnel extraction is continuous liquid/liquid extraction. FC-143 may be successfully extracted from an acidified water sample using a continuous extractor and
dichloromethane solvent.
7
8.1 Set up a different 1-Liter continuous extractor unit for each
sample and a blank. Place a few boiling chips in the round bottom
flask supported snugly by the heating mantle.
Attach the
extraction reservoir to the flask with sufficient support for safe
operation. Add approximately 300 mL of dichloromethane to the
reservoir.
8.2 Use a graduated Erlenmeyer flask to measure 1 liter of the
sample and transfer it into Fortify the sample with 0.1
the continuous extractor mL of surrogate solution
r(eseseerv6o.i2r)..
Fortify any matrix spike or blank spike sample with both surrogate
and 0.1 m L of matrix spiking solution (see 6.3).
8.3 Adjust the sample to pH <1.5 using 1+1 sulfuric acid.
8.4 Add extra dichloromethane to the reservoir if necessary to insure proper solvent circulation during the extraction. Attach the condenser to the reservoir and start cool water flow through the condenser jacket. Energize the heating mantle to maintain gentle solvent boiling for sixteen hours of continuous extraction.
8.5 When extraction time has expired, discontinuing heating and allow the unit to cool. Turn off the water to the condenser jacket. Disconnect the collection flask and transfer the extract directly to a KD apparatus using a small amount of acetone to
complete the transfer quantitatively.
8.6 Concentrate the extract using the steam bath. When the apparent volume of the extract reaches 2 mL, remove the KD from the
heat and allow it to cool.
8.7 Replace the 3-ball Snyder column with a 24/40 stopper. Disassemble the receiver and transfer the extract to a 4--dram vial. Add seven mL of 1 N NaOH and three mL of fresh dichloromethane to the receiver before reconnecting it to the KD flask. Gently rotate the apparatus to rinse the interior walls. Vapor pressure will be produced which should be released frequently at the stoppered joint. After several seconds of rotation, return the apparatus to the upright position and allow a few minutes for the liquids to drain into the receiver. Remove the receiver from the KD and transfer the rinse liquids to the 4-dram vial. Additional fresh dichloromethane may be required to establish equal volumesTMDf the
two liquid phases in the vial.
Note: The stopper, the KD, the Snyder, and the empty receiver
should be rinsed with water without delay to avoid future problems
with this reusable glassware. Do not allow the basic water to dry
on the glass surfaces.
.
8.8 Finish the sample preparation' in the same manner as that used for separatory funnel extractions. Continue at procedure step 7.16.
8
Instrument Calibration - Primary Analysis
9 .i Figure 1 through Figure 4 are example chromatograms with
recommended instrument conditions included in the legends.^ Figure
1 is a typical instrument calibration standard.
Figure 3
illustrates the multiple peak pattern of FC-143 using the primary
analytical column.
9.2 Instrument calibration and sample analysis must be performed using multiple internal standards. The three internal standard compounds listed in section 6.4 are recommended to establish both relative retention times (RUT) and relative response factors (ERF). Internal standards appearing in a chromatogram will establish primary search windows for those target compounds' nearby in the chromatogram. Relative retention times are calculated using
equation 2 .
RRT = RTtaxgeV RTIS
Eq.2
The relative response factor is calculated as follows.
Absolute Response Factor = RF = Amount/Area
Eq.3
Relative Response Factor = RRF = RFtar3Dt/RFIS
Eq.4
Note: Amount in equation 3 refers to the mass (e.g. ug) of compound mixed into the solution inj e c t e d .
9.3 Prepare initial calibration standards at a minimum of three concentration levels for both FC-143 and the surrogates. Analyze each level of calibration standard. For each compound, tabulate the RRF at each level (see equation 4). If the RRF over the working range is constant enough (less than 20% relative standard deviation), then the midpoint value may be used for calculation. Alternatively, the calibration data may be used to construct a calibration curve of relative response against relative amounts.
9.4 Continuing calibration standards must be injected after every ten sample extracts and at the end of an analytical sequence. Relative response factors must be compared to the initial calibration values. Treat the continuing calibration standard as an unknown sample and compute the concentration, injected. If the calculated concentration differs more than 20% from the true value, then initial calibration must be repeated and all affected samples
must be reanalyzed.
Instrument Calibration - Confirmation Analysis
10.0 FC-143 tentatively identified b y the primary analysis should be confirmed by at least one qualitative technique. Confirmation data may be supplied by analysis using a dissimilar analytical
9
column, or by gas chromatography with mass spectrometry detection (if the target concentration is adequate), or by a second derivatization forming a different ester. Confirmation analysis must be qualitative and may be quantitative.
10.1 GC Dissimilar Column Confirmation
10.1.1 Column: HP-5 (or equivalent) 5% phenyl methyl polysiloxane stationary phase, 25 meters in length, 0.32 mm inside diameter, 0.5 urn film thickness.
10.1.2 Chromatographic conditions:
Oven Program: 50 C for 2 minutes
to 90 C at 5 DC/minute
to 280 C at 10 DC/minute
(hold 2 minutes)
Injector Zone: 200 C
Detector Zone: 320 C
_
Helium carrier inlet pressure: 5.5 psig
1.0 uL splitless injection, split on at 0.5 minutes
10.1.3 The primary initial and continuing calibration standards and criteria are used for GC dissimilar column confirmation (sections 9-2 - 9.4)
10.1.4 The sample extracts requiring confirmation and the associated method blanks should be analyzed at the same dilutions as for the primary analysis.
10.2 GC/MS Confirmation
10.2.1 GC/MS analysis normally requires a minimum concentration of 10 ug/mL in the field sample extract. The sample extracts requiring confirmation and the associated method blanks must be analyzed. A reference standard for the target compound must also be analyzed by GC/MS. The concentration of the reference standard must be at a level that would demonstrate the ability to confirm the target.
10.2.2 To confirm the identification of FC-143, the background corrected mass spectrum of the compound must be obtained from the sample extract and compared with a mass spectrum from a stock or calibration standard analyzed under
the same chromatographic conditions.
10.3 Confirmation by Analysis of an Alternate Ester
10.3.1 Both standards and sample extracts may be derivat'ized
to an alternate ester before analysis according to the
following reaction.
RC00H + R'OH ----> RCOOR' + HOH
10
Choice of the alcohol to which acetyl chloride is added (section 5.5) will determine which ester is prepared at procedure step 7.22. The methyl, ethyl, n--propyl, and n --butyl esters of FC-143 have been prepared by using the corresponding alcoholic HC1. Only the ethyl ester has been processed through full method validation.
10.3.2 Before samples are confirmed by this technique, the laboratory must demonstrate the ability to successfully process and analyze representative standards.
10.4 Simultaneous Primary and Confirmation Analysis
10.4.1 A gas chromatograph using only one injection port, one column oven, and one detector can produce a chromatogram from two dissimilar columns installed in parallel. The resulting chromatogram is a superimposed (overlaid) composite of the separations achieved with each independent column. Each compound in the sample will produce two chromatographic peaks thus providing enhanced qualitative information for every
injection.
10.4.2 The recommended primary and confirmation columns may be connected in parallel using two "Y" connectors (part number 20405, Restek Corporation, Beliefonte, PA) and short sections of uncoated deactivated fused silica tubing. Alternatively, a 2--hole ferrule may be used at the injection port and at the
detector.
10.4.3 A n example chromatogram is presented in Figure 4. The
# symbol precedes the name of a chromatographic peak
produced by the confirmation column. Named peaks produced by
the primary, column will not include the
symbol.
Therefore, the FC-143 peak was produced by the Rtx-200 column,
and the # FC-143 peak was produced by the HP-5 column.
10.4.4 When a 2-hole graphite ferrule is used at the detector, peak identification may be assisted by temporarily removing one column from the detector for selected injections.
Quality Control
. ..,,at
. . . ,. . .
11.1 Every laboratory must operate a formal quality control
program which meets or exceeds the requirements of the data quality
objectives.
11.2 Before using this method, laboratory capability must be demonstrated by analyzing spiked samples. Seven replicates spiked at sufficiently low FC-143 concentration must be analyzed to estimate the method detection limit (Table 1 and Table 2). Figure 2 is an example chromatogram of blank water spiked to estimate the method detection limit using the separatory funnel extraction
11
technique. Also four additional replicates must be spiked at the matrix spike level and analyzed to create additional precision and
accuracy data available for audits (Table 3).
11.3 Ongoing performance evaluation will be accomplished by
analyzing a matrix spike and a matrix spike duplicate for every twenty field samples. A control sample will be extracted and
analyzed with every set of matrix spikes. Surrogates_ will be spiked into every field sample, matrix spike > matrix spike
duplicate, control sample, and laboratory blank.
Surrogate
recoveries will be control charted to determine control limits for
corrective action.
Gas Chromatography
12.1 The gas chromatographic conditions listed in Figure 1 ape recommended for primary analysis. Peaks tentatively identified in a sample require confirmation analysis (section 10.0).
12.2 Qualitative Analysis.
12.2.1 Peak identification is based upon relative retention time comparison to calibration data. Library retention times are established, on a daily basis. A nearby internal standard serves as a time reference for all chromatographic acquisitions. A target compound should be identified in a sample only if the chromatographic peak matches the predicted retention time within 0.05 minutes.
Predicted RT = (RRTlibraIY)(RTISBMple)
Eq.5
12.2.2
Equation 5 should serve only as a recommended
treatment for peak identification. The experienced analyst
nmst consider other chromatographic features such as peak
shape, resolution, distance from the time reference, and
multiple peak pattern recognition.
12.3 Quantitative Analysis.
12.3.1 Sample extracts should be prepared for analysis by
mixing 50 uL of internal standard solution (see G.4) into 0.5
mL of the sample extract.
^
12.3.2 The concentration of FC-143 identified in a sample may be calculated using equation 6. It is recommended that the nearest internal standard not suffering from co-elution be used for calculation, but the discretion of the analyst should
prevail.
sample
:
(ugISBafflpla)(Area/"*16)(RRFX) (AreaISEO"pla) (L extracted) (D.F. )
Eq.6
12
PPByaaiapl8 = concentration of target (in the^ sample)
_
u 9isSMiplB = mass of material (I.S.) mixed with the solution
injected.
#
AreaxBTM plB = peak area of target (in the sample)
RRFX = relative response factor (equation 4) j-ea b">p1b = peak area of internal standard ( m the sample)
L extracted = Total raw sample producing total extract
D.F. = dilution factor of the extract
12.3.3 The concentration of FC-143 identified in a sample may alternatively be calculated using the initial calibration
curve.
13
TABLE 1 LOW LEVEL SPIKING STUDY - SEPARATORY FUNNEL
FC-143 So iked
0.0 ppb
0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb
0.2 ppb 0.4 ppb
FC-143 Cone. Found fucr/L}
ND
0.052 0.052 0.061 0.059 0.058 0.052 0.062
0.11 0.23
FC-143 % Recoverv
N/A
62% 58% 61% 59% 58% 62% 62%
55%' 58%
ET-C9 SURR.
52%
51% 44% 49% 53% 50% 51% 54%
48% 53%
ET-C11 SURR-
47%
48% 41% 46% 47% 48% 47% 50%
48% 46%
ND - Not detected
Mean Concentration. FC-143 (seven 0.1 ppb replicates) = 0.059 ppb Standard Deviation FC-143 (seven 0.1 ppb replicates) = 0.0035 ppb Three std. dev. FC-143 (seven 0.1 ppb replicates) = 0.011 ppb
One liter of blank water was spiked and extracted
14
TABLE 2 LOW LEVEL SPIKING STUDY - CONTINUOUS EXTRACTOR
FC-143 Soiled
0.0 ppb 0.0 ppb
0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb 0.1 ppb
0.2 ppb 0.4 ppb 1 ppb 10 ppb 100 ppb
FC-143 Cone. Found iucr/L)
ND ND
0.041 0.039 0.045 0.042 0.045 0.043 0.041
0.076 0.155 0.670 8.94 96.3
FC-143 % Recoverv
N/A N/A
41% 39% 45% 42% 45% 43% 41%
38% 39% 67% 89% 96%
ET-C9 SURR.
50% 61%
64% 64% 69% 57% 67% 64% 67%
64% 60% 61% 81% 90%
ET-C11 SURR.
49% 64%
65% 58% 67% 50% 60% 64% 70%
68% 65% 68% 81% 83%
ND - Not detected
Mean Concentration FC-143 (seven 0.1 ppb replicates) = 0.042 ppb Standard Deviation FC-143 (seven 0.1 ppb replicates) = 0.0022 ppb Three std. dev. FC-143 (seven 0.1 ppb replicates) = 0.0066 ppb
One liter of blank water was spiked and extracted
15
TABLE 3 PRECISION & ACCURACY STUDY - SEPARATORY FUNNEL
FC-143 Spiked
1 PPb 1 PPb 1 PPb 1 PPb 1 PPb
FC-143 Cone. Found fucr/Li
1.10 0.88 1.01 1.03 0.84
FC-143 % Recovery
110% 88%
101% 103%
84%
ET-C9 SURR.
100% 78% 94% 98% 78%
ET-C11 5URR.
88% 69% 87% 92% 65%
One liter of blank water was spiked and extracted
16
?
>***.
Figure 1. GC/ECD chromatogram of instrument standard at 1 ug/mL
FC-143.
F u s e d _ silica capillary column;
Rtx-200 (Restek
Corporation), 30m x 0.32mm I.D., 0.5um film thickness. Column
programs 50C for 2 minutes to 90C at 5C/minute then to 280 C at
10C/minute and hold 2 minutes. Injector zones 200C. Detector
zone: 320C. Helium carrier inlet pressure = 5.5psig. luL
splitless injection, split on at 0.5 minutes.
17
Cll - Surrogate
Figure 2. GC/ECD chromatogram of blank water (extracted by separatory funnel) spiked at 0.1 ug/L FC-143. Fused, silica capillary column: Rtx-200 (Restek Corporation), 30m x 0.32mm I.D.^, 0.5um film thickness. Column program: 50C for 2 minutes to 90C at 5C/minute then to 280C at 10C/minute and hold 2 minutes. Injector zone: 200DC. Detector zone: 320C. Helium carrier inlet pressure = 5.5psig. luL splitless injection, split on at 0.5 minutes.
18
Figure 3. GC/ECD chromatogram of instrument standard at l _ug/mL and multiple peak pattern of F C --143. Fused silica capillary column: Rtx-200 (Restek Corporation)/ 30m x 0.32mm I-D.r 0.5urn. film thickness. Column program: 50DC for 2 minutes to 90C at 5C/minute then to 280C at 10C/minute and hold 2 minutes. Injector zone: 200C. Detector zone: 320C._ Helium carrier inlet pressure = 5.5psig, luL splitless injection, split on at 0.5 minutes.
19
I
(U +J te a>
Cil - Surrogate
Figure 4. GC/ECD chromatogram of instrument standard at 1 ug/mL
FC-143. Dual (parallel) capillary columns: Rtx-200 (Restek
Corporation), 30m x 0.32mm I.D., 0.5um film thickness. HP-5
(Hewlett Packard), 25m x 0.32mm I.D., 0.5um film thickness. Column
program: 50C for 2 minutes to 90C at 5C/minute then to 2B0C at
10C/minute and hold for 4 minutes. Injector zone: 200C.
Detector zone: 320C. Helium carrier inlet pressure = 5.5psig.
luL splitless injection, split on at 0.5 minutes. Peak naming
convention: The "#" symbol preceding the compound name indicates
a peak produced by the HP-5 capillary column.
.
20